The detection of cryptococcal capsular antigen (CrAg) is very sensitive and specific, however false-negative results have been reported, mostly in cerebrospinal fluid. We report the case of an HIV-infected patient with CD4 = 42 cells/mL, asthenic, negative serum CrAg lateral flow assay (LFA) and culture-proven cryptococcaemia. Despite the high accuracy of LFA, false-negative result is possible. Careful clinical evaluation and close follow-up are relevant.Entities:
Keywords: AIDS; Cryptococcosis; False negative reaction; Fungal antigens; Prozone effect
Year: 2019 PMID: 31737475 PMCID: PMC6849355 DOI: 10.1016/j.mmcr.2019.10.009
Source DB: PubMed Journal: Med Mycol Case Rep ISSN: 2211-7539
Fig. 1A) Lateral flow assay test showing the positive reaction (Day +173). B) Chest computed tomography (CCT) scan: a nodule of 2.2 × 1.2 cm in the upper lobe of the right lung (Day +173). C) Control CCT scan: nodular atelectatic opacity in the right upper lobe of healing aspect (Day + 502).
Laboratory exams results.
| Exams | Day 0 | Day +47 | Day +173 |
|---|---|---|---|
| TL CD4+ count (cells/mL) | 42 (5%) | - | 223 (11%) |
| Viral load (copies/mL) | 805,439 (5.9) | - | Undetectable |
| CrAg Latex agglutination - IMMY | Negative | - | Positive 1:64 |
| CrAg LFA - IMMY | - | Positive | |
| Blood Culture | Negative | Negative | |
| Haemoglobin (g/dL) | - | 11 | 16.3 |
| Haematocrit (%) | - | 35 | 48 |
| White blood cell (cells/mm3) | - | 4590 | 3630 |
| Platelets (cells/L) | - | 233 × 109 | 151 × 109 |
| Albumin (g/dL) | - | 2.79 | 3.72 |
| CrAg LFA (CSF) - IMMY | - | - | Negative |
Time zero: first hospitalization, day 0.
Time 1: first outpatient consultation, day +47.
Time 2: second outpatient consultation, day +173.
Immuno-Mycologics, OK, USA.