Eknath D Pawar1, Amit Kumar Yadav2, Arohi Sharma3, Abhishek Harsoor4.
Abstract
INTRODUCTION: Revision rates of total hip arthoplasty have decreased after introducing total hip arthroplasty using ceramic component due to reduce wear and osteolysis. CASE REPORT: 29 year old male case of bilateral hip avascular necrosis operated for bilateral hip THA presented 30 months after with complaint of squeaking, x-ray showing left ceramic liner fracture without any history of trauma. Revision of ceramic liner and head with ceramic on polyethylene was done. DISCUSSION ANDEntities:
Keywords: Ceramic on ceramic; Ceramic on polyethylene; Total hip arthroplasty
Year: 2019 PMID: 31734476 PMCID: PMC6864327 DOI: 10.1016/j.ijscr.2019.10.079
Source DB: PubMed Journal: Int J Surg Case Rep ISSN: 2210-2612
Fig. 1Showing postoperative AP radiograph a normal seated acetabular component and liner.
Fig. 2Shows a left side ceramic liner fracture.
Fig. 3Intraoperative photographs shows Fractured Ceramic liner.
Fig. 4Radiograph after revision of ceramic Head and liner.