Literature DB >> 31723725

Accuracy and Reliability of Infrared Analyzers for Measuring Human Milk Macronutrients in a Milk Bank Setting.

Maryanne T Perrin1, Jannette Festival2, Shaina Starks3, Lesley Mondeaux4, Elizabeth A Brownell5,6, Amy Vickers3.   

Abstract

BACKGROUND: Infrared (IR) analysis is an emerging technology that may be a useful tool for milk banks to manage the nutrient variability in donor human milk.
OBJECTIVE: To evaluate the accuracy, reliability, and comparability of commercial infrared analyzers for measuring human milk macronutrients in a milk bank setting.
METHODS: Three nonprofit milk banks received blinded test kits of human milk that had been assessed using reference methods. Four infrared instruments were used to measure macronutrients as follows: 1 filtered mid-IR, 2 Fourier-transformed full-spectra mid-IR, and 1 near-IR. Twenty-five unique samples were read concurrently for the accuracy arm. An identical sample was read daily for 1 mo for the reliability arm.
RESULTS: Values for R 2 describing relationships with reference methods for total fat, crude protein, and lactose, were as follows: filtered mid-IR, 0.98, 0.94, and 0.48; Fourier-transformed full-spectra mid-IR, 0.97, 0.93, and 0.36 for instrument 1 and 0.98, 0.98, and 0.31 for instrument 2; and near-IR 0.93, 0.93, and 0.12. There was no significant difference between instruments for crude protein and total fat measurements. There were significant differences in carbohydrate measurements between instruments. For 1 mo of daily measurements in the reliability arm, CVs for filtered mid-IR were ≤4.6%, for Fourier-transformed full spectra mid-IR were ≤1.7%, and for near-IR were ≤5.1%.
CONCLUSIONS: Infrared analysis is an accurate and reliable method for measuring crude protein and total fat in a milk bank setting. Carbohydrate measurements are less accurate and are significantly different between instruments, which will likely lead to differences in derived calorie values.
Copyright © The Author(s) 2019.

Entities:  

Keywords:  donor milk; human milk; infrared; macronutrients; milk bank

Year:  2019        PMID: 31723725      PMCID: PMC6838652          DOI: 10.1093/cdn/nzz116

Source DB:  PubMed          Journal:  Curr Dev Nutr        ISSN: 2475-2991


Introduction

There is significant variability in the macronutrient content of human milk between women at the same stage of lactation, with fat often reported as the most variable macronutrient (1–3). While macronutrient variability is of limited concern with ad libitum feeding of term infants, it may have a significant impact on preterm infants who are fed enterally at prescribed rates and are at risk for poor growth (4). The American Academy of Pediatrics and the Canadian Pediatric Society recommend that preterm infants who do not have access to their mother's own milk receive pasteurized donor human milk (DHM) from a milk bank due to its protective effect against necrotizing enterocolitis (NEC) (5–7). Milk banks screen donors serologically and for lifestyle risk factors, pool milk between multiple donors, pasteurize milk, and conduct postpasteurization microbiological tests before dispensing DHM to neonatal intensive care units (NICUs) (8). Three small studies suggest that the fat content of DHM can also be highly variable (3, 9, 10), which may contribute to the poor growth observed in preterm infants fed DHM (7, 11, 12). Knowing the macronutrient content of a donor's milk could improve the ability of milk banks to strategically combine donors to reach nutrient goals for DHM going to vulnerable infants in the NICU. Infrared (IR) analyzers are tools that have been used in the dairy industry to measure the macronutrient content of bovine milk (13). Some milk banks within the Human Milk Banking Association of North America (HMBANA) have adapted IR analyzers for use with human milk. Recently, the US FDA approved the first IR human milk macronutrient test system as a medical device (14). HMBANA milk banks operate as food manufacturers, a role that differs operationally and regulatorily from clinical entities that operate medical devices. There is a growing body of research that suggests that, with appropriate sample handling, IR analyzers are accurate and reliable for measuring fat and protein in human milk (15–23); however, the majority of these studies were conducted in laboratory or clinical settings. Therefore, the purpose of this study was to 1) evaluate the accuracy of commercially available IR analyzers for measuring macronutrients in DHM in a milk bank setting; 2) compare results between a variety of IR technologies used at multiple milk banks; and 3) evaluate the reliability of commercially available IR analyzers for measuring macronutrients over 30 d in a milk bank setting.

Methods

This method validation study was classified as non–human subject research and provided an exempt status by the Institutional Review Board at the University of North Carolina Greensboro (UNCG). It was also approved by the HMBANA Research Committee.

Test kit preparation

Human milk test kits were prepared in the laboratory at the University of North Carolina Greensboro (UNCG). De-identified raw human milk used to prepare the kits was received from 3 HMBANA milk banks (NorthernStar Mothers Milk Bank, Calgary, Alberta; Northwest Mothers Milk Bank, Portland, OR; and Mothers' Milk Bank North Texas, Fort Worth, TX) whose donors consent to having their milk used for research purposes. The milk was from 12 unique, approved donors and had been classified as research milk due to various deferral requirements (e.g., donor use of medications). Test kit samples were prepared by thawing milk in a warming oven and shaking water baths, and pooling milk from 1 to 2 donors to reach volumes ranging from 200 to 1800 mL, with higher volumes needed for samples used in the reliability arm. Samples that contained milk from 2 donors used different combinations of donor milks, with the goal of creating a wide range of nutrients within the test kits. Each sample was continuously stirred with a magnet during the process of preparing test kit aliquots to ensure consistency. The test kits consisted of 10 unique calibration samples, with macronutrient values provided to the milk bank for instrument calibration; 25 unique, blinded samples for the accuracy arm of the study; and 30 identical, blinded samples for the reliability arm of the study. All test kits contained the same samples for calibration and the accuracy arm. Because IR instruments use up the sample in the process of analysis and samples cannot be remeasured, large sample volumes were required to create 30 d of identical samples for the reliability arm; therefore, samples for the reliability arm of the study were identical within a milk bank, but different between milk banks. Each sample in the test kit contained sufficient volume to be measured in duplicate. Completed test kits were frozen and shipped overnight on dry ice to participating milk banks and to reference labs. Due to volume limitations, only a subset of the 25 unique test kit samples in the accuracy arm were analyzed by reference methods: crude protein and nonprotein nitrogen (n = 20), total fat (n = 20), and lactose (n = 25).

Milk bank setting 1

The NorthernStar Mothers Milk Bank (Calgary, Alberta, Canada) has been operating since 2012 and processes approximately 200,000 oz (5,914 L) of DHM annually. Two human milk test kits were sent to the NorthernStar milk bank for testing on a near-IR instrument (Near) (Unity SpectraStar XL, Unity Scientific) and on a filtered mid-IR instrument (FiltMid) (Miris HMA Human Milk Analyzer, Miris). Sample handling processes are summarized in .
TABLE 1

Summary of calibration and sample handling by instrument

FiltMidFTMid1FTMid2Near
OperatorTrained milk bank technicianTrained milk bank technicianTrained milk bank technicianTrained milk bank technician
Calibration setManufacturer calibration samplesHuman milk samples (10) analyzed by USDAStudy calibration samples (10)Study calibration samples (10)
Volume required for duplicate readings6 mL50 mL60 mL6 mL
ThawingOvernight in refrigeratorOvernight in refrigeratorOvernight in refrigeratorOvernight in refrigerator
WarmingTo 40°C in a bead warmer (heater model JBN5 US, Miris)Temperatures at analysis 4°–8°C. Pilot study on effect of temperature showed no difference in results at starting temperatures of 4°C, 24°C, and 40°C (unpublished data).To 20°C in water bath (Precision 183, Fisher Scientific)To 40°C in bead warmer (heater model JBN5 US, Miris)
MixingPer manufacturer instructionsSonicated at 75% amplitude 2 min with 6-mm probe submerged ∼75% into sample (VCX 130, Sonics and Materials)Sonicated at 75% amplitude 3 min with 6-mm probe submerged ∼75% into sample (VCX 130, Sonics and Materials)Sonicated at 50% amplitude for 15 s (ultrasonic processor model CV18, Miris); sample gently inverted 10 times before loading

FiltMid (Miris HMA), FTMid1 (MilkoScan), FTMid2, (LactoScope), and Near (SpectraStar). FiltMid, filtered midinfrared; FTMid, Fourier-transformed full-spectra midinfrared; Near, near infrared.

Summary of calibration and sample handling by instrument FiltMid (Miris HMA), FTMid1 (MilkoScan), FTMid2, (LactoScope), and Near (SpectraStar). FiltMid, filtered midinfrared; FTMid, Fourier-transformed full-spectra midinfrared; Near, near infrared.

Milk bank setting 2

The Mothers' Milk Bank of North Texas has been operating since September 2004 and processes approximately 600,000 oz (17,742 L) of DHM annually. One human milk test kit was sent to the North Texas milk bank for testing on a Fourier-transformed full spectra mid-IR instrument (FTMid1) (MilkoScan FT120, FOSS). Due to loss of study calibration samples, calibration was done with in-house calibration samples created by the North Texas milk bank and analyzed by the USDA Federal Milk Market. Sample handling processes are summarized in Table 1.

Milk bank setting 3

Northwest Mothers Milk Bank has been operating since 2013 and processes approximately 375,000 oz (11,089 L) of DHM annually. One human milk test kit was sent to the Northwest milk bank for testing on a Fourier-transformed full spectra mid-IR instrument (FTMid2) (LactoScope FTA; Perten Instruments). Additionally, Perten Instruments set up a custom channel on the LactoScope to estimate lactose using the calibration curve data, as IR technology does not distinguish between lactose and indigestible carbohydrates. Sample handling processes are summarized in Table 1.

Measurement

Reference methods were used to measure the macronutrients in the test kits as follows. Total nitrogen and nonprotein nitrogen (NPN) were measured by Eurofins using the Kjeldahl method (AOAC 991.20 and AOAC 991.21, respectively) to assess the protein in human milk. Total nitrogen was converted to Crude Protein using a protein conversion factor for milk foods of 6.38 (24, 25). True protein was calculated for reference samples as (crude protein − measured NPN). Total fat was measured by the USDA Federal Milk Market using the Mojonnier ether extraction method. Lactose was measured in triplicate at UNCG using AOAC method 2006.06 adapted to run in a 96-well plate. This method uses the enzymes β-galactosidase, galactose mutarose, and β-galactosidase dehydrogenase (K-LACGAR, Megazyme International) to convert lactose to β-d-galactose, which oxides NAD+ to NADH. This enzymatic approach has been reported as accurate with HPLC methods (26). Average CVs for lactose measurements in triplicate were 3.2%. It is important to note that IR technology, which was originally developed for the dairy industry, measures crude protein, which does not account for NPN (e.g., urea and nucleotides). In bovine milk, the amount of NPN is limited (<5%), while in human milk, NPN averages 20–30% and can be >50% (17, 27–29). Similarly, IR technology measures total carbohydrates and does not differentiate between lactose and human milk oligosaccharides (HMOs). Bovine milk contains limited indigestible carbohydrates while in human milk, HMOs can account for 15–25% of carbohydrates (30). In the context of analyzing human milk in a food manufacturing setting, we chose to measure lactose for our reference values instead of total carbohydrates because metabolizable nutrients are recommended for deriving calorie values in foods (24).

Data analysis

Accuracy for IR instruments was defined based on information published in the Miris HMA Human Milk Analyzer User Manual since this device has been approved for clinical use by the FDA: total fat ±12% of reference methods; crude protein, true protein, and carbohydrates ± 15% of reference methods (31). Coefficients of determination (R2) were calculated to assess relationships with reference methods and analyzer results were plotted as percentage differences from reference methods. Descriptive statistics on macronutrient measurements were computed by analyzer type and differences between analyzers were evaluated using an ANOVA with a Tukey's honest significant difference test for multiple comparisons. Intra-assay CVs were computed to assess the reliability of instruments over 1 mo of replicate measures. Regression analysis was used to probe for significant changes over time. All analysis was conducted with SAS Enterprise Edition 9.4 (SAS Corporation).

Results

Assessment of accuracy to reference methods

According to the reference methods used, the study samples contained the following ranges of macronutrients: total fat 2.52–4.11 g/dL (n = 20), crude protein 1.01–1.96 g/dL (n = 20), true protein 0.70–1.57 g/dL (n = 20), and lactose 5.09–6.79 g/dL(n = 25). Scatter plots of percentage differences between IR measurements and reference methods are presented in . There was a high degree of accuracy (±15%) for crude protein measurements across all analyzers, with a small number of samples falling slightly outside the range with the filtered mid-IR (3/20, 15%) and near-IR instruments (1/20, 5%). None of the IR instruments measure true protein. Instead, true protein values were derived via a calculation by reducing crude protein by a flat percentage. Accurate true protein values could be obtained by applying a 30% deduction to crude protein on the Fourier-transformed full-spectra mid-IR instruments and the near-IR instrument, while a 20% deduction was appropriate on the filtered mid-IR instrument. Similarly, there was a high degree of accuracy (±12%) for total fat measurements across all analyzers. Only the Fourier-transformed full-spectra mid-IR instruments reported carbohydrate values that fell within ±15% of reference values for lactose, and there was a trend to overreporting at lower lactose concentrations. The filtered mid-IR and near-IR instruments reported carbohydrate values that were 20–50% higher than reference lactose values, with greater differences at lower lactose concentrations.
FIGURE 1

Accuracy of infrared instruments compared with the reference methods FiltMid (Miris HMA), FTMid1 (MilkoScan), FTMid2, (LactoScope), and Near (SpectraStar). FiltMid, filtered midinfrared; FTMid, Fourier-transformed full-spectra midinfrared; Near, near infrared.

Accuracy of infrared instruments compared with the reference methods FiltMid (Miris HMA), FTMid1 (MilkoScan), FTMid2, (LactoScope), and Near (SpectraStar). FiltMid, filtered midinfrared; FTMid, Fourier-transformed full-spectra midinfrared; Near, near infrared. Relationship with reference methods, reported as R2 values from regression analysis for total fat, crude protein, and lactose, were as follows: FiltMid, 0.91, 0.94, and 0.48; FTMid1, 0.97, 0.93, and 0.36, FTMid2, 0.98, 0.98, and 0.31, and Near, 0.93, 0.93, and 0.12. One sample measured multiple times on FiltMid was >30% below reference values for fat, suggesting the fat loss may have occurred in kit preparation. When this sample was excluded, the R2 value for fat was 0.98 on FiltMid.

Comparison of results between analyzers

We did not observe a significant difference between the crude protein or total fat value obtained with different infrared instruments (P > 0.05). There was a significant difference in the carbohydrate measurements obtained between infrared instruments (P < 0.0001) ().
TABLE 2

Comparison of macronutrient results by infrared analyzer

FiltMid (n = 25)FTMid1 (n = 24)FTMid2 (n = 25)Near (n = 24) P value
Crude protein, g/dL1.29 ± 0.32 (0.90–2.00)1.43 ± 0.28 (1.03–2.00)1.42 ± 0.30 (1.01–2.04)1.40 ± 0.26 (1.07–1.96)0.3254
Fat, g/dL3.34 ± 0.65 (1.85–4.65)3.44 ± 0.58 (2.55–4.72)3.46 ± 0.49 (2.71–4.60)3.44 ± 0.55 (2.49–4.80)0.8771
Carbohydrate, g/dL7.78a ± 0.17 (7.40–8.10)6.38b ± 0.22 (5.89–6.72)5.88c ± 0.37 (5.06–6.55)7.90a ± 0.16 (7.62–8.17)<0.0001

Values represent mean ± SD (range) unless otherwise indicated. Data were evaluated with ANOVA and Tukey's honest significant difference for multiple comparisons. Values in the same row with the same superscript are not significantly different. FiltMid (Miris HMA), FTMid1 (MilkoScan), FTMid2, (LactoScope), and Near (SpectraStar). FiltMid, filtered midinfrared; FTMid, Fourier-transformed full-spectra midinfrared; Near, near infrared.

Comparison of macronutrient results by infrared analyzer Values represent mean ± SD (range) unless otherwise indicated. Data were evaluated with ANOVA and Tukey's honest significant difference for multiple comparisons. Values in the same row with the same superscript are not significantly different. FiltMid (Miris HMA), FTMid1 (MilkoScan), FTMid2, (LactoScope), and Near (SpectraStar). FiltMid, filtered midinfrared; FTMid, Fourier-transformed full-spectra midinfrared; Near, near infrared.

Assessment of reliability over 1 mo

Macronutrients measured daily over 1 mo are plotted by IR instrument in . The only significant (P < 0.05) time-related trends that we observed were a decline in crude protein in the Fourier-transformed full-spectra mid-IR instrument (LactoScope) (slope = −0.0008 g/d; −0.06% daily change from reference value) and a decline in lactose values in the Near instrument (slope = −0.0147 g/day; −0.22% daily change from reference value). These temporal changes were very small, resulting in approximately 2–6% total change over the course of 30 d. The intra-assay CVs for 1 mo of daily measurements for crude protein, fat, and carbohydrates, respectively, were as follows: FiltMid, 4.2%, 4.6%, and 0.7%; FTMid1, 1.2%, 0.4%, and 0.3%; FTMid2, 0.9%, 1.7%, and 0.8%; Near, 5.1%, 3.0%, and 2.2%.
FIGURE 2

Reliability of macronutrient measurements over 1 mo of daily use, by infrared instrument. Each milk bank received a different set of test samples, so macronutrient values between analyzers were not expected to be the same. FiltMid (Miris HMA), FTMid1 (MilkoScan), FTMid2, (LactoScope), and Near (SpectraStar). FiltMid, filtered midinfrared; FTMid, Fourier-transformed full-spectra midinfrared; Near, near infrared; ProCrude, crude protein.

Reliability of macronutrient measurements over 1 mo of daily use, by infrared instrument. Each milk bank received a different set of test samples, so macronutrient values between analyzers were not expected to be the same. FiltMid (Miris HMA), FTMid1 (MilkoScan), FTMid2, (LactoScope), and Near (SpectraStar). FiltMid, filtered midinfrared; FTMid, Fourier-transformed full-spectra midinfrared; Near, near infrared; ProCrude, crude protein.

Discussion

We observed high levels of accuracy in measuring total fat and crude protein in raw human milk across 3 different milk banks using a variety of commercially available infrared technologies, including devices developed specifically for human milk (Miris HMA), and devices developed for the dairy industry (MilkoScan, LactoScope, SpectraStar) and adapted by milk banks to measure human milk. Our findings are in agreement with others that have reported accuracy in measuring total fat and crude protein using near-IR (16, 18, 21), filtered mid-IR (15, 18–20, 23), and full-spectra mid-IR instruments (17, 32), when devices have been calibrated to human milk. Limited studies have been conducted in a milk bank setting (19, 23, 32, 33), and few studies have compared multiple types of infrared technology (18). IR analyzers can only estimate true protein values using crude protein measurements. The average nonprotein nitrogen composition of our reference samples was 28% (range 20–34%). We found that a 20% reduction in crude protein was appropriate for the filtered mid-IR instrument and that a 30% reduction was needed for the other instruments, suggesting that conversion factors to obtain accurate true protein values may differ by instrument. In contrast to our findings, a few studies have described poor accuracy when using IR technology to measure crude protein and fat in human milk (34–36). Existing research using infrared analyzers for measuring macronutrients in human milk differs with respect to sample thawing, warming, and homogenization methods, as well as instrument calibration, which may explain differing conclusions (37). We observed significant differences in the accuracy and comparability of carbohydrate measurements across devices. The Fourier-transformed full-spectra mid-IR devices mostly fell into ±15% of the reference values for lactose, while the near-IR and filtered mid-IR devices reported values that were 20–50% higher than reference lactose values. This is likely because, while IR technology measures total carbohydrates and cannot differentiate between lactose and HMOs, the full-spectra instrument calibration equations developed by the milk banks attempts to derive lactose values from total carbohydrate measurements based on known lactose values of the calibration samples. None of the instruments in our study showed strong agreement with reference methods for lactose (R2 < 0.5). Others have reported low accuracy in measuring lactose in human milk with infrared analysis (15, 17, 18, 32, 34). During approximately 1 month of daily use, we found all IR devices provided consistent results, with small, but statistically significant, temporal changes in some components (in the range of 0.06% to 0.22%); however, these values are not likely to be clinically significant. The small temporal changes we did observe highlight the importance of monitoring instruments daily and recalibrating as needed to identify and respond to potential instrument drifts. In the current regulatory environment, milk banks operate as food manufacturers, and thus they are subject to the Code of Federal Regulation, Title 21, Food and Drugs (38). Among other things, this regulation requires milk banks to follow Good Manufacturing Practices (part 110) and to maintain clean and accurate equipment (part 110.40). Our study provides evidence that a variety of infrared analyzers can be used in a milk bank setting to accurately measure crude protein and total fat. Use of these instruments may enable milk banks to strategically pool milk from different donors to reach protein and fat targets for DHM. Significant differences in the measurements and accuracy of lactose between devices may lead to differences in calorie calculations, which is an important area for future research given the nutritional vulnerability of infants receiving DHM.

Limitations

We used a small number of reference samples that did not include more extreme concentrations of fat and protein. This is an important area for future research, including testing cream products that have a higher percentage of total solids, which may impact IR accuracy. We defined accuracy based on thresholds for an instrument that has been approved as a clinical device by the FDA. Different levels of accuracy may be more appropriate for manufacturers of medical foods. There was one outlier for total fat (>30% below reference value) in the filtered mid-IR instrument; this sample was measured multiple times with similar results, suggesting that the error may have occurred during preparation of the test kits, highlighting how important drawing a representative sample is for generating meaningful results. We did not measure total carbohydrates, as we felt lactose was a better indicator of the energy available to an infant; therefore, instruments that measure total carbohydrates may have shown better performance results if they and been assessed against total carbohydrate reference values.

Conclusions

A variety of infrared instruments, including near-IR, filtered mid-IR, and full-spectra mid-IR, can accurately measure crude protein and total fat in a milk bank setting when instruments have been calibrated to human milk and when appropriate sample handling protocols are used. Infrared instruments may be useful tools to help milk banks determine how to combine milk from multiple donors to achieve target fat and protein amounts. Calculations to arrive at true protein measurements may differ between instruments. There are significant differences in carbohydrate measurements between instruments that may lead to differences in calorie values. Research is needed regarding the accuracy of IR devices to derive calorie values for human milk.
  31 in total

1.  Evaluation of a mid-infrared analyzer for the determination of the macronutrient composition of human milk.

Authors:  Ylenia S Casadio; Tracey M Williams; Ching T Lai; Sofia E Olsson; Anna R Hepworth; Peter E Hartmann
Journal:  J Hum Lact       Date:  2010-09-24       Impact factor: 2.219

Review 2.  "Bed Side" Human Milk Analysis in the Neonatal Intensive Care Unit: A Systematic Review.

Authors:  Gerhard Fusch; Celia Kwan; Gynter Kotrri; Christoph Fusch
Journal:  Clin Perinatol       Date:  2016-12-29       Impact factor: 3.430

3.  A longitudinal study of the protein, nitrogen, and lactose contents of human milk from Swedish well-nourished mothers.

Authors:  B Lönnerdal; E Forsum; L Hambraeus
Journal:  Am J Clin Nutr       Date:  1976-10       Impact factor: 7.045

4.  Calibration Adjustment of the Mid-infrared Analyzer for an Accurate Determination of the Macronutrient Composition of Human Milk.

Authors:  Hélène Billard; Laure Simon; Emmanuelle Desnots; Agnès Sochard; Cécile Boscher; Alain Riaublanc; Marie-Cécile Alexandre-Gouabau; Clair-Yves Boquien
Journal:  J Hum Lact       Date:  2015-06-02       Impact factor: 2.219

5.  Validation of mid-infrared spectroscopy for macronutrient analysis of human milk.

Authors:  S Parat; S Groh-Wargo; S Merlino; C Wijers; D M Super
Journal:  J Perinatol       Date:  2017-04-20       Impact factor: 2.521

6.  Comparison of mid-infrared transmission spectroscopy with biochemical methods for the determination of macronutrients in human milk.

Authors:  Dolores Silvestre; Miriam Fraga; María Gormaz; Ester Torres; Máximo Vento
Journal:  Matern Child Nutr       Date:  2012-07-12       Impact factor: 3.092

7.  A longitudinal study of human milk composition in the second year postpartum: implications for human milk banking.

Authors:  Maryanne T Perrin; April D Fogleman; David S Newburg; Jonathan C Allen
Journal:  Matern Child Nutr       Date:  2016-01-18       Impact factor: 3.092

8.  Changes in carbohydrate composition in human milk over 4 months of lactation.

Authors:  G V Coppa; O Gabrielli; P Pierani; C Catassi; A Carlucci; P L Giorgi
Journal:  Pediatrics       Date:  1993-03       Impact factor: 7.124

9.  Growth in VLBW infants fed predominantly fortified maternal and donor human milk diets: a retrospective cohort study.

Authors:  Tarah T Colaizy; Susan Carlson; Audrey F Saftlas; Frank H Morriss
Journal:  BMC Pediatr       Date:  2012-08-17       Impact factor: 2.125

10.  Rapid measurement of macronutrients in breast milk: How reliable are infrared milk analyzers?

Authors:  Gerhard Fusch; Niels Rochow; Arum Choi; Stephanie Fusch; Susanna Poeschl; Adelaide Obianuju Ubah; Sau-Young Lee; Preeya Raja; Christoph Fusch
Journal:  Clin Nutr       Date:  2014-05-17       Impact factor: 7.324

View more
  2 in total

1.  The Nutritional Composition and Energy Content of Donor Human Milk: A Systematic Review.

Authors:  Maryanne T Perrin; Mandy B Belfort; James I Hagadorn; Jacqueline M McGrath; Sarah N Taylor; Lauren M Tosi; Elizabeth A Brownell
Journal:  Adv Nutr       Date:  2020-07-01       Impact factor: 8.701

Review 2.  Using Nature to Nurture: Breast Milk Analysis and Fortification to Improve Growth and Neurodevelopmental Outcomes in Preterm Infants.

Authors:  Katherine Marie Ottolini; Elizabeth Vinson Schulz; Catherine Limperopoulos; Nickie Andescavage
Journal:  Nutrients       Date:  2021-11-29       Impact factor: 5.717

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.