A robotic flexible endoscope with shared autonomy: a study of mockup cholecystectomy.

Chengzhi Song1, Xin Ma2, Xianfeng Xia1, Philip Wai Yan Chiu1,2, Charing Ching Ning Chong1, Zheng Li3,4.   

Abstract

BACKGROUND: Endoscope is the eye of surgeon in minimally invasive surgery (MIS). Prevailing handheld endoscopes are manually steered, which can cause endoscope-instrument fencing. Robotic endoscopes can reduce the fatigue but could not reduce collisions. Handheld endoscopes with a flexible bending tip can reduce the shaft pivoting and collisions. However, its steering is challenging. In this paper, we present a robotic flexible endoscope with auto-tracking function and compare it with the conventional rigid endoscopes.
METHODS: A robotic flexible endoscope (RFE) with shared autonomy is developed. The RFE could either track the instruments automatically or be controlled by a foot pedal. A mockup cholecystectomy was designed to evaluate the performance. Five surgeons were invited to perform the mockup cholecystectomy in an abdominal cavity phantom with a manual rigid endoscope (MRE), a robotic rigid endoscope (RRE), and the RFE. Space occupation, time consumption, and questionnaires based on the NASA task load index were adopted to evaluate the performances and compare the three endoscope systems. An ex vivo experiment was conducted to demonstrate the feasibility of using the RFE in a biological tissue environment.
RESULTS: All surgeons completed the mockup cholecystectomy with the RFE independently. Failure occurred in the cases involving the RRE and the MRE. Inside the body cavity, the space occupied when using the RFE is 17.28% and 23.95% (p < 0.05) of that when using the MRE and the RRE, respectively. Outside the body cavity, the space occupied when using the RFE is 14.60% and 15.53% (p < 0.05) of that by using MRE and RRE. Time consumed in the operations with MRE, RRE, and RFE are 28.3 s, 93.2 s and 34.8 s, respectively. Questionnaires reveal that the performance of the RFE is the best among the three endoscope systems.
CONCLUSIONS: The RFE provides a wider field of view (FOV) and occupies less space than rigid endoscopes.

Entities:  

Keywords:  Cholecystectomy; Minimally invasive surgery; Robotic flexible endoscope; Shared autonomy

Mesh:

Year:  2019        PMID: 31722046     DOI: 10.1007/s00464-019-07241-8

Source DB:  PubMed          Journal:  Surg Endosc        ISSN: 0930-2794            Impact factor:   4.584


  1 in total

1.  Robotic Camera Assistance and Its Benefit in 1033 Traditional Laparoscopic Procedures: Prospective Clinical Trial Using a Joystick-guided Camera Holder.

Authors:  Sebastian W Holländer; Hans Joachim Klingen; Marliese Fritz; Peter Djalali; Dieter Birk
Journal:  Surg Technol Int       Date:  2014-11
  1 in total
  3 in total

1.  Robotic hepatic resection in postero-superior region of liver.

Authors:  Zhiming Zhao; Zhuzeng Yin; Lichao Pan; Chenggang Li; Minggen Hu; Wan Yee Lau; Rong Liu
Journal:  Updates Surg       Date:  2020-10-08

2.  Autonomous robotic laparoscopic surgery for intestinal anastomosis.

Authors:  H Saeidi; J D Opfermann; M Kam; S Wei; S Leonard; M H Hsieh; J U Kang; A Krieger
Journal:  Sci Robot       Date:  2022-01-26

3.  A Confidence-Based Supervised-Autonomous Control Strategy for Robotic Vaginal Cuff Closure.

Authors:  Michael Kam; Hamed Saeidi; Michael H Hsieh; J U Kang; Axel Krieger
Journal:  IEEE Int Conf Robot Autom       Date:  2021-10-18
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.