Literature DB >> 31702087

Assessment of the Adaptation of Interim Crowns using Different Measurement Techniques.

Chin-Chun Peng1, Kwok-Hung Chung1, Van Ramos1.   

Abstract

PURPOSE: To compare the internal fit and marginal discrepancy of acrylic resin interim crowns fabricated by different manufacturing methods, and to test the consistency of measuring marginal discrepancy and internal fit between different measuring techniques.
MATERIALS AND METHODS: A dentoform mandibular left first molar was prepared for an all-ceramic crown. Thirty-six interim crowns were fabricated and divided into three groups (n = 12): group BAC (Bis-acrylic composite, fabricated manually), group CAM (CAD/CAM polymethylmethacrylate resin, milled), and group 3DP (3D printed methacrylic oligomers, printed). The internal fit of the interim crowns was evaluated by the silicone replica technique and by X-ray microcomputed tomography (µCT) technique. The marginal discrepancy of the interim crowns was evaluated by the vinyl polysiloxane (VPS) (Aquasil Ultra XLV) impression technique and by optical coherence tomography (OCT) technique. Data were statistically analyzed using ANOVA and Turkey tests at α = 0.05. Pearson correlation test was used to evaluate the correlation between the different measurement techniques and marginal discrepancy/internal fit.
RESULTS: The manually fabricated interim crowns (group BAC) had significantly greater discrepancy of internal fit than did the digitally fabricated crowns (group CAM and group 3DP) measured by both silicone replica technique and µCT 2-dimensional (2D) image measurement. There were no statistically significant differences in the cement space volume values obtained by the µCT image technique between group BAC and group 3DP (p = 0.285). The coefficient of determination between the two volumetric measurement techniques was low (R2 = 0.30). For marginal discrepancy, the manually fabricated interim crowns had a wider absolute marginal discrepancy than both digitally fabricated groups (p < 0.05). In both the VPS impression and OCT assessment, there was no statistically significant difference between group CAM and group 3DP (p = 0.798 and 0.994, respectively). The coefficient of determination between the VPS impression and OCT techniques for marginal discrepancy measurement was low (R2 = 0.23).
CONCLUSIONS: Digitally fabricated interim crowns (group CAM and group 3DP) had better internal fit and smaller marginal discrepancy than manually fabricated interim crowns (group BAC). For comparison of the different evaluation techniques, the silicone replica technique and µCT measurements had low correlation for internal fit assessment, as did the PVS impression and OCT techniques for marginal discrepancy test.
© 2019 by the American College of Prosthodontists.

Entities:  

Keywords:  Internal fit; X-ray microcomputed tomography technique; marginal discrepancy; optical coherence tomography technique; silicone replica technique

Year:  2019        PMID: 31702087     DOI: 10.1111/jopr.13122

Source DB:  PubMed          Journal:  J Prosthodont        ISSN: 1059-941X            Impact factor:   2.752


  4 in total

Review 1.  3D printing restorative materials using a stereolithographic technique: a systematic review.

Authors:  Alvaro Della Bona; Viviane Cantelli; Vitor T Britto; Kaue F Collares; Jeffrey W Stansbury
Journal:  Dent Mater       Date:  2021-01-19       Impact factor: 5.304

2.  Evaluation of Internal Fit and Marginal Adaptation of Provisional Crowns Fabricated with Three Different Techniques.

Authors:  Jie Wu; Hongjun Xie; Alireza Sadr; Kwok-Hung Chung
Journal:  Sensors (Basel)       Date:  2021-01-22       Impact factor: 3.576

3.  Marginal and internal fit and intaglio surface trueness of interim crowns fabricated from tooth preparation of four finish line locations.

Authors:  Keunbada Son; Young-Tak Son; Ji-Min Lee; Kyu-Bok Lee
Journal:  Sci Rep       Date:  2021-07-06       Impact factor: 4.379

4.  Effect of Luting Cement and Convergence Angle of the Preparation on the Internal Fit of Zirconia Restorations.

Authors:  Andrés Sánchez-Monescillo; Carlos González-Serrano; José González-Serrano; João Malta Barbosa; Carlos López-Suárez; Sillas Duarte
Journal:  Materials (Basel)       Date:  2021-12-18       Impact factor: 3.623

  4 in total