Particle size study of nine metered dose inhalers, and their deposition probabilities in the airways.

A Bouchikhi1, M H Becquemin, J Bignon, M Roy, A Teillac.   

Abstract

This study deals with the particle size measurement of nine aerosol metered dose inhalers. Calibration was made possible by the use of a laser particle velocimeter (aerodynamic Particle Sizer from TSI). The count median aerodynamic diameters (CMAD) show little variation, from 0.63 to 0.73 micron, with standard deviations (sigma g) between 1.2 and 1.8. Aerodynamic diameter aerosol diagram analysis showed multimodal mass distribution for all the tested dose inhalers. Calculations for the airway deposition probabilities (extrathoracic, tracheobronchial and alveolar) refer to the studies made by W. Stahlhofen and co-workers. As most aerosol metered dose inhalers have a predominantly bronchial therapeutic destination, the deposition at the bronchial level could be enhanced with the following parameters: inspired volume of 1500 ml, inspiratory time of 2 sec, aerosol mass median aerodynamic diameter (MMAD) of 7.5 microns, with a monodispersed distribution. The respective influences of the excipients and propellents used for the aerosolization of these dose metered inhalers are also discussed.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 3169224

Source DB:  PubMed          Journal:  Eur Respir J        ISSN: 0903-1936            Impact factor:   16.671


  5 in total

Review 1.  Non-impactor-based methods for sizing of aerosols emitted from orally inhaled and nasal drug products (OINDPs).

Authors:  Jolyon Mitchell; Richard Bauer; Svetlana Lyapustina; Terrence Tougas; Volker Glaab
Journal:  AAPS PharmSciTech       Date:  2011-07-22       Impact factor: 3.246

2.  Bioequivalence of generic aerosol bronchodilators: what are the issues?

Authors:  M Spino
Journal:  CMAJ       Date:  1989-11-01       Impact factor: 8.262

3.  Effect of terbutaline on mucociliary clearance in asthmatic and healthy subjects after inhalation from a pressurised inhaler and a dry powder inhaler.

Authors:  J Mortensen; S Groth; P Lange; F Hermansen
Journal:  Thorax       Date:  1991-11       Impact factor: 9.139

4.  The Effect of Electronic Cigarette User Modifications and E-liquid Adulteration on the Particle Size Profile of an Aerosolized Product.

Authors:  Haley A Mulder; Jesse L Patterson; Matthew S Halquist; Leon Kosmider; Joseph B McGee Turner; Justin L Poklis; Alphonse Poklis; Michelle R Peace
Journal:  Sci Rep       Date:  2019-07-15       Impact factor: 4.379

5.  Orally Inhaled Drug Particle Transport in Computerized Models of Laryngotracheal Stenosis.

Authors:  Dennis Onyeka Frank-Ito; Seth Morris Cohen
Journal:  Otolaryngol Head Neck Surg       Date:  2020-10-13       Impact factor: 3.497

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.