Evaluation of scattered radiation from fluoroscopy using small OSL dosimeters.

Hajime Ito1, Ikuo Kobayashi2,3, Kazutoshi Watanabe2, Shigehiro Ochi2, Noriyuki Yanagawa2,4.   

Abstract

Recently, there has been a significant amount of interest in studying the importance of radiation doses to the eye lens during endoscopic retrograde cholangiopancreatography (ERCP). A study that focused on measuring the scattered radiation using an ionization chamber survey meter reported that a lead curtain was useful in reducing the scattered radiation. The over-couch X-ray tube system tends to deliver higher doses to the head and neck of the staff involved in the procedure than the under-couch X-ray tube position. In this study, a small optically stimulated luminescence (OSL) dosimeter called the nanoDot was used to evaluate and measure the amount of radiation; this dosimeter was developed by Landauer Ltd. and was specifically designed for point measurements. There are numerous studies that have reported the usefulness of personal OSL dosimeters other than the nanoDot to measure scattered radiation. Here, we evaluated the amount of scattered radiation, along with the degree of reduction achieved with the use of a protective curtain, while employing a personal OSL dosimeter and nanoDot. When the scattered radiation dose was measured using the nanoDot, the maximum recorded value without a protective curtain was 0.363 mGy and that with a protective curtain was 0.026 mGy, both at the height of 100 cm. The maximum reduction rate of scattered radiation while using a protective curtain was approximately 93% and 97% at 100 cm and 150 cm, respectively. The measured values recorded using both personal OSL dosimeters and nanoDot machine were strongly correlated.

Keywords:  Dosimetry; Endoscopic retrograde cholangiopancreatography; Fluoroscopy; Radiation protection; Scattered radiation; Small OSL dosimeter

Mesh:

Year:  2019        PMID: 31691889     DOI: 10.1007/s12194-019-00536-4

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  7 in total

1.  [Measurement of scatter radiation on MDCT equipment using an OSL dosimeter].

Authors:  Hironobu Tomita; Kunihiko Morozumi
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2004-11

2.  [Creation of a crystalline lens radiation exposure defense cover and the effect of radiation exposure decrease on neuro-interventions].

Authors:  Toshio Take; Kaori Sato; Katsunori Kiuchi; Yasuo Nakazawa
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2007-11-20

3.  Angular dependence of the nanoDot OSL dosimeter.

Authors:  James R Kerns; Stephen F Kry; Narayan Sahoo; David S Followill; Geoffrey S Ibbott
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

Review 4.  ICRP publication 118: ICRP statement on tissue reactions and early and late effects of radiation in normal tissues and organs--threshold doses for tissue reactions in a radiation protection context.

Authors:  F A Stewart; A V Akleyev; M Hauer-Jensen; J H Hendry; N J Kleiman; T J Macvittie; B M Aleman; A B Edgar; K Mabuchi; C R Muirhead; R E Shore; W H Wallace
Journal:  Ann ICRP       Date:  2012-02

5.  Occupational radiation dose to eyes from endoscopic retrograde cholangiopancreatography procedures in light of the revised eye lens dose limit from the International Commission on Radiological Protection.

Authors:  U O'Connor; A Gallagher; L Malone; G O'Reilly
Journal:  Br J Radiol       Date:  2013-02       Impact factor: 3.039

6.  A head phantom study for intraocular dose evaluation of 64-slice multidetector CT examination in patients with suspected cranial trauma.

Authors:  Kosuke Matsubara; Kichiro Koshida; Kimiya Noto; Tadanori Takata; Masayuki Suzuki; Tetsunori Shimono; Tomoyuki Yamamoto; Osamu Matsui
Journal:  Eur J Radiol       Date:  2010-03-02       Impact factor: 3.528

7.  Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurements.

Authors:  Paul A Jursinic
Journal:  Med Phys       Date:  2007-12       Impact factor: 4.071

  7 in total
  2 in total

1.  A more accurate and safer method for the measurement of scattered radiation in X-ray examination rooms.

Authors:  Tokiko Nakamura; Suzuki Shoichi; Yasutaka Takei; Masanao Kobayashi; Vergil Cruz; Ikuo Kobayashi; Satoshi Asegawa; Kyoichi Kato
Journal:  Radiol Phys Technol       Date:  2019-12-23

2.  Occupational eye dose correlation with neck dose and patient-related quantities in interventional cardiology procedures.

Authors:  Hiroki Ishii; Koichi Chida; Ko Satsurai; Yoshihiro Haga; Yuji Kaga; Mitsuya Abe; Yohei Inaba; Masayuki Zuguchi
Journal:  Radiol Phys Technol       Date:  2022-01-23
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.