Shozo Arai1, Hironao Okada2, Hiroshi Sawada1, Yuji Takahashi1, Kumiko Kimura1, Toshihiro Itoh3. 1. National Institute of Animal Health, National Agriculture and Food Research Organization (NARO), 3-1-5 Kannondai, Tsukuba, Ibaraki 305-0856, Japan. 2. Sensing System Research Center, National Institute of Advanced Industrial Science and Technology (AIST), 1-2-1 Namiki, Tsukuba, Ibaraki 305-8564, Japan. 3. Graduate School of Frontier Sciences, The University of Tokyo, 5-1-5 Kashiwanoha, Kashiwa, Chiba 277-8563, Japan.
Abstract
We evaluated the relationship between ruminal motility measured by a force transducer and acceleration measured by bolus sensor, and we assessed the detection of ruminal motility in cattle by a bolus-type wireless sensor. The bolus sensor can be orally administered to cattle and was placed in the reticulum for continuous measurements. The probe was almost horizontal to the longitudinal axis. The bolus sensor's basic y-axis acceleration movement appeared to have a very distinct vertical pattern, occurring roughly 1-1.5 times/min with a duration of approximately 8 sec, displaying at around 500 mG. A significant positive correlation was observed between the ruminal contraction revealed by the force transducer and the acceleration shown by the bolus sensor (P<0.01). The contraction of the dorsal sac of the rumen and the acceleration signals in the reticulum occurred at practically the same time. The frequency and amplitude of ruminal contraction demonstrated by the bolus sensor and the force transducer in feeding were significantly higher than those at rest (P<0.01). The bolus sensor could also detect ruminal atony in the cattle after the administration of xylazine. A bolus-type wireless sensor may thus be useful for the measurement of ruminal motility in cattle and for detecting rumen dysfunction (e.g., ruminal atony).
We evaluated the relationship between ruminal motility measured by a force transducer and acceleration measured by bolus sensor, and we assessed the detection of ruminal motility in cattle by a bolus-type wireless sensor. The bolus sensor can be orally administered to cattle and was placed in the reticulum for continuous measurements. The probe was almost horizontal to the longitudinal axis. The bolus sensor's basic y-axis acceleration movement appeared to have a very distinct vertical pattern, occurring roughly 1-1.5 times/min with a duration of approximately 8 sec, displaying at around 500 mG. A significant positive correlation was observed between the ruminal contraction revealed by the force transducer and the acceleration shown by the bolus sensor (P<0.01). The contraction of the dorsal sac of the rumen and the acceleration signals in the reticulum occurred at practically the same time. The frequency and amplitude of ruminal contraction demonstrated by the bolus sensor and the force transducer in feeding were significantly higher than those at rest (P<0.01). The bolus sensor could also detect ruminal atony in the cattle after the administration of xylazine. A bolus-type wireless sensor may thus be useful for the measurement of ruminal motility in cattle and for detecting rumen dysfunction (e.g., ruminal atony).
Entities:
Keywords:
bolus sensor; cattle; force transducer; ruminal motility; wireless
The frequency and amplitude of ruminal contractions in cattle are influenced by metabolic
diseases such as ruminal acidosis and hypocalcaemia, as well as any of the many diseases that
cause pain or fever [2, 10, 17]. In addition to anorexia, clinical
signs of indigestion include reticuloruminal hypo-motility [11]. Reticuloruminal acidosis has shown to reduce the frequency of ruminal
contraction, eventually resulting in ruminal stasis [2,
18]. Ruminal contractions decrease in number and
strength and reticuloruminal stasis then develops, often within 15−30 min. Reticuloruminal
stasis is of particular interest in relation to the pathogenesis of production diseases
associated with gastrointestinal atony, such as displaced abomasum and ruminal tympany [2]. Therefore, the measurement of ruminal motility in its
own right has value in the management of cattle to improve their health, welfare and
production performance.Radiotelemetry has been used in research for many decades to monitor numerous physiological
parameters such as the respiratory rate, heart rate, and body temperature in animals [6]. Wireless sensors for intraruminal insertion through the
esophagus (boluses) were recently developed as a noninvasive alternative to surgery. Various
measurement systems have become available for the continuous monitoring of the ruminal and
reticular pH or temperature values in cattle [1, 14, 19, 21, 23]. These bolus
sensors are useful for detecting subacute ruminal acidosis or fever in cattle. However, it is
unknown whether a bolus sensor can carry out continuous long-term measurements of
reticuloruminal motility in cattle. Ruminal motility in cattle can be assessed physiologically
using devices such as a strain gauge force transducer or by ultrasonography [4, 7]. A force
transducer method requires surgery, and it is difficult to carry out continuous
ultrasonography measurements over a long term. More recently, bolus wireless sensors with an
accelerometer were developed, and several research groups have attempted to assess the
function of reticulorumen such as rumination or the motion of reticulorumen in cattle [3, 15, 20]. It has been unclear whether a bolus wireless sensor
with an accelerometer can assess the ruminal motility in cattle. If ruminal motility can be
measured over a long term, the measurement could be useful for detecting dysfunctions of
ruminal motility such as ruminal atony, ruminal tympany, and anorexia. We conducted the
present study to examine the relationship between ruminal motility in cattle measured by a
force transducer and acceleration measured by a bolus sensor, and we assessed the ability of a
bolus-type wireless sensor to detect ruminal motility in cattle.
MATERIALS AND METHODS
Animals
Four healthy non-lactating Holstein cows weighing 420 to 595 kg were used in this study.
All cattle were housed in individual stanchion stalls, and were given commercial cattle
concentrate, hay and free access to water. Food was offered twice daily (9:00 and 16:00).
The handling of the animals used in this study was approved by the Institutional Care and
Use Committee for Laboratory Animals of the National Institute of Animal Health
(Japan).
The bolus-type wireless sensor and system
The configuration of the wireless sensor and system were modified based on our methods
[3, 20]. We
designed a wireless sensor probe that can be administered to cattle orally. The probe is
cylinder-shaped and consists of a three-axis accelerometer (ADXL362, Analog Devices,
Chelmsford, MA, U.S.A.), a temperature sensor, a microcontroller unit (C8051F930, Silicon
Labs, Austin, TX, U.S.A.), a transmitter (Si4461-B1, Silicon Labs), a radiofrequency (RF)
antenna, and a battery. The probe housing was made from polytetrafluorethylene (PTFE) with
glue and PTFE seal tape. The diameter of the probe is 22 mm, its length is 76 mm, and its
weight is 55 g (its density is 1.9 g/cm3) (Fig. 1A).
Fig. 1.
A: Bolus sensor and radio-telemetry system. B: System operation. A1: Bolus sensor.
A2: Collar repeater. A3: Receiver.
A: Bolus sensor and radio-telemetry system. B: System operation. A1: Bolus sensor.
A2: Collar repeater. A3: Receiver.The probe can be easily administered orally via a catheter into the rumen. Power is
provided by a heavy-duty 3 V lithium battery. The acceleration is probed at 1 sec interval
and in a ± 2 G range with 10-bit resolution. Transmissions are in the 429-MHz band. This frequency band does not require any license in Japan, but
the maximum effective radiated power is restricted to 12.14 dBm. The battery has a mean
life of 1 month when measurements of three-axis acceleration were transmitted continuously
every 1 sec. This system uses a collar repeater to relay the transmissions to a receiver
that stores the data on a cloud-connected server or a personal computer with special
software (Monitoring data system ver.4, GSI Corp., Tokyo, Japan) (Fig. 1B). The communication distance from the collar repeater of
the cattle to the receiver was about 60 m. The bolus sensor is a disposable device. The
collar repeater can be reused by changing the battery.
Measurement of acceleration by the bolus sensor
The acceleration data from the bolus sensor are transmitted to the receiver in real time.
Three-axis acceleration data were recorded to a personal computer with the GSI software
cited above connected to the receiver. The sampling rates for acceleration was 1 Hz. In
three-axis acceleration, the x- and z-axes indicate the radial acceleration of the probe,
and the y-axis indicates vertical acceleration. The average values of the acceleration
(mG) and the peak number of acceleration (times/min) using the bolus sensor were
calculated each 10 min. The y-axis data were used as the measurement value of
acceleration.
Measurement of ruminal motility by the force transducer
Ruminal motility in the cattle was measured by a strain gauge force transducers [4]. The size of the force transducer was 18 × 35 mm, and
the force transducer contained the strain gauge (Foil Strain Gauge; NEC San-ei
Instruments, Ltd., Tokyo, Japan). The force transducer was covered with silicone. Four
cattle were sedated intramuscularly with xylazine hydrochloride (0.05 mg/kg; Bayer,
Leverkusen, Germany). Ten min later, local anesthesia was applied to the left paralumbar
fossa with an intramuscular infusion of procaine hydrochloride (8−100
ml/body, Kyoritsu Seiyaku Co., Tokyo, Japan). The force transducer was
sutured onto the serosa of the dorsal sac of the rumen. The lead wires of the force
transducer were exteriorized through the abdominal wall. The wires from the force
transducer were then subcutaneously tunneled and pulled out through the skin. The cattle
were injected intramuscularly with antibiotics (10 mg/kg, ampicillin sodium; Meiji Seika
Pharma Co., Ltd., Tokyo, Japan) after the surgery.Ruminal motility was recorded on a polygraph (MP100A, BIOPAC Systems Inc., Santa Barbara,
CA, U.S.A.) by connecting the lead wires of the transducers to the connecting cables from
amplifiers (DA100, BIOPAC Systems Inc.), was digitized by an analog-to-digital converter,
and continuously recorded on a computer for analysis. The amplitude of ruminal contraction
was calculated as the average value of the measurements of the area (g × sec) surrounded
by the contraction wave at the base-line for 10 min. The frequency of ruminal contraction
(contractions/min) was calculated as the average value for 10 min.
Experimental procedure
We first examined the correlation between ruminal motility shown by the force transducer
and the acceleration shown by the bolus sensor. The bolus sensor was orally administered
by using a custom-made gastric tube, without surgery, and the bolus sensor was confirmed
to be located in the reticulum by the movement pattern of acceleration [3]. The force transducer was sutured onto the dorsal sac
of the rumen, with surgery. The Measurements by the force transducer and bolus sensor were
carried out starting at 7 days after the surgery. The ruminal motility measured by the
force transducer and the acceleration measured by the bolus sensor were recorded
continuously and simultaneously throughout the trial. The measurement data of ruminal
motility and acceleration were collected for 30-min periods during feeding and resting for
each cattle. We also repeated the same trial on another day. The y-axis data were used as
the measurement values of acceleration. We were able to confirm the cattle’s behavior as
feeding or resting by video observations through a camera connected to a notebook
computer; these videos were recorded.We next examined the effects of the administration of xylazine hydrochloride on the
ruminal motility and acceleration. The cattle were intramuscularly administered xylazine
hydrochloride (0.05 mg/kg; Bayer) once at 12:00. The measurement data of ruminal motility
and acceleration were collected for 120-min periods before and after xylazine
hydrochloride administration. The y-axis data were used as the measurement values of
acceleration.
Statistical analysis
Data are expressed as the mean ± standard deviation (SD). The relationship between
acceleration shown by the bolus sensor and the ruminal motility shown by the force
transducer was examined by obtaining Pearson’s correlation coefficient and by performing
an analysis of regression. Student’s t-test was used to evaluate
differences in acceleration, the amplitude of concentration, the peak number of
acceleration, and the frequency of contraction between feeding and resting. Student’s
t-test was also used to evaluate differences in acceleration shown by
the bolus sensor and the ruminal motility shown by the force transducer between before
(−10 min) and after the administration of xylazine. A value of P<0.05
was considered as statistically significant.
RESULTS
Acceleration data revealed by the bolus sensor
Figure 2 illustrates the three-axis acceleration data from the reticulum for a 30-min period
when the cattle was resting. The baseline y-axis data stayed level, but baseline data of
the x- and z-axes followed a large curve over the day. The probe was in an almost
horizontal position with respect to the longitudinal axis. The basic movement of
acceleration of the y-axis appeared to have a very distinct vertical pattern (Figs. 2 and
3), occurring roughly 1 time/min, with a duration of approximately 8 sec, and
displaying two local maximums at around 500 mG (Fig.
3). The acceleration of the x- and z-axes showed an indistinct vertical pattern
(Fig. 2).
Fig. 2.
Changes in the acceleration data of the x-, y-, and z-axes for a 30-min period when
the cattle was at rest. The x- and z-axes indicate the acceleration in the radial
direction. The y-axis indicates the acceleration in the vertical direction.
Fig. 3.
Measurement results of ruminal motility obtained with the force transducer and the
acceleration data obtained with the bolus sensor. The acceleration data are the
y-axis.
Changes in the acceleration data of the x-, y-, and z-axes for a 30-min period when
the cattle was at rest. The x- and z-axes indicate the acceleration in the radial
direction. The y-axis indicates the acceleration in the vertical direction.Measurement results of ruminal motility obtained with the force transducer and the
acceleration data obtained with the bolus sensor. The acceleration data are the
y-axis.
Relationship between ruminal motility and acceleration data
Figure 3 shows the measurement results of
ruminal motility obtained with the force transducer and the acceleration data (y-axis)
obtained with the bolus sensor. The waveforms of ruminal motility and that of acceleration
revealed by the bolus sensor occurred at nearly the same time (Fig. 3), and patterns were clearly visible. The acceleration shown
by the bolus sensor was significantly positively correlated with the amplitude of ruminal
contraction shown by the force transducer (P<0.01, r=0.966) (Fig. 4). The peak number of accelerations obtained with the bolus sensor was significantly
positively correlated with the frequency of ruminal contraction obtained with the force
transducer (P<0.01, r=0.961) (Fig.
4). The acceleration data used the y-axis values.
Fig. 4.
A: The relationship between acceleration and the amplitude of ruminal contraction.
B: The relationship between the peak number of acceleration and the frequency of
ruminal contraction. Significant correlations were observed between the two methods
(A: r=0.966, P<0.01, B: r=0.961,
P<0.01).
A: The relationship between acceleration and the amplitude of ruminal contraction.
B: The relationship between the peak number of acceleration and the frequency of
ruminal contraction. Significant correlations were observed between the two methods
(A: r=0.966, P<0.01, B: r=0.961,
P<0.01).
Ruminal motility and acceleration data during feeding and resting
The acceleration data shown by the bolus sensor during feeding (464 ± 168 mG) was
significantly higher than that shown during resting (P<0.01) (385 ±
109 mG) (Table 1). The amplitude of ruminal contraction measured by the force transducer
during feeding (473 ± 127 g × sec) was significantly higher than that during resting
(P<0.01) (414 ± 88 g × sec) (Table 1). The peak number of accelerations during feeding (1.5 ± 0.1 times/min)
was significantly higher than that during resting (P<0.01) (1.1 ± 0.1
times/min) (Table 1). The frequency of ruminal
contraction during feeding (1.5 ± 0.1 contractions/min) was significantly higher than that
during resting (P<0.01) (1.1 ± 0.1 contractions /min) (Table 1). The acceleration data used the y-axis
values.
Table 1.
Measurement values of the bolus sensor and the force transducer during feeding
and resting in cattle
Item
Bolus sensor
Force transducer
Feeding
Resting
Feeding
Resting
Acceleration (mG)
464 ± 168a)
385 ± 109b)
-
-
Amplitude of contraction (area: g × sec)
-
-
473 ± 127a)
414 ± 88b)
Peak number of acceleration (times/min)
1.5 ± 0.1a)
1.1 ± 0.1b)
-
-
Frequency of contraction (contractions/min)
-
-
1.5 ± 0.1a)
1.1 ± 0.1b)
Data are expressed as means ± SD . a, b) P<0.01.
Data are expressed as means ± SD . a, b) P<0.01.
Changes in ruminal motility and acceleration data after the administration of
xylazine
After the administration of xylazine hydrochloride, the changes in the waveform of
ruminal motility shown by the force transducer and those of acceleration shown by the
bolus sensor were almost the same (Fig. 5). The ruminal atony occurred approximately 10 min after xylazine hydrochloride
administration (Fig. 5). Compared to the −10 min
values (before the xylazine hydrochloride administration), the acceleration and the
amplitude of ruminal contraction were significantly decreased at 10−40 min after the
xylazine hydrochloride administration (P<0.05) and were then
significantly increased at 80 min (P<0.05) (Fig. 6). Compared to the −10 min values, the peak number of accelerations and the
frequency of ruminal contractions were significantly decreased at 10−70 min after the
administration of xylazine hydrochloride (P<0.05) (Fig. 6). Regarding the cattle’s clinical signs,
ruminal tympany occurred at 30−40 min after the administration of xylazine hydrochloride
and then recovered at 60−70 min.
Fig. 5.
Changes in ruminal motility and acceleration after the administration of xylazine
hydrochloride in one of the four cattle. Top: Acceleration shown by the bolus
sensor. Bottom: Ruminal motility shown by the force transducer.
Fig. 6.
A: Changes in the amplitude of contraction revealed by the force transducer and
acceleration revealed by the bolus sensor after the administration of xylazine
hydrochloride in cattle. B: Changes in the frequency of contraction revealed by the
force transducer and the peak number of acceleration revealed by the bolus sensor
after the administration of xylazine hydrochloride in cattle.
*P<0.05 compared to −10 min (before the administration of
xylazine). ■, Force transducer; □, Bolus sensor.
Changes in ruminal motility and acceleration after the administration of xylazine
hydrochloride in one of the four cattle. Top: Acceleration shown by the bolus
sensor. Bottom: Ruminal motility shown by the force transducer.A: Changes in the amplitude of contraction revealed by the force transducer and
acceleration revealed by the bolus sensor after the administration of xylazine
hydrochloride in cattle. B: Changes in the frequency of contraction revealed by the
force transducer and the peak number of acceleration revealed by the bolus sensor
after the administration of xylazine hydrochloride in cattle.
*P<0.05 compared to −10 min (before the administration of
xylazine). ■, Force transducer; □, Bolus sensor.
DISCUSSION
The intended location of the bolus sensor in this study was the reticulum, which is a small
compartment at the front of the rumen. The bolus-type wireless sensor that was inserted into
the rumen was left continuously in the reticulum [5,
16, 22]. It
seems that the bolus sensor is held in a comparatively free situation in digesta of the
reticulum [16]. Since the accelerometer is subject to
Earth’s gravity, it is possible to monitor its orientation. We observed that the baseline
y-axis acceleration data of the bolus sensor was kept level, whereas the baseline data of x-
and z-axes showed a large curve. Once in its intended location in the reticulum, the probe
is in an almost horizontal position with respect to the longitudinal axis. Acceleration
measurements may be influenced by factors such as the flow of the contents of the
reticulorumen and reticuloruminal motion [3, 15]. In the present study, there were many periods during
which relatively clear data dominated the signal. The cylindrical shape of the probe allows
it to rotate freely around its longitudinal axis, and this is reflected in the collected
data (x- and z-axes). Probe rotations occur with varying frequency and tend to be
concomitant with reticuloruminal motion. A bolus sensor can be confirmed to be located in
the reticulum by palpation through the rumen cannula [3, 22]. It is also possible to determine
the location of the probe from measurement data alone [3]. When inserted in the rumen, the probe will eventually (within a few days)
spontaneously transfer from the rumen to the reticulum, where it remains indefinitely. There
appears to be a discernible difference in the movement pattern of acceleration between the
rumen and the reticulum. The pattern in the reticulum exhibits more feature-rich movements,
where each movement generally displays two or three local maximums, whereas the pattern in
the rumen displays smoother movements with fewer maximums and a slightly different rhythm
[3]. Our present analyses revealed more feature-rich
movements of acceleration with two local maximums within 1 day after the oral administration
of the bolus sensor in all four cattle.We observed that the waveforms of ruminal motility and acceleration were almost the same,
and a significant positive correlation was observed between the amplitude of ruminal
contraction and the acceleration (r=0.966, P<0.01). A significant
positive correlation was also revealed between the frequency of ruminal contraction and the
peak number of acceleration (r=0.961, P<0.01). Thus, the contraction of
the dorsal sac of the rumen and the acceleration signals in the reticulum occurred at nearly
the same time. This phenomenon may be explained as follows: the flow of ruminal contents
which occurred by compression of the dorsal sac of the rumen pushed and moved the bolus
sensor in the reticulum. Braun et al. reported that the duration time of
the contraction of the dorsal sac of the rumen was 8.2 ± 1.04 sec [7]. In the present study, the duration of the acceleration movement shown
by the bolus sensor was about 8 sec. It seemed that the movement of the dorsal sac of the
rumen and the movement of the acceleration of digesta in the reticulum were almost the same.
The general pattern of reticuloruminal motility consists of regular contraction sequences
[8, 24], that
successively engage the various parts of the reticulum and rumen. There are primary and
secondary contraction cycles; the former serves primarily to mix the ingesta, support
optimal microbial fermentation, and transport ingesta into the omasum, whereas the latter
cycles are involved in the eructation of gas [8, 9]. The primary cycles started with biphasic reticular
contraction followed immediately by contraction of the ruminal atrium. The dorsal sac of the
rumen, the caudodorsal blind sac and the left longitudinal groove contracted at practically
the same time, after which contraction of the caudoventral blind sac, and the ventral sac of
the rumen occurred. Secondary contraction cycles do not involve the reticulum or ruminal
atrium; they consist of contraction of the dorsal sac and then the ventral sac of the rumen
[7, 9]. In the
present study, the movement of acceleration shown by the bolus sensor generally displayed
two local maximus. Since the bolus sensor was located in the reticulum, these local maximums
may be reflecting biphasic reticular contraction. It is likely that the major waveform of
acceleration shown by the bolus sensor indicates the flow of ruminal contents which occurred
by compression of the dorsal sac of the rumen, because a significant positive correlation
was observed between the acceleration and ruminal motility.It was reported that the frequency of contraction of the dorsal sac of the rumen was 1.1 ±
0.2 contractions/min during resting [7]. Herein we
observed similar results obtained by the force transducer (1.1 ± 0.1 contractions/min) and
the bolus sensor (1.1 ± 0.1 times/min). The frequency and amplitude of the ruminal
contractions during feeding were significantly higher than those during resting. Similar
results were obtained by another measurement methods [12]. Ruminal motility is also detectable by auscultation. Ruminal sounds can be
heard with a stethoscope placed on the sublumbar triangle. Ruminal sounds are generated when
a cow is fed a diet with high fiber content that moves along the ruminal wall during
contraction of the rumen. High producing dairy cows fed a ration with relatively low fiber
content produce fewer frictional ruminal sounds than cows fed hay [8, 17]. Because acceleration data
measured by a bolus sensor indicate the flow of ruminal contents, the acceleration data may
be reflecting ruminal sounds.Our analyses demonstrated that ruminal atony occurred approximately 10 min after an
administration of xylazine hydrochloride. The detection of ruminal atony by the bolus sensor
was almost synchronous with that detected by the force transducer. Xylazine, an
α2-adrenoreceptor agonist, is often used as a sedative for cattle. The
pharmacologic effects of xylazine are an excessive depression of the central nervous system,
ruminal atony and bradycardia [25]. In disorders of
ruminal motility, general anesthesia and disease at any site that produces pain or fever can
inhibit the hindbrain reflex centers responsible for evoking primary and secondary cycle
contractions of the reticulorumen [17]. Clinical
cases of ruminal stasis should be interpreted as being caused by a depression of gastric
centers, a lack of excitatory reflex inputs, an increase in inhibitory reflex inputs,
blockade of the motor pathways as in hypocalcemic animals but probably not in the case of
vagus indigestion. In addition to their specific reflex inputs, the gastric centres are
influenced by nervous activity in other parts of the central nervous system [17].The bolus senor used in the present study could detect impaired forestomach motility such
as ruminal atony in real time. In addition, ruminal tympany occurred when the ruminal atony
continued for 30−40 min. Ruminal tympany (bloat) is of particular significance in fattening
beef cattle, but it may also be a problem in dairy herds on high concentrate rations [13]. In healthy cattle, gaseous distension of the
reticulorumen reflex elicits an increase in the force and frequency of secondary cycles; it
evokes the opening of the cardia and initiates the other respiratory, esophageal, pharyngeal
and buccal actions of the eructation sequence. Normally the cardiac region is cleared when
the dorsal ruminal sac contractions of secondary cycles force the gas layer forward, thereby
depressing the fluid level below the level of the cardia. In cases of systemic disease in
which ruminal hypomotility or stasis are secondary features, a persistent, mild bloat may be
evident, because although the fermentation rate is probably low, the weak ruminal motility
may be inadequate to move the gas layer and to clear the cardia [17]. We observed that the bolus sensor was useful for evaluating aspects
of ruminal motility such as the frequency, amplitude, and timing of hypomotility.
Evaluations of ruminal motility using a bolus sensor may be useful as a non-invasive method
of investigating reticuloruminal stasis in cows with ruminal tympany, acidosis, fever or
hypocalcaemia [2, 10, 17].Since the bolus sensor is located continuously in the reticulum, the sensor probe should be
durable for as long as possible. Our probe housing was made from PTFE, and the sensor could
be recorded continuously for ≥1−2 months. Several studies have described probe housing made
from stainless steel or resin, and the studies’ authors noted that the sensors used with
this type of probe housing could be recorded continuously for several months [14, 23]. It has
been unclear whether a sensor probe can have even more long-term (e.g., several years)
durability and safety. Further investigations are needed to assess the durability and safety
of the bolus sensor described herein, in order to spread the uses of sensor technology for
cattle.In conclusion, the bolus sensor we have described can be orally administered without
surgery, and it was left in the reticulum. The basic movement of acceleration of the y-axis
shown by the bolus sensor appeared to have a very distinct vertical pattern. For the
analyses of ruminal motility, the acceleration of the y-axis should be used. A significant
positive correlation was observed between the ruminal contraction measured by the force
transducer and the acceleration measured by the bolus sensor. The contraction of the dorsal
sac of the rumen and the acceleration signals in the reticulum occurred at practically the
same time. The bolus sensor also detected ruminal atony after the administration of
xylazine. We therefore suggest that a bolus-type wireless sensor may be useful for the
measurement of ruminal motility and for detecting dysfunctions of the rumen such as ruminal
atony and ruminal tympany.
Authors: Andrew W Hamilton; Chris Davison; Christos Tachtatzis; Ivan Andonovic; Craig Michie; Holly J Ferguson; Laura Somerville; Nicholas N Jonsson Journal: Sensors (Basel) Date: 2019-03-07 Impact factor: 3.576
Authors: Mahmoud Fadul; Luigi D'Andrea; Maher Alsaaod; Giuliano Borriello; Antonio Di Lori; Dimitri Stucki; Paolo Ciaramella; Adrian Steiner; Jacopo Guccione Journal: PLoS One Date: 2022-03-04 Impact factor: 3.240