| Literature DB >> 31677354 |
Manu Verma1, Karen Sooy2, George Just2, Mark Nixon1, Ruth Morgan1, Ruth Andrew1, Karen E Chapman1, Natalie Z M Homer1,2.
Abstract
RATIONALE: The activity of the glucocorticoid activating enzyme 11β-hydroxysteroid dehydrogenase type-1 (11βHSD1) is altered in diseases such as obesity, inflammation and psychiatric disorders. In rodents 11βHSD1 converts inert 11-dehydrocorticosterone (11-DHC) into the active form, corticosterone (CORT). A sensitive, specific liquid chromatography/tandem mass spectrometry method was sought to simultaneously quantify total 11-DHC and total and free CORT in murine plasma for simple assessment of 11βHSD1 activity in murine models.Entities:
Mesh:
Substances:
Year: 2020 PMID: 31677354 PMCID: PMC7540072 DOI: 10.1002/rcm.8610
Source DB: PubMed Journal: Rapid Commun Mass Spectrom ISSN: 0951-4198 Impact factor: 2.586
Instrument settings for CORT, 11‐DHC and internal standards
| Chromatographic conditions (flow rate 0.5 mL/min) | ||
|---|---|---|
| Time (min) | Mobile phase A (%): water (0.1% FA, v/v) | Mobile phase B (%): acetonitrile (0.1% FA, v/v) |
| 0 | 70 | 30 |
| 1.5 | 70 | 30 |
| 3 | 10 | 90 |
| 4 | 10 | 90 |
| 6 | 70 | 30 |
| 10 | 70 | 30 |
Conditions established following electrospray ionisation at 5 kV, 550°C.
FA = formic acid; DP = declustering potential (V); CE = collision energy (eV); CXP = collision cell exit potential (V).
Quantitative ion with qualitative ion shown in brackets.
Figure 1Extracted ion chromatograms of 11‐DHC, CORT, epi‐CORT and d4F demonstrating chromatographic resolution of isobaric epimers CORT and epi‐CORT and of 11‐DHC and CORT
Recovery and indices of intra‐ and inter‐day precision and accuracy for quantitation of mouse plasma enriched at LOQ (0.25 and 0.20 ng/mL), low (2.5), medium (10) and high (150) levels of 11‐DHC and CORT
| Amount (ng/mL) | % recovery | Intra‐day | Inter‐day | ||
|---|---|---|---|---|---|
| Mean ± RSD | Precision, RSD (%) | Accuracy, RME (%) | Precision, RSD (%) | Accuracy, RME (%) | |
| 0.25 (11‐DHC) 0.20 (CORT) | 87.2 ± 3.4 89.1 ± 2.0 | 10.13 9.80 | 15.32 17.30 | 12.92 18.37 | 12.39 10.92 |
| 2.5 (11‐DHC) 2.5 (CORT) | 89.5 ± 2.1 86.1 ± 3.1 | 4.19 4.70 | 3.94 8.11 | 14.74 11.21 | 10.27 8.45 |
| 10 (11‐DHC) 10 (CORT) | 90.2 ± 5.4 86.3 ± 2.4 | 4.31 5.22 | 8.23 7.32 | 3.97 4.23 | 10.20 2.50 |
| 150 (11‐DHC) 150 (CORT) | 89.5 ± 3.4 87.3 ± 2.4 | 6.23 7.90 | 1.20 2.50 | 9.44 12.30 | 5.56 3.50 |
RSD = relative standard deviation; RME = relative mean error.
Figure 2Concentrations of corticosterone (CORT) in mouse plasma, 3 h following ip injection of vehicle (0.9% saline) or 100 μg/kg lipopolysaccharide (LPS). Measured amounts were within the validated range. Data are mean ± SEM (n = 7/group)