BACKGROUND: Subpectoral biceps tenodesis can be performed with cortical fixation using different repair techniques. The goal of this technique is to obtain a strong and stable reduction of biceps tendon in an anatomic position. PURPOSE/HYPOTHESIS: The purpose of this study was to compare (1) displacement during cyclic loading, (2) ultimate load, (3) construct stiffness, and (4) failure mode of the biceps tenodesis fixation methods using onlay techniques with an all-suture anchor versus an intramedullary unicortical button. It was hypothesized that fixation with all-suture anchors using a Krackow stitch would exhibit biomechanical characteristics similar to those exhibited by fixation with unicortical buttons. STUDY DESIGN: Controlled laboratory study. METHODS: Ten pairs of fresh-frozen cadaveric shoulders (N = 20) were dissected to the humerus, leaving the biceps tendon-muscle unit intact for testing. A standardized subpectoral biceps cortical (onlay) tenodesis was performed using either an all-suture anchor or a unicortical button. The biceps tendon was initially cycled from 5 to 70 N at a frequency of 1.5 Hz. The force on the tendon was then returned to 5 N, and the tendon was pulled until ultimate failure of the construct. Displacement during cyclic loading, ultimate failure load, stiffness, and failure modes were assessed. RESULTS: Cyclic loading resulted in a mean displacement of 12.5 ± 2.5 mm for all-suture anchor fixation and 29.2 ± 9.4 mm for unicortical button fixation (P = .005). One all-suture anchor fixation and 2 unicortical button fixations failed during cyclic loading. The mean ultimate failure load was 170.4 ± 68.8 N for the all-suture anchor group and 125.4 ± 44.6 N for the unicortical button group (P = .074), with stiffness 59.3 ± 11.6 N/mm and 48.6 ± 6.8 N/mm (P = .091), respectively. For the unicortical button, failure occurred by suture tearing through tendon in 100% of the specimens. For the all-suture anchor, failure occurred by suture tearing through tendon in 56% and knot failure in 44% of the specimens. CONCLUSION: The all-suture anchor fixation using a Krackow stitch for subpectoral biceps tenodesis provided ultimate load and stiffness similar to unicortical button fixation using a nonlocking whipstitch. The all-suture anchor fixation technique was shown to be superior in terms of displacement during cyclic loading when compared with the unicortical button fixation technique. However, the results of this study help to show that the fixation method used on the humeral side is less implicative of the overall construct strength than stitch location and technique, as the biceps tendon tissue and stitch configuration seem to be the limiting factor in subpectoral onlay tenodesis techniques. CLINICAL RELEVANCE: All-suture anchors have a smaller diameter than traditional suture anchors, can be inserted through curved guides, and preserve humeral bone stock without compromising postoperative imaging. This study supports use of the all-suture anchor fixation technique for subpectoral biceps tenodesis, with high biomechanical fixation strength and low displacement, as an alternative to the subpectoral onlay biceps tenodesis technique.
BACKGROUND: Subpectoral biceps tenodesis can be performed with cortical fixation using different repair techniques. The goal of this technique is to obtain a strong and stable reduction of biceps tendon in an anatomic position. PURPOSE/HYPOTHESIS: The purpose of this study was to compare (1) displacement during cyclic loading, (2) ultimate load, (3) construct stiffness, and (4) failure mode of the biceps tenodesis fixation methods using onlay techniques with an all-suture anchor versus an intramedullary unicortical button. It was hypothesized that fixation with all-suture anchors using a Krackow stitch would exhibit biomechanical characteristics similar to those exhibited by fixation with unicortical buttons. STUDY DESIGN: Controlled laboratory study. METHODS: Ten pairs of fresh-frozen cadaveric shoulders (N = 20) were dissected to the humerus, leaving the biceps tendon-muscle unit intact for testing. A standardized subpectoral biceps cortical (onlay) tenodesis was performed using either an all-suture anchor or a unicortical button. The biceps tendon was initially cycled from 5 to 70 N at a frequency of 1.5 Hz. The force on the tendon was then returned to 5 N, and the tendon was pulled until ultimate failure of the construct. Displacement during cyclic loading, ultimate failure load, stiffness, and failure modes were assessed. RESULTS: Cyclic loading resulted in a mean displacement of 12.5 ± 2.5 mm for all-suture anchor fixation and 29.2 ± 9.4 mm for unicortical button fixation (P = .005). One all-suture anchor fixation and 2 unicortical button fixations failed during cyclic loading. The mean ultimate failure load was 170.4 ± 68.8 N for the all-suture anchor group and 125.4 ± 44.6 N for the unicortical button group (P = .074), with stiffness 59.3 ± 11.6 N/mm and 48.6 ± 6.8 N/mm (P = .091), respectively. For the unicortical button, failure occurred by suture tearing through tendon in 100% of the specimens. For the all-suture anchor, failure occurred by suture tearing through tendon in 56% and knot failure in 44% of the specimens. CONCLUSION: The all-suture anchor fixation using a Krackow stitch for subpectoral biceps tenodesis provided ultimate load and stiffness similar to unicortical button fixation using a nonlocking whipstitch. The all-suture anchor fixation technique was shown to be superior in terms of displacement during cyclic loading when compared with the unicortical button fixation technique. However, the results of this study help to show that the fixation method used on the humeral side is less implicative of the overall construct strength than stitch location and technique, as the biceps tendon tissue and stitch configuration seem to be the limiting factor in subpectoral onlay tenodesis techniques. CLINICAL RELEVANCE: All-suture anchors have a smaller diameter than traditional suture anchors, can be inserted through curved guides, and preserve humeral bone stock without compromising postoperative imaging. This study supports use of the all-suture anchor fixation technique for subpectoral biceps tenodesis, with high biomechanical fixation strength and low displacement, as an alternative to the subpectoral onlay biceps tenodesis technique.
The long head of the biceps is recognized as a common contributor to anterior shoulder
pain and is often associated with other shoulder pathologies, including SLAP (superior
labrum anterior and posterior) lesions, rotator cuff tears, and subacromial impingement.[13,22] Both tenotomy and tenodesis are effective in ameliorating pain associated with
the long head of the biceps tendon. However, decreased muscle function and cosmetic
defects are seen at a higher rate after tenotomy compared with tenodesis.[8,21,22,28,31] Lower reoperation rates are seen after subpectoral fixation when compared with
suprapectoral fixation, and it is believed that releasing the tendon from its sheath and
the bicipital groove relieves the patient of most associated pain.[16,22,25]There is no clear consensus on whether bone tunnel or cortical surface (onlay) healing
confers better outcomes. Clinical outcome studies comparing interference screw fixation
(intramedullary) and suture anchor fixation techniques (onlay) for subpectoral biceps
tenodesis found no significant difference in patient outcomes between the 2 techniques.[16,19] However, interference screws are found to be associated with various
complications, including humeral fractures at the drill hole, persistent pain, and
bioabsorbable screw reactions.[7,17,26] Subpectoral biceps tenodesis using an onlay fixation technique is a reasonable
alternative to mitigate these risks.Fixation using all-suture anchors in the shoulder has become more popular owing to
increased preservation of bone stock with unicortical drilling of only 1.8 mm, improved
postoperative imaging, easier revision surgery, potentially lower fracture risk, and
similar biomechanical properties in comparison with suture anchors for labral repairs.[15,32] While most studies have reported the use of all-suture anchor onlay techniques in
the glenoid, the all-suture anchor has only been presented as a fixation device for
subpectoral tenodesis with bicortical drilling and intramedullary tendon fixation.[4] The purpose of this study was to compare the (1) displacement during cyclic
loading, (2) ultimate load to failure, (3) construct stiffness, and (4) failure mode of
the biceps tenodesis fixation methods using the all-suture anchor and unicortical button
onlay techniques. It was hypothesized that fixation with all-suture anchors would
exhibit biomechanical characteristics similar to those exhibited by fixation with
unicortical buttons when used for an onlay biceps tenodesis.
Methods
Specimen Preparation
Ten total pairs (N = 20) of male fresh-frozen cadaveric shoulders (mean age, 58.8
years [range, 51-64 years], body mass index, 18-35 kg/m2) were used
in this study. All specimens were devoid of any history of shoulder injury or
surgery, osteoarthritis, degenerative joint disease, and osteoporosis. The
shoulders from each pair were randomly allocated into 2 groups: one using a
unicortical button fixation method for biceps tenodesis (n = 10) and the other
using an all-suture anchor fixation (n = 10). Randomization was used to minimize
the effects of anatomic differences between right and left shoulders. All
specimens were dissected of all soft tissue and muscle to the level of the
shoulder capsule, leaving only the biceps muscle and tendon and capsular
structures intact. The humerus was then disarticulated from the glenoid, and the
transverse humeral ligament was removed to release the biceps tendon from the
intertubercular groove and sectioned 40 mm below the upper border of the
pectoralis major insertion. The humerus was then inverted and the proximal
aspect was potted in a cylindrical mold with polymethyl methacrylate (Fricke
Dental International) to the level of the inferior border of the intertubercular
groove while orienting the longitudinal axis of the humerus parallel to the
longitudinal axis of the cylindrical mold. A saline spray was used throughout
preparation and testing to keep the biceps tendon tissue superficially
hydrated.
Surgical Technique
For all repairs, the biceps tenodesis site on the humerus was marked 50 mm below
the palpable entrance of the bicipital groove. This placement is approximately
20 mm distal to the proximal edge of the pectoralis major tendon, which is 10 mm
proximal to the musculotendinous junction (MTJ) of the long head of the biceps
(Figure 1).[9,12] The long head of the biceps tendon was cut 20 mm proximal to the MTJ and
all tendons were sutured as described in the following “Fixation Technique”
section 20 mm distally beginning 10 mm proximal to the MTJ (Figure 2), as this is the location of the
strongest fixation strength.[29]
Figure 1.
Entrance point of unicortical drilling (yellow dot), 50 mm below the
distal entrance of the bicipital groove (b)
approximately 20 mm below the proximal edge of the pectoralis major
tendon (Pec major) (a) to ensure the anatomic length
and tension of the biceps muscle.
Figure 2.
Final repair construct with whipstitched biceps tendon, beginning 10 mm
proximal to (A) and including 10 mm of the
musculotendinous junction (B).
Entrance point of unicortical drilling (yellow dot), 50 mm below the
distal entrance of the bicipital groove (b)
approximately 20 mm below the proximal edge of the pectoralis major
tendon (Pec major) (a) to ensure the anatomic length
and tension of the biceps muscle.Final repair construct with whipstitched biceps tendon, beginning 10 mm
proximal to (A) and including 10 mm of the
musculotendinous junction (B).
Unicortical Button Fixation Technique
The unicortical button used in the study was an implantable titanium suture
button of 2.6 × 12–mm size with 2 suture holes (BicepsButton; Arthrex). For
intramedullary unicortical fixation, a 3.2-mm hole was drilled 50 mm distal from
the entrance of the bicipital groove as previously described (Figure 3A).[3] A nonabsorbable high-strength suture loop (FiberLoop No. 2; Arthrex) was
then placed in the MTJ with an initial locking stitch and a nonlocking
whipstitch 20 mm proximally with a total of 8 throws. The free suture ends were
manually pretensioned to seat the sutures, and both strands were passed through
1 button hole, then reshuttled through the opposite button hole in the reverse
direction. The button was passed through the previously drilled hole in the
anterior cortex of the humeral shaft according to the manufacturer's guide
(Figure 3B). The
button was released in an intramedullary manner and the shuttled strands were
gently pulled, allowing the button to “flip” and be seated in contact with the
anterior cortex of the humeral shaft (Figure 3C). The suture ends were pulled
to tighten the biceps tendon against the bone. Subsequently, the shuttled
strands were passed through the tendon and tied to each other by use of a knot
pusher with an initial sliding Weston knot followed by 4 reverse half-hitches
(Figure 3D).
Figure 3.
Unicortical button placement illustration: (A) unicortical 3.2-mm
drilling of the anterior cortex of the humeral shaft; (B) insertion of
the button; (C) one suture strand was gently pulled, allowing the button
to “flip” and be seated to the anterior cortex; (D) final unicortical
button fixation with biceps tendon.
Unicortical button placement illustration: (A) unicortical 3.2-mm
drilling of the anterior cortex of the humeral shaft; (B) insertion of
the button; (C) one suture strand was gently pulled, allowing the button
to “flip” and be seated to the anterior cortex; (D) final unicortical
button fixation with biceps tendon.
All-Suture Anchor Fixation Technique
A modified all-suture anchor fixation technique was performed using a
single-loaded all-suture soft anchor (1.8-mm FiberTak; Arthrex) with a braided
high-strength suture (No. 2 FiberWire CL; Arthrex).[10] For unicortical fixation, a drill guide was placed 50 mm distal from the
entrance of the bicipital groove and a 1.8-mm hole was drilled (Figure 4A). The all-suture
anchor was inserted into the bone tunnel through the drill guide (Figure 4B) and seated by
hand before it was impacted into its final position. The suture was then gently
pulled to expand the anchor and seat it securely against the anterior cortex of
the humeral shaft (Figure 4C). One suture limb was then used to make a locking Krackow stitch[11] about the proximal biceps tendon 20 mm distally and back, starting 10 mm
proximal to the MTJ with 8 throws in total. After completion of the Krackow
stitch, the suture was pretensioned manually to seat the sutures in the biceps
tendon. The free suture limb was pulled to shuttle the biceps tendon to the bone
and then was passed through the tendon. Both suture ends were tied by the use of
a knot pusher with an initial sliding Weston knot followed by 4 reverse
half-hitches (Figure 4D).
Figure 4.
All-suture anchor placement illustration: (A) unicortical 1.8-mm drilling
through the femoral shaft with a drill guide; (B) placed all-suture
anchor; (C) gentle pulling of the suture ends to expand (arrows) and
seat the all-suture anchor to the anterior cortex; (D) final all-suture
anchor fixation with biceps tendon.
All-suture anchor placement illustration: (A) unicortical 1.8-mm drilling
through the femoral shaft with a drill guide; (B) placed all-suture
anchor; (C) gentle pulling of the suture ends to expand (arrows) and
seat the all-suture anchor to the anterior cortex; (D) final all-suture
anchor fixation with biceps tendon.
Biomechanical Testing
Following repair, the biceps tissue was secured by the use of a ribbed, custom
soft tissue clamp to the actuator of the dynamic testing machine (ElectroPuls
E10000; Instron Systems) 2 cm distal to the fixation site of the biceps tendon.
The humerus was fixed to the base of the dynamic testing machine, allowing the
biceps to be pulled vertically along the longitudinal axis of the humeral shaft.
This was done to replicate anatomic force vectors on the biceps tendon. The
setup for biomechanical testing is shown in Figure 5. The biceps tendon was initially
cycled 500 times from 5 to 70 N at a frequency of 1.5 Hz.[3] The force on the tendon was then returned to 5 N and the tendon was
pulled at a rate of 30 mm/min until ultimate failure of the construct.
Displacement was determined by the displacement of the actuator, while force was
recorded by the tensile testing machine’s load cell (Dynacell Biaxial Dynamic
Load Cell; maximum load capacity, ±10 kN; maximum torque capacity, ±100 N·m;
manufacturer-reported accuracy, 0.5% of reading), throughout the testing.
Ultimate load was defined as the highest load attained during testing. Stiffness
was calculated by interpolating a line through the region of the
force-displacement curve located between 30% and 70% of the yield load (yield
load was defined as the first time the force dropped by over 5% of the ultimate
load during testing). Following failure of the construct, failure mode was
qualitatively reported.
Figure 5.
Setup for biomechanical testing.
Setup for biomechanical testing.
Statistical Analysis
The statistical power of the analysis was calculated according to a prior study.[3] Paired t tests were used to assess the primary
comparison of displacement during cyclic loading, ultimate load, and stiffness
between the button and all-suture anchor fixation techniques. A Welch
t test was used to compare the same measurements between
failure mode. P < .05 was deemed statistically significant.
The statistical software R version 3.5.0 was used for all plots and analyses (R
Core Team, with additional package ggplot2).
Results
Two specimens (1 from each group) were excluded for technical reasons (error in the
load cell software, resulting in increased tension during cyclic loading) prior to
testing. Data analysis was performed on the remaining 18 specimens. Two specimens in
the cortical button group (suture pulled through tendon) and 1 specimen in the
all-suture anchor group (knot failure) failed during cyclic loading prior to
beginning the pull-to-failure portion of the test.The mean displacement during cyclic loading was 29.2 ± 9.4 mm for the unicortical
button fixation and 12.5 ± 2.5 mm for the all-suture anchor fixation. The all-suture
anchor fixation had a significantly lower displacement when compared with the button
fixation (P = .005).The mean ultimate load for the unicortical button fixation was 125.4 ± 44.6 N and the
mean stiffness was 48.6 ± 6.8 N/mm (Table 1 and Figure 6). The all-suture anchor fixation
showed a mean load to failure of 170.4 ± 68.8 N and a mean stiffness of 59.3 ± 11.6
N/mm (Figure 6). The
difference in ultimate load (P = .074) and stiffness
(P = .091) for both fixation groups was not significant.
Table 1
Summary of Biomechanical Strength for Both Fixation Groups
Fixation Type
Cyclic Displacement (mm)
Stiffness (N/mm)
Ultimate Load (N)
Button
29.2 ± 9.4
48.6 ± 6.8
125.4 ± 44.6
All-suture anchor
12.5 ± 2.5
59.3 ± 11.6
170.4 ± 68.8
Values are given as mean ± SD.
Figure 6.
Boxplots comparing displacement during cyclic loading, ultimate load, and
stiffness between groups, with the thick horizontal line representing the
median and the box representing the interquartile range.
Summary of Biomechanical Strength for Both Fixation GroupsValues are given as mean ± SD.Boxplots comparing displacement during cyclic loading, ultimate load, and
stiffness between groups, with the thick horizontal line representing the
median and the box representing the interquartile range.
Failure Mode
In all unicortical button fixations, the construct failed on the tendon side with
the suture cutting or tearing through the tendon (Figure 7). In 5 (55.6%) of the all-suture
anchor fixation specimens, failure occurred by means of suture cutting or
tearing through the tendon. In the remaining 4 (44.4%), knot failure was
observed. No statistically significant differences were seen in displacement
(P = .285), ultimate load (P = .445), or
stiffness (P = .699) between those specimens with all-suture
anchor fixation that failed due to the knot versus those that failed due to
tearing through the tendon. No failure of the bone (ie, fracture) or pullout of
either the button or all-suture anchor occurred.
Figure 7.
A right shoulder from the biceps button group during cyclic loading,
showing large displacement after slipping and tearing of the suture
through the tendon.
A right shoulder from the biceps button group during cyclic loading,
showing large displacement after slipping and tearing of the suture
through the tendon.
Discussion
The most important finding of this study is that all-suture anchor fixation using a
Krackow stitch for subpectoral biceps tenodesis provides ultimate load and stiffness
similar to unicortical button fixation using a nonlocking whipstitch. The all-suture
anchor fixation technique was shown to be superior in terms of displacement during
cyclic loading when compared with the button fixation technique. When observing the
failure mechanism, the unicortical button technique (without continuous locking
stitches) showed an increased rate of suture cutting through the tendon, with
significantly higher displacement (Figure 6) when compared with all-suture fixation technique (with
continuous locking stitches). This finding supports the idea that stitch location
and configuration is critical in subpectoral onlay biceps tenodesis, as the stitched
tendon will not be pressed into the bone with an interference screw.[6,29]There is no clear consensus on whether the bone tunnel or the cortical surface
(onlay) healing confers better outcomes. In a rabbit model, Tan et al[30] compared tendon-to-bone healing for both the fixation techniques and found no
significant difference between groups with respect to failure load, stiffness, and
mean volume of newly formed bone. Histological analysis demonstrated direct
tendon-to-bone healing on the outer cortical surface. In the intracortical fixation
group, only 5% of the newly formed bone was located intramedullary, while 95% was
present on the cortical surface.Several cadaveric studies have reported on different supra- and subpectoral fixation
techniques for biceps tenodesis, showing the interference screw to provide the
strongest biomechanical stability.[14,18,20,23] However, complications have been reported, including implant failure,
bioabsorbable screw reactions, and especially humeral fractures.[5,7,17,26] Sears et al[26] reported a case series of humeral fractures following subpectoral biceps
tenodesis. They concluded that the potential stress riser effect created by the
cortical defect, location, and depth of the drill hole may be reduced by limiting
the size of the cortical defect. This stress riser effect and fracture risk may be
increased especially in young overhead athletes with repetitive humeral torque.[24] Buchholz et al[3] introduced a bone-preserving onlay technique using unicortical button
fixation for subpectoral biceps tenodesis. These authors showed that the unicortical
button withstands similar loads when compared with the “gold standard” interference
screw fixation, with ultimate loads of 218 ± 40 N and 212 ± 28 N for the button and
screw, respectively.[3]The present study used a similar testing protocol to compare the onlay unicortical
button technique with an all-suture anchor onlay fixation technique for subpectoral
biceps tenodesis. The all-suture technique requires only a 1.8-mm unicortical drill
hole compared with 3.2 mm for the button technique. Results have confirmed the
hypothesis that the all-suture anchor fixation is biomechanically similar to the
button technique for subpectoral biceps tenodesis in terms of ultimate load, with
means of 170.4 ± 68.8 N and 125.4 ± 44.6 N (ns) for the all-suture anchor and button
fixations, respectively. Ultimate loads of both constructs exceeded 110 N, which is
the force required to hold 1 kg at 90° of elbow flexion and has been proposed as a
good estimate of force during daily activities.[3] Concerns exist of decreased construct stiffness due to the soft component of
all-suture anchors; however, this study presented comparable results in terms of
stiffness for all-suture anchor fixation (59.3 ± 11.6 N/mm) compared with titanium
button fixation (48.6 ± 6.8 N/mm; ns).While ultimate load and stiffness of both constructs are convincing, significantly
higher displacement was observed during cyclic loading for unicortical button
fixation at the MTJ (29.2 ± 9.4 mm) compared with the all-suture anchor fixation
(12.5 ± 2.5 mm). However, there was no button failure observed at the humeral site,
indicating that the tendon tissue and stitch configuration may be the limiting
factors. While both techniques utilized 8 throws through the tendon, a locking
Krackow stitch technique was used for tendon fixation of the all-suture anchor
group, and a suture-loop system with only 1 locking stitch was used for the button
group. We observed that, with the lack of a terminal locking stitch in the
suture-loop system, the most proximal throws of the whipstitch pulled through the
tendon and increased the overall displacement during cyclic testing. This phenomenon
was not observed in the all-suture fixation group. Although the use of 2 different
stitch techniques on the tendon may be construed as one of the study limitations, it
reflects clinical reality. This study did not only compare 2 implants against each
other, but 2 subpectoral biceps tenodesis techniques according to their daily
clinical use. While Spiegl et al[29] emphasized the importance of suture location for proximal biceps tenodesis,
biomechanical performance of stitch configurations and alternative sutures (ie,
tape) should be addressed in future investigations of biceps tenodesis
techniques.The results of this study are in line with those of previously published studies.
Arora et al[1] found ultimate loads of 174 ± 38 N with a stiffness of 73 ± 26 N/mm for
unicortical button fixation, compared with 125.4 ± 44.6 N and 48.6 ± 6.8 N/mm in the
present study. However, the displacement differed greatly: 9 mm in the Arora et al
study compared with 29 mm in the present study. This disparity may be because Arora
and colleagues used a locking suture-loop system, leading to a potentially stronger
suture fixation in the tendon tissue. Furthermore, Arora et al pretensioned the
construct over a 2-minute period prior to measurement of cyclic displacement, which
may have caused initial elongation to go undetected. The current study refrained
from adding a preconditioning protocol, as there is no preconditioning in clinical
practice. Sethi et al[27] biomechanically compared different interference screw fixations to
unicortical button (with button placed at the posterior cortex) fixation for
subpectoral biceps tenodesis. They found a mean ultimate load of 99 ± 17 N and a
mean displacement of 15 ± 8 mm for the unicortical button group. The ultimate load
was 43% lower than that observed by Arora et al, while the displacement was 40%
greater. Similar to the current study, Sethi et al used a suture configuration
without continuous locking stitches, resulting in uniform suture tearing through the
tendon. These findings indicate that suture configuration may compromise the results
of onlay subpectoral biceps tenodesis techniques.To our knowledge, biomechanical performance of subpectoral biceps tenodesis onlay
techniques using all-suture anchor fixation had not been evaluated prior to this
study. Therefore, this study is the first to establish results comparing all-suture
anchor fixation with an established onlay technique for subpectoral biceps
tenodesis. The all-suture anchor onlay fixation technique proved to be
biomechanically similar to the button technique in this cadaveric model.
Limitations
We acknowledge several limitations to this study. First, as mentioned previously,
continuous locking stitches were not utilized in the unicortical button group,
which may predispose the construct to greater displacement and decreased
ultimate failure loads. However, the fixation technique used is consistent with
the manufacturer’s instructions[2] and reflects daily clinical practice. Second, as a time-zero cadaveric
study, the evolution of biomechanical properties in vivo with potential healing
of the tendon to the bone could not be studied. However, the results may provide
an idea of the construct strength in the early postoperative phase when no
advanced healing has occurred.
Conclusion
The all-suture anchor fixation using a Krackow stitch for subpectoral biceps
tenodesis provides ultimate load and stiffness similar to unicortical button
fixation using a nonlocking whipstitch. The all-suture anchor fixation technique was
shown to be superior in terms of displacement during cyclic loading when compared
with the unicortical button fixation technique. However, the results of this study
help to show that the fixation method used on the humeral side is less implicative
of the overall construct strength than the stitch location and technique, as the
biceps tendon tissue and stitch configuration seem to be the limiting factors in
subpectoral onlay tenodesis techniques.
Authors: Augustus D Mazzocca; David Chowaniec; Mark P Cote; Joel Fierra; John Apostolakos; Michael Nowak; Robert A Arciero; Knut Beitzel Journal: Arthroscopy Date: 2012-02-01 Impact factor: 4.772
Authors: Nicholas R Slenker; Kevin Lawson; Michael G Ciccotti; Christopher C Dodson; Steven B Cohen Journal: Arthroscopy Date: 2012-01-28 Impact factor: 4.772
Authors: Shane J Nho; Stefanie N Reiff; Nikhil N Verma; Mark A Slabaugh; Augustus D Mazzocca; Anthony A Romeo Journal: J Shoulder Elbow Surg Date: 2010-05-14 Impact factor: 3.019
Authors: Andrew R Hsu; Neil S Ghodadra; Matthew T Provencher; Paul B Lewis; Bernard R Bach Journal: J Shoulder Elbow Surg Date: 2010-11-04 Impact factor: 3.019
Authors: Ulrich J Spiegl; Sean D Smith; Simon A Euler; Peter J Millett; Coen A Wijdicks Journal: Knee Surg Sports Traumatol Arthrosc Date: 2014-06-27 Impact factor: 4.342
Authors: Simon A Euler; Marilee P Horan; Michael B Ellman; Joshua A Greenspoon; Peter J Millett Journal: Arch Orthop Trauma Surg Date: 2016-01-25 Impact factor: 3.067
Authors: L Willemot; R Elfadalli; K C Jaspars; M H Ahw; J Peeters; N Jansen; G Declerq; O Verborgt Journal: Acta Orthop Belg Date: 2016-08 Impact factor: 0.500
Authors: Lucca Lacheta; Marilee P Horan; Philip C Nolte; Brandon T Goldenberg; Travis J Dekker; Peter J Millett Journal: Orthop J Sports Med Date: 2022-06-21
Authors: Alexander Otto; Sebastian Siebenlist; Joshua B Baldino; Matthew Murphy; Lukas N Muench; Julian Mehl; Elifho Obopilwe; Mark P Cote; Andreas B Imhoff; Augustus D Mazzocca Journal: JSES Int Date: 2020-09-21