| Literature DB >> 31662907 |
Guiswe Gnowe1, Henri Paul Ekobena Fouda2, Mbo Amvene Jéremie3, Takmou Pascal4, Bonaventure Babinne Graobe1.
Abstract
BACKGROUND: The purpose of this study was to estimate the doses delivered to adult patients during chest examination for comparison with those elsewhere and to establish a local diagnostic reference level for the chest. The doses delivered in the standard X-ray examinations are not sufficiently optimized and controlled. The working protocols for the same exam given differ for similar morphotypes within the same hospital structure.Entities:
Year: 2019 PMID: 31662907 PMCID: PMC6778874 DOI: 10.1155/2019/3619498
Source DB: PubMed Journal: Radiol Res Pract ISSN: 2090-195X
Sociodemographic data.
| Radiography | Sex | Age (years) | Weight (kg) | |
|---|---|---|---|---|
| Chest | M | 55 | 29–53 (40.92 ± 7.42) | 64–80 (70.84 ± 4.7) |
| F | 50 | 23–56 (35.58 ± 10.19) | 60–78 (66.66 ± 4.71) | |
| Total |
| |||
M: male. F: female.
Tube yield in mR and mGy · (mAs)−1.
| Radiography | Incidence | Output (mR) | Output mGy · (mAs)−1 |
|---|---|---|---|
| Min-max | Min-max | ||
| Chest | PA | 75.22–81.75 | 0.65–0.71 |
PA: posteroanterior. Min: minimum. Max: maximum.
Technical parameters used to conduct this exam.
| Radiography | Incidence | kV | mAs | FFD | FSD | ESD | 3rd quartile | SD | |
|---|---|---|---|---|---|---|---|---|---|
| Min-max | Max/min | ||||||||
| Chest | PA | 120–125 (122.8 ± 1.72) | 2.5–4 (3.17 ± 0.46) | 140–180 (180 ± 0.09) | 160–200 (128 ± 0.20) | 0.08–0.09 | 1.04 | 0.18 | 0.21 |
kV: kilovolt. mAs: milliampere second. FFD: focus-film-distance. FSD: focus-skin-distance. ESD: entrance skin dose. SD: standard deviation.
Comparison of the entrance skin dose (mGy) with those obtained elsewhere and at DRLs.
| Radiography | Incidence | Our study | IRSN (DRL) | Sudan [ | Iran [ | Mali [ |
|---|---|---|---|---|---|---|
| Chest | PA | 0.18 ± 0.21 | 0.3 | 0.53 ± 0.24 | 0.70 ± 0.38 | 3.84 |