Radiation doses to patients undergoing standard radiographic examinations: a comparison between two methods.

V Tsapaki1, I A Tsalafoutas, I Chinofoti, A Karageorgi, E Carinou, V Kamenopoulou, E N Yakoumakis, E D Koulentianos.   

Abstract

The objective of the study was to derive a mathematical method for calculating the entrance surface dose (ESD) from exposure factors for all tube potentials used in clinical practice and to compare the calculated ESDs (ESD(C)) with those measured (ESD(TLD)) using thermoluminescent dosemeters (TLDs). The exposure parameters of 43 patients who underwent (a) posteroanterior (PA) and lateral (LAT) chest examination (13 patients), (b) supine abdomen (10 patients), (c) erectus abdomen (10 patients), or (d) urinary tract examination (10 patients) were recorded. Patient ESD was directly measured by TLDs and calculated from exposure factors. The differences between ESD(C) and ESD(TLD) were quite small and could be explained by the uncertainties involved in both methods, in all but the PA chest examination where the ESD(C) was about 50% larger than ESD(TLD). However, in PA chest the ESD(TLD) was close to the minimum detectable dose of TLDs, questioning the accuracy of ESD(TLD). Further investigation showed that using the high tube potential technique (130 kV) in the PA chest examination resulted in very short exposure times, in the region of 4 ms. In such short exposure times, the X-ray generator operation presented stability problems that led to loss of output linearity and consequently to false calculation of ESD. The calculation method offers a reliable and cheap alternative to the measurement of ESD by TLD, provided that the exposure times are not as short as in the PA chest examinations recorded in this study, so that the output linearity with tube current-time product (mAs) is maintained.

Entities:  

Mesh:

Year:  2006        PMID: 16885174     DOI: 10.1259/bjr/87150291

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  6 in total

1.  The empirical formula for calculating the incident air Kerma in intraoral radiographic imaging.

Authors:  Zoran Mirkov; Katarina M Rajković; Jovan B Stanković; Dario Faj
Journal:  Dentomaxillofac Radiol       Date:  2021-07-20       Impact factor: 2.419

Review 2.  How to Measure/Calculate Radiation Dose in Patients?

Authors:  Reinhard Loose; Michael Wucherer
Journal:  Cardiovasc Intervent Radiol       Date:  2021-03-03       Impact factor: 2.740

3.  Optimization of patient radiation protection in pelvic X-ray examination in Ghana.

Authors:  Eric K Ofori; William K Antwi; Diane N Scutt; Matt Ward
Journal:  J Appl Clin Med Phys       Date:  2012-07-05       Impact factor: 2.102

4.  Exposure Levels of Adult Patients during Radiographic Examinations: Sinuses and Coastal Grill Cases at the Ngaoundere Regional Hospital, Cameroon.

Authors:  Guiswe Gnowe; Fouda H P Ekobena; Amvene J Mbo; Guena M Neossi
Journal:  Radiol Res Pract       Date:  2019-04-30

5.  Performance Evaluation of a Multichannel All-In-One Phantom Dosimeter for Dose Measurement of Diagnostic X-ray Beam.

Authors:  Hyesu Jeon; Wook Jae Yoo; Sang Hun Shin; Guwon Kwon; Mingeon Kim; Hye Jin Kim; Young Beom Song; Kyoung Won Jang; Won Sik Youn; Bongsoo Lee
Journal:  Sensors (Basel)       Date:  2015-11-11       Impact factor: 3.576

6.  Assessment of the Local Exposure Level during Adult Chest X-Rays at the Ngaoundere Regional Hospital, Cameroon.

Authors:  Guiswe Gnowe; Henri Paul Ekobena Fouda; Mbo Amvene Jéremie; Takmou Pascal; Bonaventure Babinne Graobe
Journal:  Radiol Res Pract       Date:  2019-09-22
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.