Accuracy and precision of 3D-printed implant surgical guides with different implant systems: An in vitro study.

Matthew Yeung1, Aous Abdulmajeed2, Caroline K Carrico3, George R Deeb4, Sompop Bencharit5.   

Abstract

STATEMENT OF PROBLEM: Implant guided surgery systems promise implant placement accuracy and precision beyond straightforward nonguided surgery. Recently introduced in-office stereolithography systems allow clinicians to produce implant surgical guides themselves. However, different implant designs and osteotomy preparation protocols may produce accuracy and precision differences among the different implant systems.
PURPOSE: The purpose of this in vitro study was to measure the accuracy and precision of 3 implant systems, Tapered Internal implant system (BioHorizons) (BH), NobelReplace Conical (Nobel Biocare) (NB), and Tapered Screw-Vent (Zimmer Biomet) (ZB) when in-office fabricated surgical guides were used.
MATERIAL AND METHODS: A cone beam computed tomography (CBCT) data set of an unidentified patient missing a maxillary right central incisor and intraoral scans of the same patient were used as a model. A software program (3Shape Implant Studio) was used to plan the implant treatment with the 3 implant systems. Three implant surgical guides were fabricated by using a 3D printer (Form 2), and 30 casts were printed. A total of 10 implants for each system were placed in the dental casts by using the manufacturer's recommended guided surgery protocols. After implant placement, postoperative CBCT images were made. The CBCT cast and implant images were superimposed onto the treatment-planning image. The implant positions, mesiodistal, labiopalatal, and vertical, as well as implant angulations were measured in the labiolingual and mesiodistal planes. The displacements from the planning in each dimension were recorded. ANOVA with the Tukey adjusted post hoc pairwise comparisons were used to examine the accuracy and precision of the 3 implant systems (α=.05).
RESULTS: The overall implant displacements were -0.02 ±0.13 mm mesially (M), 0.07 ±0.14 mm distally (D), 0.43 ±0.57 mm labially (L), and 1.26 ±0.80 mm palatally (P); 1.20 ±3.01 mm vertically in the mesiodistal dimension (VMD); 0.69 ±2.03 mm vertically in the labiopalatal dimension (VLP); 1.69 ±1.02 degrees in mesiodistal angulation (AMD); and 1.56 ±0.92 degrees in labiopalatal angulation (ALP). Statistically significant differences (ANOVA) were found in M (P=.026), P (P=.001), VMD (P=.009), AMD (P=.001), and ALP (P=.001). ZB showed the most displacements in the M and vertical dimensions and the least displacements in the P angulation (P<.05), suggesting statistically significant differences among the M, VMD, VLP, AMD, and ALP. NB had the most M variation. ZB had the least P deviation. NB had the fewest vertical dimension variations but the most angulation variations.
CONCLUSIONS: Dimensional and angulation displacements of guided implant systems by in-office 3D-printed fabrication were within clinically acceptable limits: <0.1 mm in M-D, 0.5 to 1 mm in L-P, and 1 to 2 degrees in angulation. However, the vertical displacement can be as much as 2 to 3 mm. Different implant guided surgery systems have strengths and weaknesses as revealed in the dimensional and angulation implant displacements.
Copyright © 2019 Editorial Council for the Journal of Prosthetic Dentistry. Published by Elsevier Inc. All rights reserved.

Entities:  

Year:  2019        PMID: 31653399     DOI: 10.1016/j.prosdent.2019.05.027

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  11 in total

1.  [Research advances in the use of digital surgical guides in implantology].

Authors:  Xiao-Hua Wang; Ai-Peng Liu; Wen-Zheng Deng
Journal:  Hua Xi Kou Qiang Yi Xue Za Zhi       Date:  2020-02-01

2.  3D printed fracture reduction guides planned and printed at the point of care show high accuracy - a porcine feasibility study.

Authors:  Andreas Hecker; Sophie C Eberlein; Frank M Klenke
Journal:  J Exp Orthop       Date:  2022-09-27

3.  Accuracy of Computer-Assisted Dynamic Navigation in Implant Placement with a Fully Digital Approach: A Prospective Clinical Trial.

Authors:  Cornelia Edelmann; Martin Wetzel; Anne Knipper; Ralph G Luthardt; Sigmar Schnutenhaus
Journal:  J Clin Med       Date:  2021-04-21       Impact factor: 4.241

Review 4.  The Modern and Digital Transformation of Oral Health Care: A Mini Review.

Authors:  Muhammad Syafiq Alauddin; Ahmad Syukran Baharuddin; Mohd Ifwat Mohd Ghazali
Journal:  Healthcare (Basel)       Date:  2021-01-25

5.  Does the macro design of an implant affect the accuracy of template-guided implantation? A prospective clinical study.

Authors:  Sigmar Schnutenhaus; Cornelia Edelmann; Heike Rudolph
Journal:  Int J Implant Dent       Date:  2021-04-26

6.  Material Extrusion Based Fabrication of Surgical Implant Template and Accuracy Analysis.

Authors:  Chengyu Zhang; Yanping Yuan; Jimin Chen
Journal:  Materials (Basel)       Date:  2022-02-25       Impact factor: 3.623

7.  Comparing digital and traditional guides in first molar implant surgery: A randomized clinical trial.

Authors:  Yajie Yang; Chen Hu; Yanli Zhang; Linlin Wang; Longquan Shao; Jie You
Journal:  Technol Health Care       Date:  2022       Impact factor: 1.205

8.  Case report: Fabrication of a dental implant guide based on tetrahedron positioning technology.

Authors:  Jie Lin; Zhenxiang Lin; Zhiqiang Zheng
Journal:  BMC Oral Health       Date:  2021-07-07       Impact factor: 2.757

9.  Effect of Tooth Types on the Accuracy of Dental 3D Scanners: An In Vitro Study.

Authors:  Keunbada Son; Kyu-Bok Lee
Journal:  Materials (Basel)       Date:  2020-04-09       Impact factor: 3.623

10.  Accuracy of Computer-Assisted Dynamic Navigation as a Function of Different Intraoral Reference Systems: An In Vitro Study.

Authors:  Sigmar Schnutenhaus; Anne Knipper; Martin Wetzel; Cornelia Edelmann; Ralph Luthardt
Journal:  Int J Environ Res Public Health       Date:  2021-03-21       Impact factor: 3.390

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.