| Literature DB >> 31649988 |
Hilde L Von Volkmann1,2, Ingeborg Brønstad1,2, Torunn Fiskerstrand3,4, Oddrun Anita Gudbrandsen2.
Abstract
BACKGROUND: Reported concentrations of serotonin in platelet-poor plasma (PPP) in healthy subjects vary widely due to different pre-analytical procedures. AIM: To examine how different pre-analytical conditions affect the measured concentration of serotonin in PPP.Entities:
Keywords: 5-HT; 5-Hydroxytryptamine; Plasma; Platelet-poor plasma; Pre-analytical procedure; Serotonin
Year: 2019 PMID: 31649988 PMCID: PMC6804510 DOI: 10.1016/j.plabm.2019.e00136
Source DB: PubMed Journal: Pract Lab Med ISSN: 2352-5517
Overview of the preparation methods of platelet-poor plasma for serotonin analyses.
| Pre-analytical method | Sampling tube | First centrifugation (EDTA blood) | Second centrifugation (supernatant) | References |
|---|---|---|---|---|
| A | 3 ml VACUETTE® K2EDTA tube | 200× | 4500× | Recommended by IBL [ |
| B | 3 ml VACUETTE® K2EDTA tube | 150× | 550× | Yubero-Lahoz [ |
| C | 3 ml VACUETTE® K2EDTA tube | 100× | 14.500× | Frost [ |
| D | 3 ml VACUETTE® K2EDTA tube | Heat shock: | none | Maurer-Spurej [ |
| E | 5 ml VACUETTE® | 1800× | none | |
| F | 5 ml VACUETTE® K2EDTA gel tube | 1800× | none |
Measured serotonin concentrations in platelet-poor plasma from healthy donors (N = 9).
| Pre-analytical method | Measured Serotonin Concentration (ng/mL)1 | |
|---|---|---|
| A | 0.9 ± 0.5a | |
| B | 2.5 ± 2.3a | |
| C | 1.0 ± 0.5a | |
| D | 9.8 ± 11.1b | |
| E | 41.4 ± 18.6 | |
| F | 10.7 ± 1.6b |
Serotonin concentration is presented as mean ± standard deviation. Serotonin concentrations in PPP prepared by method E was discarded from statistical analysis since results deviated strongly from results obtained with the other pre-analytical methods.
1Methods A, B, C, D and F are compared using one-way ANOVA (p = 1.4 × 10−4) with Tukey post-hoc test; means in a column with different letters are significantly different (p < 0.05).