To be reaffirmed in 1987 for lack of a ready replacement, the flexural (elastic-bending) test protocol of ADA Specification No. 32 is judged inadequate. The protocol is problematic because of potentials for erroneous use of the theoretical component, incompatible with the flexible titanium alloy and multistrand stainless steel wires marketed subsequent to the preparation of the specification, and obscure to the clinician because it dictates quantifications of mechanical (pertaining to material only) rather than structural properties (including wire shape and size influences). A five-point elastic-bending test is proposed that stimulates wire activation toward engagement of a single, malaligned tooth crown. An experimental study was undertaken to determine values of transverse stiffness and corresponding elastic range for a broad sample of orthodontic wires and in the process to evaluate the proposed alternative test. Reduced test results are presented; comparisons of rankings and ratios from available theoretical developments and other experimental outcomes, including findings from the existing standard test, were completed. The difficulties with the existing protocol are largely eliminated with the alternative test; a test fixture and procedures are relatively straightforward to fabricate and follow, and the structural characteristics quantified are more meaningful to the practitioner.

Entities:  

Mesh:

Year:  1988        PMID: 3163884     DOI: 10.1016/0889-5406(88)90078-9

Source DB:  PubMed          Journal:  Am J Orthod Dentofacial Orthop        ISSN: 0889-5406            Impact factor:   2.650


  4 in total

1.  Bending and torquing accuracy of the bending art system (BAS).

Authors:  H Fischer-Brandies; W Orthuber; L Pohle; D Sellenrieck
Journal:  J Orofac Orthop       Date:  1996-02       Impact factor: 1.938

2.  [A computer-controlled flexing test for determining the elastic parameters of highly flexible orthodontic wires].

Authors:  R Plietsch; C Bourauel; D Drescher; B Nellen
Journal:  Fortschr Kieferorthop       Date:  1994-04

3.  Comparison of Superelasticity of Nickel Titanium Orthodontic Arch wires using Mechanical Tensile Testing and Correlating with Electrical Resistivity.

Authors:  Aravind Sivaraj
Journal:  J Int Oral Health       Date:  2013-06-23

4.  Mechanical properties of beta-titanium wires.

Authors:  Júlio A Gurgel; Célia R M Pinzan-Vercelino; John M Powers
Journal:  Angle Orthod       Date:  2011-02-07       Impact factor: 2.079

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.