Assessment of signal-to-noise ratio and contrast-to-noise ratio in 3 T magnetic resonance imaging in the presence of zirconium, titanium, and titanium-zirconium alloy implants.

Husniye Demirturk Kocasarac1, Emine Sebnem Kursun-Cakmak2, Gulbahar Ustaoglu3, Seval Bayrak4, Kaan Orhan5, Marcel Noujeim6.   

Abstract

OBJECTIVE: We quantitatively compared the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) in T1 weighted imaging (WI) and T2 WI sequences in 3 Tesla (T) magnetic resonance imaging (MRI) using zirconium, titanium (grades 4 and 5), and titanium-zirconium alloy implants to evaluate the effect of implant type and imaging sequence. STUDY
DESIGN: MRI was acquired using a 3 T magnet with a 16-channel head coil. Implants of each type were mounted in gel and scanned in axial, coronal, and sagittal planes using fast spin echo sequences in T1 WI (TR = 600, TE = 12 milliseconds) and T2 WI (TR = 3000, TE = 80 milliseconds) sequences. Data were transferred to Synapse 3-D software, and images were measured twice by an oral and maxillofacial radiologist blinded to the type of implants.
RESULTS: Zirconium implants resulted in the lowest SNR and CNR values (P < .05). No significant differences were identified between titanium (grades 4 and 5) and titanium-zirconium implants. The T2 WI sequence had a significantly higher SNR and CNR than T1 WI. There was no difference in intraobserver agreement between T1 WI and T2 WI.
CONCLUSIONS: CNR and SNR at 3 T MRI are dependent on implant type and imaging sequence. Titanium (grades 4 and 5) and titanium-zirconium implants and the T2 WI sequence produced higher SNR and CNR values.
Copyright © 2019 Elsevier Inc. All rights reserved.

Entities:  

Year:  2019        PMID: 31628073     DOI: 10.1016/j.oooo.2019.08.020

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol


  2 in total

1.  Magnetic resonance imaging artefacts caused by orthodontic appliances and/or implant-supported prosthesis: a systematic review.

Authors:  Katrine Mølgaard Johannsen; João Marcus de Carvalho E Silva Fuglsig; Brian Hansen; Ann Wenzel; Rubens Spin-Neto
Journal:  Oral Radiol       Date:  2022-09-30       Impact factor: 1.882

2.  Trabecular Bone Assessment Using Magnetic-Resonance Imaging: A Pilot Study.

Authors:  Lauren Bohner; Pedro Tortamano; Norbert Meier; Felix Gremse; Johannes Kleinheinz; Marcel Hanisch
Journal:  Int J Environ Res Public Health       Date:  2020-12-11       Impact factor: 3.390

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.