OBJECTIVE: Gradually increasing number of minimally invasive spine surgery (MISS), there is an increasing risk of radiation exposure to medical personnel during the surgery. We measured the radiation exposure of the operating room personnel during MISS, tried to find the riskiest person, and checked the effectiveness of a new lead-composite shielding curtain.Entities:
Keywords: Minimally invasive surgical procedures; Protective device; Radiation exposure
Year: 2019 PMID: 31614404 PMCID: PMC7136108 DOI: 10.14245/ns.1938282.141
Source DB: PubMed Journal: Neurospine ISSN: 2586-6591
Fig. 1.Positions of medical staffs during operation. The stars indicate the location of dosimeter measuring radiation exposure dose in anteroposterior (A) and lateral (B) views.
Fig. 2.The use of a shielding curtain. (A) Two frame bars mounted on the operating table. (B, C) A shielding curtain attached to the frame bars. (D) The patient positioned with shielding curtain during surgery. Written informed consent was obtained from the medical staffs.
Demographic data of the patients
| Variable | Nonshield group (n = 35) | Shield group (n = 35) | p-value |
|---|---|---|---|
| Sex, male:female | 15:20 | 13:22 | |
| Age (yr) | 64.8 ± 12.4 | 66.3 ± 13.0 | 0.30 |
| Body mass index (kg/m2) | 23.7 ± 2.2 | 24.0 ± 2.4 | 0.22 |
| BMD (T-score) | -1.3 ± 1.2 | -1.1 ± 1.3 | 0.20 |
| No. of surgery types | |||
| Discectomy (1 level) | 10 | 8 | |
| 1-Level fusion | 16 | 18 | |
| 2-Level fusion | 5 | 6 | |
| 3-Level fusion | 4 | 3 | |
| Radiation time (sec/level) | |||
| 1 Level discectomy | 11.7 ± 2.6 (n = 10) | 10.6 ± 4.1 (n = 8) | 0.96 |
| 1 Level fusion | 54.2 ± 15.9 (n = 38) | 50.9 ± 12.2 (n = 39) | 0.31 |
Values are presented as mean±standard deviation or number.
BMD, bone mineral density.
Radiation exposures in the nonshield and shield groups and protection rates
| Variable | AP view | Lateral view | ||||
|---|---|---|---|---|---|---|
| Nonshield (μSv/hr) | Shield (μSv/hr) | Protection (%) | Nonshield (μSv/hr) | Shield (μSv/hr) | Protection (%) | |
| Operator | 679.7 ± 186.4[ | 39.8 ± 16.5[ | 94.1 | 2,330.9 ± 967.4 | 86.2 ± 55.7[ | 96.3 |
| Assistant | 210.5 ± 144.4 | 196.4 ± 98.4 | 6.7 | 160.6 ± 121.9 | 154.7 ± 88.5 | 3.6 |
| Nurse | 46.9 ± 27.9[ | 11.1 ± 10.4[ | 76.3 | 97.1 ± 48.6 | 25.4 ± 17.3[ | 73.9 |
| Anesthesiologist | 16.8 ± 10.3[ | 15.4 ± 9.9 | 8.5 | 61.2 ± 28.6 | 62.5 ± 23.4 | -2.1 |
Values are presented as mean±standard deviation.
Protection rate=(nonshield radiation dose – shield radiation dose)/nonshield radiation dose×100.
AP, anteroposterior.
p<0.001 vs. nonshield group.
p<0.001 vs. lateral view.