Literature DB >> 27171657

Fluoroscopic Radiation Exposure in Spinal Surgery: In Vivo Evaluation for Operating Room Personnel.

Daniel S Mulconrey1.   

Abstract

STUDY
DESIGN: Prospective in vivo investigation of fluoroscopic radiation exposure during spinal surgery.
OBJECTIVE: To quantify the total amount of radiation dosage and identify techniques to maintain safe levels of fluoroscopic exposure in the operating room. SUMMARY OF BACKGROUND DATA: No previous study has performed an in vivo examination of fluoroscopic radiation exposure to the spinal surgeon and operating room personnel. Previous similar studies were in vitro, used older versions of fluoroscopy, and increased fluoro times associated with pedicle screw placement.
METHODS: Thirty-five surgeries were evaluated in 18 males and 17 females (mean age 52.4 y; range, 26.0-79.4). Surgeries included 37 lumbar levels fused, 45 lumbar decompressions, 8 anterior cervical fusions, and 19 transforaminal lumbar interbody fusion procedures. Spinal instrumentation was implemented in all fusion procedures (104 lumbar pedicle screws, 14 iliac, 22 anterior cervical). Radiation dosimetry was obtained through unprotected badges placed on surgeon's chest, first assistant chest, cranial and caudal end of operating table.
RESULTS: Total fluoroscopic time was 37.01 minutes. Mean fluoroscopic time with lumbar spine instrumentation was greater than decompression alone (1.74 vs. 0.22 min). Total fluoroscopic radiation exposure was obtained for surgeon (1225 mrem), first assistant (369 mrem), cranial table (92 mrem), and caudal table (150 mrem). Mean dose/min (mrem/min) was calculated for surgeon (33.1), first assistant (9.97), cranial table (2.48), and caudal table (4.05). To remain below the maximum yearly permissible level of radiation, the estimated total number of minutes for the surgeon would be 453.
CONCLUSIONS: The results of this in vivo study indicate fluoroscopic dosage to the spine surgeon remains below the annual maximum limit of radiation exposure. Increasing distance from radiation source led to a significantly diminished in vivo dosimetry reading. Monitoring fluoroscopic time and maintaining a distance from the beam source, radiation exposure to the spine surgeon may be kept within current safety standards.

Mesh:

Year:  2016        PMID: 27171657     DOI: 10.1097/BSD.0b013e31828673c1

Source DB:  PubMed          Journal:  Clin Spine Surg        ISSN: 2380-0186            Impact factor:   1.876


  17 in total

1.  Are computer numerical control (CNC)-manufactured patient-specific metal templates available for posterior thoracic pedicle screw insertion? Feasibility and accuracy evaluation.

Authors:  Xiangxue Kong; Lei Tang; Qiang Ye; Wenhua Huang; Jianyi Li
Journal:  Eur Spine J       Date:  2017-07-17       Impact factor: 3.134

2.  Radiation exposure for the surgical team in a hybrid-operating room.

Authors:  Konard Schuetze; A Eickhoff; C Dehner; M Schultheiss; F Gebhard; P H Richter
Journal:  J Robot Surg       Date:  2018-05-10

3.  Spine surgery is kyphosing to spine surgeon.

Authors:  Ezan A Kothari; Timur M Urakov
Journal:  Acta Neurochir (Wien)       Date:  2020-02-10       Impact factor: 2.216

4.  Radiation exposure to the patients in thoracic and lumbar spine fusion using a new intraoperative cone-beam computed tomography imaging technique: a preliminary study.

Authors:  Ludovic Kaminski; V Cordemans; O Cartiaux; M Van Cauter
Journal:  Eur Spine J       Date:  2017-02-06       Impact factor: 3.134

Review 5.  Does less invasive spine surgery result in increased radiation exposure? A systematic review.

Authors:  Elizabeth Yu; Safdar N Khan
Journal:  Clin Orthop Relat Res       Date:  2014-06       Impact factor: 4.176

6.  Decreasing radiation dose with FluoroLESS Standalone Anterior Cervical Fusion.

Authors:  Kingsley R Chin; Fabio J R Pencle; Kathleen A Quijada; Moawiah S Mustafa; Luai S Mustafa; Jason A Seale
Journal:  J Spine Surg       Date:  2018-12

7.  Fluoroscopic Cranial Radiation Exposure in Spine Surgery: A Prospective Single-Center Evaluation in Operating Room Personnel.

Authors:  Laila Malani Mohammad; James Messegee; M Omar Chohan; Christopher L Taylor
Journal:  Int J Spine Surg       Date:  2019-02-22

8.  Radiation Exposure in Posterior Lumbar Fusion: A Comparison of CT Image-Guided Navigation, Robotic Assistance, and Intraoperative Fluoroscopy.

Authors:  Erik Wang; Jordan Manning; Christopher G Varlotta; Dainn Woo; Ethan Ayres; Edem Abotsi; Dennis Vasquez-Montes; Themistocles S Protopsaltis; Jeffrey A Goldstein; Anthony K Frempong-Boadu; Peter G Passias; Aaron J Buckland
Journal:  Global Spine J       Date:  2020-02-27

9.  Puncture Reduction in Percutaneous Transforaminal Endoscopic Discectomy with HE's Lumbar LOcation (HELLO) System: A Cadaver Study.

Authors:  Guoxin Fan; Xiaofei Guan; Qi Sun; Annan Hu; Yanjie Zhu; Guangfei Gu; Hailong Zhang; Shisheng He
Journal:  PLoS One       Date:  2015-12-16       Impact factor: 3.240

10.  Efficiency of lead aprons in blocking radiation - how protective are they?

Authors:  Seung-Jae Hyun; Ki-Jeong Kim; Tae-Ahn Jahng; Hyun-Jib Kim
Journal:  Heliyon       Date:  2016-05-27
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.