Majid Moshirfar1,2,3, Benjamin Buckner3, David B Rosen4, Madeline B Heiland4, Yasmyne C Ronquillo3, David F Skanchy5, Harry Y Liu6, Tim Melton3, Liliana Werner1, Phillip C Jr Hoopes1,3.
Abstract
The purpose of this case series is to report visual outcomes in patients who underwent explantation of the Raindrop® hydrogel corneal inlay. Retrospective chart review comprising four cases of explantation of the Raindrop® corneal shape-changing hydrogel inlay: pre-implantation, pre-explantation, and post-explantation values for uncorrected distance visual acuity, uncorrected near visual acuity, and corrected distance visual acuity (CDVA) were measured; keratometric and tomographic data were collected using the Pentacam system (Oculus, Inc). Three eyes were explanted for progressive haze after implantation that persisted even after removal; one eye was explanted due to poor visual acuity with no haze formation. All patients experienced decreased unaided and corrected distance visual acuity. Persistent increase in corneal thickness and mean keratometry was noted post-explantation. All four patients regained their original near visual acuities, but one patient had persistent one-line loss in CDVA. There are long lasting tomographic corneal changes following Raindrop inlay explantation. In addition, persistent increased corneal thickness could be related to semi-permanent changes in corneal structure and may account for residual haze experienced by patients. After explantation, patients may not return to baseline CDVA.Entities:
Keywords: Corrected Distance Visual Acuity; Explantation; Presbyopia; Raindrop Corneal Inlay; Visual Prognosis
Year: 2019 PMID: 31598515 PMCID: PMC6778472
Source DB: PubMed Journal: Med Hypothesis Discov Innov Ophthalmol ISSN: 2322-3219
Figure 1A: Light microscopy of explanted Raindrop® inlay, explanted from case 1 at 20x magnification. The specimen appears to be quite intact. Note that the deposits appear to be mostly related to the balanced salt solution as they are also observed on the glass slide outside of the area of the device. B: Scanning electron microscopy photograph of the inlay, showing the fern-like deposits on its surface
Figure 2Photographs of case 1 taken immediately before (A), 1 day after (B), and 1 month after (C) explantation of the Raindrop® corneal inlay. Note that the haze appears to be subsiding slightly in the more direct lighting conditions but lingers with the tangential illumination. The observer may mistakenly believe the inlay is still in place when it is in fact explanted and there is only residual haze and fibrosis
Changes in Visual Acuity Pre- and Post-explantation of the Raindrop® Hydrogel Corneal Inlay
| UDVA | UNVA | CDVA | |||||||
|---|---|---|---|---|---|---|---|---|---|
| Patients |
|
|
|
|
|
|
|
|
|
| 1 | 20/25 | 20/300 | 20/80 | 20/100 | 20/40 | 20/63 | 20/20 | 20/40+2 | 20/25+ |
| 2 | 20/20 | 20/25 | 20/20- | 20/50 | 20/40 | 20/50 | 20/20 | 20/25+2 | 20/20- |
| 3 | 20/20 | 20/70-2 | 20/40+2 | 20/100 | 20/50 | 20/100 | 20/20 | 20/25- | 20/20 |
| 4 | 20/20-2 | 20/50 | 20/50 | 20/100 | 20/80 | 20/100 | 20/20 | 20/40 | NA |
Abbreviations: CDVA: corrected distance visual acuity; UDVA: uncorrected distance visual acuity; UNVA: uncorrected near visual acuity; NA: not available; Pre-inlay: before implantation of inlay; Pre-explant: before explantation of inlay; Post-explant: after explantation of inlay
Changes in keratometry and CCT Pre- and Post-explantation of the Raindrop® Hydrogel Corneal Inlay
| Pre-Implantation | Pre-explantation | Post-explantation | ||||
|---|---|---|---|---|---|---|
| Patients |
|
|
|
|
|
|
| 1 | 43 | 537 | 45.2 | 554 | 43.7 | 555 |
| 2 | 45 | 554 | 45.2 | 565 | NA | NA |
| 3 | 43.2 | 580 | 48.8 | 632 | 43.9 | 596 |
| 4 | 42.4 | 559 | 44.4 | 586 | NA | NA |
Abbreviations: Km: mean keratometry; CCT: central corneal thickness; D: diopter; μm: micrometer; NA: not available
Changes in manifest refraction Pre- and Post-explantation of the Raindrop® Hydrogel Corneal Inlay
| Patient | MRSE (D) Pre-implantation | MRSE (D) Pre-explantation | MRSE (D) post-explantation | Change from baseline |
|---|---|---|---|---|
| 1 | -0.375 | -1.75 | -1.375 | -1 |
| 2 | -0.125 | -0.25 | -0.25 | -0.125 |
| 3 | +0.75 | -0.125 | +0.5 | -0.25 |
| 4 | +0.5 | -0.125 | NA | NA |
Abbreviations: MRSE: manifest refraction spherical equivalent; D: diopter; NA: not-available.
Figure 3PENTACAM data for case 1 of the Raindrop® Hydrogel Corneal Inlay
Figure 4Densitometry from PENTACAM for case 1 of the Raindrop® Hydrogel Corneal Inlay at various time-points