Literature DB >> 31594008

Establishing Reliability Criteria for Liver ElastPQ Shear Wave Elastography (ElastPQ-SWE): Comparison Between 10, 5 and 3 Measurements.

Davide Roccarina1, Laura Iogna Prat1, Elena Buzzetti1, Marta Guerrero Misas1, Francesco Marcello Aricó2, Francesca Saffioti1,3, Matteo Rosselli1, Massimo Pinzani1, Aileen Marshall1, Douglas Thorburn1, Emmanuel Tsochatzis1.   

Abstract

PURPOSE: ElastPQ is a new elastography technique for non-invasive liver fibrosis staging. However, it does not have validated reliability criteria. We tested the reliability of a different number of measurements in patients with chronic liver disease and explored whether the application of quality criteria improves the diagnostic performance.
MATERIALS AND METHODS: All patients underwent liver stiffness assessment (LSM) with ElastPQ and Fibroscan (F-TE). The mean, median, standard deviation (SD) and interquartile range (IQR) of 10, 5 and 3 measurements were retrospectively collected for each patient and compared to each other. Liver histology was available in a subset of patients.
RESULTS: Overall, 400 patients met the inclusion criteria. Non-alcoholic fatty liver disease (NAFLD) was the most represented etiology (75 %), followed by primary sclerosing cholangitis (PSC), primary biliary cholangitis (PBC) and autoimmune hepatitis (AIH). The correlation of medians was significantly better between 10 and 5 measurements than between 10 and 3. The difference of medians was significant only in the comparison between 10 and 3 measurements. The correlation between ElastPQ and F-TE was equally good for 10 and 5 measurements and significantly improved after an IQR/median ≤ 30 % was applied. The diagnostic performance of ElastPQ was better with the median value of 10 and 5 measurements and improved if LSM values were obtained with IQR/M ≤ 30 %.
CONCLUSION: The median value of 5 valid LSMs suffices for the reliable estimation of liver stiffness using ElastPQ. The quality criterion of IQR/M ≤ 30 % should also be followed when using this technique. Thieme. All rights reserved.

Entities:  

Mesh:

Year:  2019        PMID: 31594008     DOI: 10.1055/a-1010-6052

Source DB:  PubMed          Journal:  Ultraschall Med        ISSN: 0172-4614            Impact factor:   6.548


  2 in total

1.  Shear wave elastography using sound touch elastography and supersonic shear imaging for liver measurements: a comparative study.

Authors:  Hongjin Xiang; Wenwu Ling; Lin Ma; Lulu Yang; Tang Lin; Yan Luo
Journal:  Quant Imaging Med Surg       Date:  2022-05

2.  Portal hypertensive enteropathy: multimodality assessment through computed tomography and magnetic resonance enterography.

Authors:  Giuseppe Cicero; Francesco Marcello Aricò; Anna Viola; Velio Ascenti; Silvio Mazziotti
Journal:  Radiol Case Rep       Date:  2022-10-05
  2 in total