Literature DB >> 31590209

Fabrication, Measurement and Time Decay of the Electromagnetic Properties of Semi-Solid Water-Based Phantoms.

Carlos Mendes1,2, Custódio Peixeiro3.   

Abstract

This paper presents a complete and detailed description of the fabrication and measurement of the electromagnetic properties of water-based semi-solid phantoms with emphasis on the analysis of the time evolution of the complex permittivity of several samples stored in different conditions. A known recipe for a 2/3 muscle equivalent phantom is used as test material, and the several phantom sample properties are measured with an in-house developed coaxial probe technique. It is shown that the storing condition is of paramount importance to extend the lifetime of a given phantom. This behavior stems from the way the storing condition affects the water evaporation rate of the sample. In particular, while an unprotected sample can preserve its electromagnetic properties only for a few days, a very well-sealed one can last at least up to a year.

Entities:  

Keywords:  2/3 muscle; BAN; coaxial probe technique; complex permittivity measurement; physical phantom

Year:  2019        PMID: 31590209      PMCID: PMC6806226          DOI: 10.3390/s19194298

Source DB:  PubMed          Journal:  Sensors (Basel)        ISSN: 1424-8220            Impact factor:   3.576


1. Introduction

When the performance of an antenna placed near the human body must be measured, it is customary to use a phantom to mimic the electromagnetic properties of the human body. An excellent survey of different types of phantoms and how they can be fabricated can be found in [1]. For a phantom to be realistic it must mimic the electromagnetic properties of the human tissues in the frequency range of interest. As the main component of most biological tissues is water, the use of water as the main component of a phantom tends to guarantee a good approximation to the electromagnetic behavior of the envisaged biological tissue. By adding certain components to the water, according to an appropriated recipe, the water properties can then be modified to meet those of the intended tissue. One problem with a water-based phantom is its conservation as some evaporation will necessarily happen. To minimize the water loss and extend the lifetime span of the phantom it must be properly stored and handled. Therefore, the goal of this paper is to describe, in a complete and detailed way, the fabrication and measurement of the electromagnetic properties of water-based semi-solid phantoms and to access their lifetime span for different storing conditions. The authors have recently presented a wearable fabric antenna [2,3] to work in the Industrial, Scientific and Medical (ISM) band at 2.45 GHz. The antenna was tested when placed on a homogeneous phantom mimicking a human torso with intended properties 2/3 of muscle, i.e., complex permittivity equal to 2/3 of that expected for muscle. At 2.45 GHz the muscle characteristics are and [4], where and are, respectively, the real and imaginary parts of the relative complex permittivity and is the loss tangent. Therefore, the 2/3 muscle equivalent tissue is characterized by and . This model has proven to be a good approximation for a layered inhomogeneous phantom with skin, fat, muscle and bone [5,6,7,8]. The material used in the fabrication of the torso phantom presented in [2,3] was also used in the samples analyzed in this work. The organization of the paper is as follows: in Section 2 the recipe and fabrication procedure are described. Section 3 describes in detail the coaxial open-ended measurement technique used. The presentation of the experimental results obtained and an analysis of the time degradation of the dielectric properties are presented in Section 4. Final conclusions are given in Section 5. An uncertainty analyses is presented in the Appendix A.

2. Homogeneous Phantom Fabrication

The human body is made of many different tissues with unique electromagnetic properties. The characterization of these tissues has already been thoroughly studied by several researchers and their characterizations at different frequencies are currently available. A comprehensive study of body tissue characterization is presented in [9] where the complex permittivity of various human tissues is measured for frequencies ranging from 10 Hz to 100 GHz and fitted to a 4 term Cole–Cole model. When performing measurements, the intricate details of the human body are often replaced by a homogeneous volume representing a part of the human body and with a permittivity of some kind of average behavior of the real material in the intended frequency range. In the literature there are several formulas [10,11,12] than can be used to produce phantoms mimicking the electromagnetic properties of biological tissues. The ingredients used and the fabrication procedure follow that described in [12]. This procedure was selected since it is based on cheap and easy to obtain ingredients, has a fabrication procedure which requires minimal equipment and allows easy adjustment of the electromagnetic properties of the phantom. Moreover, it leads to phantoms with mechanical (rheological) properties suitable for the envisaged applications [1,12]. Also, the phantom produced can take any desired shape since it will conform to the shape of the container where it will be stored. This was particularly important for the work presented by the authors, where the shape on a human torso was targeted to mimic the transmission between antennas placed on the torso. Deionized water (the dominant ingredient), agar, polyethylene powder, sodium chloride (NaCl), TX-151 and sodium azide are the ingredients used which allow the obtaining of a final material that simulates the characteristics of the high-water-content human tissues, such as muscle, brain or internal organs, although not so suitable for low-water-content tissues such as fat and bone. In this phantom, retention of the shape is made possible with agar which also prevents water from separating. The polyethylene powder is used to adjust the relative permittivity while the conductivity is mainly adjusted by the concentration of sodium chloride. Since the agar solution and the polyethylene powder cannot be mixed directly, TX-151 is used to increase the viscosity. Sodium azide is added as a preservative. Using different proportions of the ingredients, this recipe has successfully been used to realize different materials. In [12], muscle and brain equivalent materials for frequencies up to 2.5 GHz are created. In [13], the same recipe is used to obtain a head-equivalent phantom for the 3–6 GHz frequency range, a real-shaped phantom of the upper half of the body at 2.6 GHz and an abdomen phantom of a pregnant women at 150 MHz. In [14,15] a recipe for a material capable of representing muscle for frequencies up to 12 GHz with 2/3 muscle properties is used to represent a real-shaped phantom of the upper half of the body of an adult male. In [16,17] a skin equivalent realistic hand for 60 GHz is presented. A model for propagation around a cylinder of 2/3 muscle is proposed and tested in [18]. The fabrication procedure, adapted from [12], is as follows: Deionized water is placed on a kettle and heated on a gas burner. During the heating process, sodium chloride and sodium azide are added to the water. To prevent evaporation, the heating process must be performed with the kettle’s lid on. Once the water starts to reach the boiling point, agar is slowly added and dissolved. When the mixture reaches the boiling point, fire is extinguished immediately. Small quantities of TX-151 and polyethylene powder are sprinkled into the liquid several times and quickly mixed with an electrical mixer at a speed fast enough to assure a uniform mixture but not so fast that air bubbles are formed. After all ingredients are added, the mixture is poured into the container, which must be immediately closed (with a plastic lid or cling-film) to prevent evaporation. Finally, the mixture is allowed to completely cool down to room temperature (for about a day) to completely solidify. After that period, the phantom can be removed from the container but must be kept in a hermetical environment to reduce water evaporation. Photographs of the main steps of the phantom material fabrication are shown in Figure 1. These photographs detail the fabrication of the torso phantom used by the authors in [2,3], which is the same process used for the smaller samples.
Figure 1

Photographs of the several steps of the phantom’s fabrication. (a) Water heating. (b) Ingredient mixing. (c) Container filling.

Two batches of phantom material were fabricated, on two different occasions, with the quantities presented in Table 1. The smaller batch is directly obtained from [15] and aims at a material with 2/3 muscle properties in the range 2–9 GHz. This batch is used to produce several small test samples with different sizes and preservation conditions with the goal of measuring the degradation rate of the phantom’s properties with time for different storage conditions. The quantities of the larger batch are scaled from the smaller one to obtain a total of 12 L to build the torso phantom used in [3]. A small sample of this batch was also kept for measurements. The small samples and their respective containers/storing conditions are shown in Figure 2.
Table 1

Composition of the muscle equivalent material.

IngredientAmount (g)Amount (g)
(Original 4531.3 g)(Scaled 12,000 g)
Deionized water3375.08937.8
Agar104.6277.0
Polyethylene powder1012.62681.6
Sodium chloride7.018.5
TX-15130.179.7
Sodium azide2.05.3
Figure 2

Photograph of the samples inside the containers.

Sample 1 is approximately parallelepiped (average side 13.5 cm and average height 5.0 cm) and has been kept closed inside a plastic box. The air gap between the sample and the box cover is approximately 3.5 cm. Sample 2 and 3 are also approximately parallelepiped (average side 13.5 cm and average height 4.0 cm). Sample 2 has been kept wrapped in cling-film and sample 3 was left open without protection. Samples 4 and 5 are approximately cylindrical (average diameter 11 cm and average height 5.5 cm) and have been kept inside a plastic box with an air gap between sample and cover of 5 mm. Sample 6 is approximately parallelepiped with average side 13.5 cm and average height 6.5 cm and has been kept closed inside the plastic box for about one year. The air gap between the sample and the box cover is approximately 10 mm. The characteristics of the samples are summarized in Table 2.
Table 2

Summary of the characteristics of the several samples.

SampleShapeStorageAgeDimensionsAir Gap
(cm)(cm)
1parallelepipedplastic boxrecent13.5 × 13.5 × 53.5
2parallelepipedcling-filmrecent13.5 × 13.5 × 40
3parallelepipedopenrecent13.5 × 13.5 × 4-
4cylindricalplastic boxrecent11 × 5.50.5
5cylindricalplastic boxrecent11 × 5.50.5
6parallelepipedplastic box1 year13.5 × 13.5 × 61.0

3. Measurement Technique

The open-ended coaxial probe method [19,20,21,22,23] is used to measure the dielectric properties of the phantom. It is intended to get insight about the frequency ranges where the fabricated coaxial probes would provide useful results and where the fabricated phantom successfully mimics the human tissues. The equipment available at the time of the measurements limited the maximum usable frequency to 6 GHz.

3.1. Method Description

In simple media (linear, isotropic, homogeneous and time invariant) the complex permittivity is given by where is the vacuum permittivity, is the real part of the relative complex permittivity, is the imaginary part of the relative complex permittivity, accounting for dielectric polarization losses, is the DC conductivity, is the angular frequency and is the total imaginary part of the relative complex permittivity. The theory behind the open-ended coaxial method for measuring the electrical properties of materials is very well established [19,20,21,22,23]. It consists of the use of an open-ended coaxial probe immersed in the material under test (MUT) and to measure the reflection coefficient in a reference plane. Figure 3 shows the coaxial open-ended probe immersed in the MUT and the equivalent circuit of the coaxial-MUT interface [21]. The reflection coefficient at the open-ended side of the coaxial probe is given by where () is the load impedance (admittance) at the probe tip and () is the coaxial probe characteristic impedance (admittance). The load admittance of the equivalent circuit is [19,21] resulting on
Figure 3

Equivalent circuit at the coaxial-MUT interface [21], where , is the fringing capacity at the tip of the probe, is the probe terminal capacity when in free space and is the voltage reflection coefficient measured at the side of the probe connected to the vector network analyzer.

The coaxial probe can also be modeled as a two-port network, as shown in Figure 4, and therefore it can be written
Figure 4

Two-port model of the coaxial probe [21].

Equating (6) and (4) leads to where The complex coefficients , and can be obtained from Equation (7) by terminating the coaxial probe with 3 standard media with known and measure the corresponding reflection coefficient (i = 1, 2, 3) at the VNA reference plane. The most common choice for the 3 standards is a short-circuit, an open-circuit and a well-known reference media. For the short-circuit (), . Substituting in (6) gives Solving for leads to For the open-circuit (), . Substituting in (7) yields Finally, substituting again in (7) for a reference liquid with results on Rearranging for the it can finally be obtained Once the are obtained from measurements with the reference loads, the coaxial probe can be loaded with the MUT. Solving for

3.2. Coaxial Probe Description

A coaxial probe, shown in Figure 5, has been fabricated from the semi-rigid coaxial cable UT141. The probe has a length of approximately 20 cm and is left open on one end and terminated with an SMA male connector on the other end to allow for the connection with a VNA. The measured input reflection coefficient of the open-ended probe is shown in Figure 6. An unexpected high ripple is noticed. The SMA connector can introduce small reflections and therefore some ripple was expected, but much smaller than the one obtained since the SMA connector is specified to be used up to 18 GHz. After a detailed analysis of the fabrication process it was concluded that a discontinuity had been introduced when soldering the cable section to the SMA connector. The high temperature needed to solder the cable to the connector led to the expansion of the dielectric insulator. After cooling, the dielectric retracted and an air gap was left between the dielectrics of the connector and of the cable. This air gap corresponds to a discontinuity in the cable characteristic impedance.
Figure 5

Photograph of the fabricated coaxial probe.

Figure 6

Input reflection coefficient of the coaxial probe.

A simple transmission line circuit model can be used to check the effect of the air gap of the coaxial probe. The circuit model used is shown in Figure 7, where is the length of the probe and is the length of the air gap perturbation. As is very small (about 1 mm), and for the sake of simplicity, the corresponding section is considered lossless. Moreover, for the open-ended coaxial will be assumed. Therefore, the impedance at the input of section 1 is given by with
Figure 7

Transmission line model of the probe with air gap.

The impedance at the input of the lossless section 2 can be obtained from Finally, the reflection coefficient at the input of section 2 is The amplitude of as a function of frequency is shown in Figure 6. The following data values have been used: The attenuation coefficient in section 1 of the cable has been obtained from typical UT141 data sheets and is reproduced in Table 3. The interpolation function indicated in Table 4 has been obtained from the data shown in Table 3.
Table 3

UT141 typical attenuation.

Frequency (GHz)Attenuation Coefficient (dB/m)
0.50.26
10.38
50.91
101.37
202.09
Table 4

UT141 cable interpolated attenuation.

Frequency (GHz)Attenuation Coefficient (dB/m)
≤5 0.047184f0.003434
[1,5] 0.049366f0.005616
[5,10] 0.057179f0.023086
≥10 0.063282f0.042385
The length of the air gap ( mm) has been adjusted to reproduce the amplitude of the ripple obtained in the open-ended experimental response. The circuit model results are compared with measurements in Figure 6. Simulation results for the case without air gap () have also been introduced for reference. It seems that the attenuation is slightly underestimated above approximately 10 GHz. The adopted model of the air gap allows a proper reproduction of the high ripple obtained in the amplitude of the input reflection coefficient. The above-mentioned discrepancies are quite limited and, at least partially, have a physical explanation.

3.3. Coaxial Probe Calibration

For the probe calibration, besides the open-circuit and short-circuit, a third reference well-known media is needed. The use of at least a fourth reference well-known media is also advisable for the testing of the probe and calibration procedure. Although the authors worked with several well documented reference liquids (deionized water, methanol, ethanol, dimethyl sulfoxide (DMSO) and Water saline solutions) the results provided here were obtained with a calibration with deionized water. As methanol has permittivity values closed to the target phantom values, the uncertainty of methanol measurement will be calculated to obtain an estimate of the uncertainty of the measured phantom’s permittivity. An exchange of the role of these two liquids showed no relevant differences in the permittivity measurements. The setup used in the measurements is shown in Figure 8a for the short-circuit and in Figure 8b for one of the reference liquids. A metallic holder was used to control the alignment and depth of the probe inside the liquids. During the measuring process it was noticed that the penetration depth of the probe was not a relevant parameter [24]. In fact, water measurements with a penetration of 20 mm, 40 mm and 60 mm are indistinguishable and only slightly different from the measurements with 0 mm. Methanol and ethanol provide experimental input reflection coefficient amplitude results that are also almost independent of the probe depth. However, a type of hysteresis effect was detected, i.e., for the same depth, the results obtained depended on whether the probe had just been inserted into the liquid or had stayed there for some time. As methanol and ethanol are less viscous than the other reference liquids used, and as the coaxial probe end is not sealed, some liquid could go inside the probe between the conductors (inner and outer) and the PTFE dielectric as was also identified in [20]. From the analysis of the results obtained it was decided to use a probe depth of 20 mm for all the reference liquids. Additionally, for methanol and ethanol measurements, special care was taken to make sure that the probe stays a short period of time inside the liquids to minimize infiltration into the probe.
Figure 8

Experimental setup used for the measurements. The photographs show two examples, the short-circuit and a reference liquid. (a) Short-circuit. (b) Liquid.

The coaxial open-ended coefficients , and , defined in Equations (8)–(10), respectively, have been obtained and are shown in Figure 9a–c. They have been measured in an environment with a stable temperature of 20.6 °C and 48% humidity.
Figure 9

coefficients using deionized water as reference. (a) A1. (b) A2. (c) A3.

4. Phantom Measurements

4.1. Comparison between Measured and Target Permittivity

The ultimate purpose is to measure the complex permittivity of the phantom material described previously. However, to test the repeatability of the fabrication process and also the way the permittivity changes with time with different storing conditions, six small samples of the phantom material, as shown in Figure 2, have also been fabricated. Samples 1 to 5 have been fabricated recently while sample 6 has been fabricated about one year before the current measurements. All the samples have been kept on a room with controlled temperature and humidity. In the first 18 days after fabrication the room temperature was °C. On day 22 it was necessary to increase the temperature about 2 °C. The humidity of the room was in the range . To measure the complex permittivity each sample was placed on the top of a photography tripod and moved up until the probe was about 9 mm deep inside the sample (Figure 10). It was necessary to move the sample smoothly, without going back and forth, to make sure that the sample material was in direct contact with the coaxial probe end. If there is a small gap between probe and sample material, as shown in Figure 11, either filled with air or water solution, the complex permittivity results will be substantially different [24].
Figure 10

Photographs of the measurement setup used to measure the phantom samples using the sample left open as an example.

Figure 11

Accidental gap between the coaxial probe tip and the phantom material.

The measured complex permittivity superimposed with the expected 2/3 muscle properties [4] are shown in Figure 12. All the results have been obtained 2 days after fabrication when the temperature of the samples was almost the same as the room temperature.
Figure 12

Measured and expected relative permittivity of the phantom samples. (a) Sample 1 (Parallelliped, Plastic). (b) Sample 2 (Parallelliped, Cling-film). (c) Sample 3 (Parallelliped, Open). (d) Sample 4 (Cilindrycal, Plastic). (e) Sample 5 (Cilindrycal, Plastic). (f) Sample 6 (Parallelliped, Plastic, one year).

For the analysis of the results, it must be kept in mind that the recipe used in the phantom’s fabrication will produce a material with properties close to 2/3 muscle in the 2–9 GHz range. Therefore, the difference observed between the measured values and the target ones for frequencies below 2 GHz (specially noted in the values of ) is expected. There is a general very good agreement for the imaginary part of the relative complex permittivity of the six samples. The value is in the range 7.09–7.41 and is in the range 0.181–0.190. For the real part of the relative complex permittivity of the six samples the agreement is good between samples 2, 4, 5 and 6. Samples 1 and 3 have lower values of meaning that they have lost water by evaporation. Sample 1, which is closed in a plastic box, lost some (little) water to the air volume (3.5 cm height) above the phantom material just until it got saturated. When the box is open, the water-saturated air is replaced by fresh air that, after the box is closed, becomes saturated again. This is the process by which the sample loses water. Sample 3, which is in a completely open situation, lost (more) water continuously. It is amazing that sample 6, despite being one year old, has preserved its macroscopic dielectric characteristics unchanged. This is because it was kept in a hermetically closed box that was never open in the one-year time period. The uncertainty analysis of the results obtained is described in Appendix A.

4.2. Time Evolution of the Phantom’s Characteristics

To evaluate the time evolution of the phantom dielectric characteristics, the several phantom samples were periodically measured. Regarding all the measurements, it is important to point out that the samples that are kept inside plastic boxes (1, 4, 5 and 6) are briefly open for the measurements and are closed hermetically afterwards and sample 2 is unwrapped for the measurements and is carefully wrapped immediately after. The time evolution of the experimental results are shown in Figure 13. The results shown indicate clear tendencies that depend on the type of protection of each of the phantom samples. As time goes by, if unprotected, the phantom samples rapidly lose water content and the relative permittivity goes down quickly. This effect can be observed in sample 3 where 11 days after fabrication went down from about 40 to about 20 and from about 7 to about 4. However, there are some results that seem to be erroneous and do not follow any tendency. These results differ from the general tendency far above the uncertainty level of the measurement method. An example is, for instance, of sample 1 on days 7, 9, 11, 18 and 44. Most of the “erroneous” results are below the tendency lines but there are also a few above. One possible explanation for the erroneous results is the fact that the coaxial probe tip might not be in direct contact with the sample material because of a small cylindrical gap, as depicted in Figure 11. This gap can be filled either with air, in which case the measured relative permittivity is lower than expected, or with water, in which case the measured relative permittivity is higher than expected. There are other factors contributing to the non-repeatability errors, such as the holes left by the previous measurements and the proximity of the measurement point to the phantom borders.
Figure 13

Time evolution of the phantom samples measured relative permittivity. (a) Sample 1 (Parallelliped, Plastic). (b) Sample 2 (Parallelliped, Cling-film). (c) Sample 3 (Parallelliped, Open). (d) Sample 4 (Cilindrycal, Plastic). (e) Sample 5 (Cilindrycal, Plastic). (f) Sample 6 (Parallelliped, Plastic, one year).

It should be pointed out that during the whole process, despite the water evaporation, the mechanical properties of the phantom are kept, i.e., the samples kept their shape and did not fall apart. The sample that was left open rapidly became noticeable dry. However, the remaining samples kept their original appearance during the whole process, which means that a visual inspection of the phantom is not enough to evaluate its current electromagnetic condition. To be able to take some useful conclusions concerning the time evolution of the relative complex permittivity of the phantoms, the measured relative complex permittivity is fit to exponential tendency lines of the form where n is the number of days since the fabrication. The same values of and are used in samples 1 to 5 and have been chosen as the average of the corresponding permittivity of samples 1, 2, 4 and 5 in day 2. Sample 6, being one year old, has been given a slightly different approach. It is quite natural that have both decreased a little during the one-year lifetime. In this case, and have been used in the data fitting process. The values of , obtained using a least-squares error minimization procedure, are indicated in Table 5.
Table 5

Time constants of the permittivity tendency lines.

Sample c1 c2
10.00390.0051
20.00500.0057
30.07000.0640
40.00100.0014
50.00160.0020
60.00360.0048
The experimental results are compared with the fitted tendency lines of each sample in Figure 13. The following main conclusions can be drawn: Except for sample 3 (unprotected) the time decaying of the permittivity is almost linear. The real part and the imaginary part of the permittivity have similar time decaying rates although a slightly faster decay of the imaginary part is noticed. The useful lifetime of an unprotected phantom is very small (a few days only). The most effective protection is a hermetic sealing box that can extend a phantom lifetime many months or even a year. This long extension of the lifetime is only possible if the volume of air is contact with the phantom is small and is not renewed. Wrapping a phantom in plastic film can also extend the useful lifetime of a phantom, not as much as a hermetic box, but it can reach up to a month. Although these conclusions are obtained from the measurements of small samples of phantom material, we strongly believe that they can be qualitatively extrapolated to larger volumes such as the fabricated torso phantom.

5. Conclusions

A complete and detailed description of the fabrication and measurement of the electromagnetic properties of water-based semi-solid phantoms with emphasis on the analysis of the time evolution of the complex permittivity of several samples stored in different conditions is presented. It is shown that the storing conditions of a water-based phantom can largely influence the lifetime of the sample. If a sample is kept unprotected, its electromagnetic properties will last only for a few days. However, if hermetically stored, they will be preserved for at least a year. A sample that is kept hermetically closed and open regularly (in a daily basis) for a quick measurement can last up to a month. The phantom material is very lossy. Consequently, only the part near its surface has influence on the propagation over the phantom. Therefore, since the water evaporation rate is the main effect driving the time change of the permittivity and it is almost independent of the sample size, the conclusions obtained from the measurements of the small samples can be extrapolated to larger ones.
Table A1

, and values and uncertainty for methanol [26].

T (°C) εs±Δεs ε±Δε τ±Δτ (ps)
10 35.74±0.06 5.818±0.104 2.262±0.019
15 34.68±0.06 5.698±0.131 2.532±0.023
20 33.64±0.06 5.654±0.104 2.822±0.019
25 32.66±0.06 5.563±0.176 3.141±0.033
30 31.69±0.06 5.450±0.204 3.490±0.204
Table A2

Parameters of the uncertainty tendency lines.

Parameter Δεr Δεr Δtanδ
A1.02503.43023.717
B0.33980.27010.293
  5 in total

1.  Effect of pierced metallic objects on SAR distributions at 900 MHz.

Authors:  Jose Fayos-Fernandez; Carlos Arranz-Faz; Antonio M Martinez-Gonzalez; David Sanchez-Hernandez
Journal:  Bioelectromagnetics       Date:  2006-07       Impact factor: 2.010

2.  Formulas for preparing phantom muscle tissue at various radiofrequencies.

Authors:  C K Chou; G W Chen; A W Guy; K H Luk
Journal:  Bioelectromagnetics       Date:  1984       Impact factor: 2.010

3.  Open-Ended Coaxial Dielectric Probe Effective Penetration Depth Determination.

Authors:  Paul M Meaney; Andrew P Gregory; Jan Seppälä; Tapani Lahtinen
Journal:  IEEE Trans Microw Theory Tech       Date:  2016-01-29       Impact factor: 3.599

4.  Electromagnetic wave propagation in Body Area Networks using the Finite-Difference-Time-Domain method.

Authors:  Jonathan N Bringuier; Raj Mittra
Journal:  Sensors (Basel)       Date:  2012-07-23       Impact factor: 3.576

Review 5.  Open-Ended Coaxial Probe Technique for Dielectric Measurement of Biological Tissues: Challenges and Common Practices.

Authors:  Alessandra La Gioia; Emily Porter; Ilja Merunka; Atif Shahzad; Saqib Salahuddin; Marggie Jones; Martin O'Halloran
Journal:  Diagnostics (Basel)       Date:  2018-06-05
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.