William N Duggar1, Bart Morris1, Ali Fatemi1,2, Jemeria Bonds1, Rui He1, Madhava Kanakamedala1, Roberto Rey-Dios3, Srinivasan Vijayakumar1, Claus Yang1.
Abstract
OBJECT: The purpose of this study was to compare two methods of stereotactic localization in Gamma Knife treatment planning: cone beam computed tomography (CBCT) or fiducial. While the fiducial method is the traditional method of localization, CBCT is now available for use with the Gamma Knife Icon. This study seeks to determine whether a difference exists between the two methods and then whether one is better than the other regarding accuracy and workflow optimization.Entities:
Keywords: CBCT; Gamma Knife; frame-based; icon; radiosurgery workflow
Mesh:
Year: 2019 PMID: 31587520 PMCID: PMC6839378 DOI: 10.1002/acm2.12745
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Phantoms utilized in this study. Top — Elekta Film Phantom Bottom Left — Elekta 8 cm radius Solid Water Phantom Bottom Right — Anthropomorphic Head Phantom.
Agreement between cone beam computed tomography defined radiological focus and film test (deviation vector is the average of the three‐dimensional vector of each independent measurement).
| Coordinate | ∆X | ∆Y | ∆Z |
|---|---|---|---|
| 100, 100, 100 | −0.07 ± 0.03 mm | 0.02 ± 0.17 mm | −0.1 ± 0.17 mm |
| Deviation vector | 0.29 ± 0.06 mm | ||
Difference from expected phantom positions based on co‐registration with reference cone beam computed tomography of film pinprick phantom at stereotactic center (100, 100, 100).
| Rotation (°) | Translation (mm) | |||||
|---|---|---|---|---|---|---|
| X | Y | Z | X | Y | Z | |
| Position 100, 100 | ||||||
| Registration 1 | −0.02 | 0.01 | 0.01 | −0.02 | 0.01 | 0.01 |
| Registration 2 | 0.00 | 0.01 | −0.02 | −0.02 | 0.00 | 0.02 |
| Registration 3 | 0.00 | 0.00 | −0.04 | −0.02 | 0.00 | 0.01 |
| Average | −0.01 | 0.01 | −0.02 | −0.02 | 0.00 | 0.01 |
| Position 100, 100 R90 | ||||||
| Registration 1 | −0.01 | −0.08 | −0.46 | 0.00 | −0.08 | 0.02 |
| Registration 2 | −0.05 | −0.06 | 0.21 | 0.01 | −0.09 | 0.02 |
| Registration 3 | −0.05 | 0.00 | −0.38 | 0.04 | −0.07 | 0.05 |
| Average | −0.04 | −0.05 | −0.21 | 0.02 | −0.08 | 0.03 |
| Position 160, 160 | ||||||
| Registration 1 | −0.03 | 0.05 | 0.27 | 0.11 | 0.14 | −0.02 |
| Registration 2 | −0.02 | 0.03 | −0.09 | 0.09 | 0.14 | 0.04 |
| Registration 3 | 0.00 | 0.02 | 0.25 | 0.09 | 0.14 | 0.04 |
| Average | −0.02 | 0.03 | 0.14 | 0.10 | 0.14 | 0.02 |
| Position 160, 40 | ||||||
| Registration 1 | 0.00 | 0.02 | −0.03 | 0.09 | −0.10 | 0.00 |
| Registration 2 | −0.04 | −0.04 | 0.15 | 0.14 | −0.10 | 0.03 |
| Registration 3 | −0.02 | 0.02 | 0.16 | 0.08 | −0.07 | 0.01 |
| Average | −0.02 | 0.00 | 0.09 | 0.10 | −0.09 | 0.01 |
| Position 40, 40 | ||||||
| Registration 1 | −0.02 | −0.04 | −0.24 | −0.06 | −0.04 | 0.00 |
| Registration 2 | 0.00 | −0.02 | 0.12 | −0.06 | −0.04 | −0.01 |
| Registration 3 | −0.02 | −0.01 | 0.23 | −0.06 | −0.05 | −0.05 |
| Average | −0.01 | −0.02 | 0.04 | −0.06 | −0.04 | −0.02 |
| Position 40, 160 | ||||||
| Registration 1 | −0.03 | −0.01 | 0.06 | −0.10 | 0.05 | −0.02 |
| Registration 2 | −0.02 | 0.00 | 0.22 | −0.11 | 0.06 | −0.03 |
| Registration 3 | −0.02 | 0.00 | 0.10 | −0.10 | 0.05 | −0.03 |
| Average | −0.02 | 0.00 | 0.03 | −0.10 | 0.05 | −0.03 |
Magnetic resonance imaging to cone beam computed tomography registration results (mm).
| Frame‐based | Frameless | Chung et al. | |
|---|---|---|---|
| Deviation in X | 0.4 ± 0.3 | 0.3 ± 0.3 | 0.1 ± 0.2 |
| Deviation in Y | 0.3 ± 0.3 | 0.3 ± 0.2 | 0.3 ± 0.3 |
| Deviation in Z | 0.3 ± 0.3 | 0.3 ± 0.2 | 0.3 ± 0.3 |
| 3D deviation | 0.7 ± 0.3 | 0.6 ± 0.2 | 0.4 ± 0.2 |
Difference between fiducial localization and cone beam computed tomography (CBCT) localization on the Ball phantom.
| CTDI 2.5 | CTDI 6.3 | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Rotation (°) | Translation (mm) | Rotation (°) | Translation (mm) | ||||||||||
| X | Y | Z | X | Y | Z | X | Y | Z | X | Y | Z | ||
| CBCT 1 | 0.46 | −0.12 | 0.41 | −0.18 | −0.34 | −0.94 | CBCT 1 | 0.51 | −0.02 | −0.08 | −0.13 | −0.19 | −0.64 |
| CBCT 2 | 0.64 | 0.11 | −0.03 | −0.13 | −0.27 | −0.7 | CBCT 2 | 0.24 | 0.08 | −0.26 | −0.17 | −0.26 | −0.54 |
| CBCT 3 | 0.13 | −0.31 | 0.02 | 0.08 | −0.04 | −0.81 | CBCT 3 | 0.51 | 0.15 | 0.07 | −0.12 | −0.28 | −0.79 |
| Average | 0.41 | −0.11 | 0.13 | −0.08 | −0.22 | −0.82 | Average | 0.42 | 0.07 | −0.09 | −0.14 | −0.24 | −0.66 |
Difference between fiducial localization and cone beam computed tomography (CBCT) localization for anthropomorphic phantom.
| Rotation (°) | Translation (mm) | |||||
|---|---|---|---|---|---|---|
| X | Y | Z | X | Y | Z | |
| MRI Sim with CBCT | ||||||
| CBCT 1 | 0.12 | −0.17 | 0.16 | 0.61 | −0.17 | −0.31 |
| CBCT 2 | 0.25 | −0.29 | 0.16 | 0.71 | −0.13 | −0.3 |
| CBCT 3 | 0.22 | −0.3 | 0.13 | 0.7 | −0.12 | −0.38 |
| Average | 0.197 | −0.253 | 0.150 | 0.673 | −0.140 | −0.330 |
| CT Sim with CBCT | ||||||
| CBCT 1 | 0.11 | 0.06 | 0.13 | 0.24 | −0.44 | −0.18 |
| CBCT 2 | 0.07 | 0.01 | 0.1 | 0.27 | −0.42 | −0.2 |
| CBCT 3 | 0.06 | 0.05 | 0.09 | 0.26 | −0.35 | −0.08 |
| Average | 0.080 | 0.040 | 0.107 | 0.257 | −0.403 | −0.153 |
Abbreviation: MRI, magnetic resonance imaging.
Difference between expected phantom movement and detected phantom movement.
| Expected shifts (CBCT postshift registered to reference CBCT) | Expected — measured difference | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Rotation (°) | Translation (mm) | Rotation (°) | Translation (mm) | |||||||||
| X | Y | Z | X | Y | Z | X | Y | Z | X | Y | Z | |
| MRI sim with CBCT | ||||||||||||
| −0.01 | 0.05 | −0.07 | 0.03 | −0.56 | 0.14 | 0.09 | 0.15 | 0.07 | −0.04 | 0.01 | 0.04 | |
| 0.02 | −0.01 | −0.02 | 0.02 | 0.03 | 0.00 | |||||||
| −0.01 | −0.00 | −0.02 | −0.02 | 0.11 | 0.15 | |||||||
| Average Difference | 0.03 | 0.04 | 0.01 | −0.02 | 0.05 | 0.06 | ||||||
| CT sim with CBCT | ||||||||||||
| −0.01 | 0.05 | −0.07 | 0.03 | −0.56 | 0.14 | −0.06 | 0.06 | −0.10 | −0.08 | −0.03 | 0.02 | |
| 0.06 | −0.26 | −0.12 | 0.05 | 0.00 | 0.05 | |||||||
| 0.07 | −0.09 | −0.08 | 0.03 | 0.03 | −0.05 | |||||||
| Average Difference | 0.02 | −0.10 | −0.10 | −0.00 | 0.00 | 0.01 | ||||||
Abbreviations: CT, computed tomography; CBCT, cone beam CT.
Figure 2A comparison of fiducial localization to pretreatment cone beam computed tomography localization for patients separated based on primary dataset used for localization. Top — magnetic resonance imaging, Bottom — computed tomography.
A comparison of fiducial localization to pretreatment CBCT localization for actual patients with a mix of primary stereotactic reference image types (MRI and CT).
| Patient site | X (mm) | Y (mm) | Z (mm) | X rotation (°) | Y rotation (°) | Z rotation (°) | 3D translation (mm) | Primary stereotactic reference |
|---|---|---|---|---|---|---|---|---|
| 1 | −0.1 | 0.06 | 0.05 | −0.19 | 0.02 | 0.05 | 0.13 | CT |
| 2 | −0.23 | −0.17 | 0.19 | 0.16 | 0.56 | −0.05 | 0.34 | CT |
| 3 | −0.85 | 0.25 | 0.47 | −0.07 | 0.23 | −0.15 | 1.00 | CT |
| 4 | −0.01 | 0.22 | 0.23 | −0.01 | 0.53 | −0.11 | 0.32 | CT |
| 5 | −0.15 | 0.35 | −0.07 | 0.75 | 0.11 | −0.1 | 0.39 | MRI |
| 6 | 0.05 | 0.2 | −0.1 | −0.49 | 0.24 | 0.63 | 0.23 | MRI |
| 7 | 0.45 | −0.21 | 0.08 | 0 | −0.02 | −0.19 | 0.50 | CT |
| 8 | 0.15 | −0.04 | 0.15 | 0.09 | 0.25 | −0.04 | 0.22 | CT |
| 9 | 0.1 | −0.39 | −0.37 | 0.45 | −0.12 | 0.17 | 0.55 | MRI |
| 10 | 0.27 | 0 | 0.7 | 0.61 | 0.34 | 0.39 | 0.75 | MRI |
| 11 | 0.57 | −0.54 | −0.24 | 0.91 | 0.12 | 0.49 | 0.82 | MRI |
| 12 | 0.06 | −0.11 | 0.3 | 0.15 | −0.02 | 0.02 | 0.33 | CT |
| 13 | −0.19 | −0.16 | 0.15 | 0.77 | −0.43 | −0.03 | 0.29 | MRI |
| 14 | −0.28 | −0.03 | 0.27 | 0.49 | 0.08 | 0.17 | 0.39 | MRI |
| 15 | −0.04 | −0.03 | 0.34 | −0.01 | 0.13 | −0.01 | 0.34 | CT |
| 16 | −0.2 | −0.06 | −0.17 | −0.01 | 0.35 | −0.08 | 0.27 | CT |
| 17 | 0.14 | 0.17 | 0.2 | 0.1 | 0.38 | 0.04 | 0.30 | CT |
| 18 | −0.5 | 0.08 | −0.2 | −0.5 | 0.08 | −0.2 | 0.54 | MRI |
| 19 | 0.49 | −0.11 | −0.42 | −0.4 | −0.2 | 0.06 | 0.66 | CT |
| 20 | −0.35 | 0.11 | 0.12 | 0.07 | 0.16 | −0.18 | 0.39 | CT |
| 21 | −1.21 | 0.24 | 0.68 | 0.81 | −0.64 | −0.16 | 1.41 | MRI |
| 22 | −1.27 | 0.09 | 0.87 | 0.39 | −0.41 | −0.33 | 1.54 | CT |
| 23 | 0.41 | −0.44 | 0.48 | 0.15 | 0.1 | 0.03 | 0.77 | CT |
| 24 | 0.17 | 0.22 | 0.16 | −0.16 | 0.42 | 0.05 | 0.32 | CT |
| 25 | −0.14 | −0.75 | −0.26 | 0.66 | −0.14 | −0.12 | 0.81 | MRI |
| 26 | −0.28 | −0.06 | −0.08 | 0.47 | −0.14 | 0.24 | 0.30 | MRI |
| 27 | 0.34 | 0.19 | 0.04 | 0.39 | 0.24 | 0.38 | 0.39 | MRI |
| 28 | −0.08 | 0.36 | −0.12 | −0.48 | 0.51 | −0.15 | 0.39 | CT |
| 29 | 0.29 | 0.16 | 0.29 | −0.21 | 0.56 | 0.03 | 0.44 | CT |
| CT average (absolute) | −0.06 | 0.03 | 0.19 | −0.03 | 0.21 | −0.06 | 0.49 | |
| CT maximum (absolute) | 0.49 | 0.36 | 0.87 | 0.39 | 0.56 | 0.06 | 1.54 | |
| MRI average (absolute) | −0.12 | −0.07 | 0.04 | 0.44 | −0.02 | 0.16 | 0.57 | |
| MRI maximum (absolute) | 0.57 | 0.35 | 0.70 | 0.91 | 0.34 | 0.63 | 1.41 | |
| Significance between MRI and CT ( | 0.72 | 0.32 | 0.23 | 0.00 | 0.04 | 0.01 | 0.51 | |
| Total average (absolute) | 0.11 | 0.01 | 0.16 | 0.32 | 0.25 | 0.16 | 0.52 | |
| Total maximum (absolute) | 1.27 | 0.54 | 0.87 | 0.91 | 0.56 | 0.63 | 1.54 |
Abbreviations: CT, computed tomography; CBCT, cone beam CT; MRI, magnetic resonance imaging.