| Literature DB >> 31585965 |
Bingyao Tan1,2, Jacqueline Chua1,3, Thiyagrajan Harish1, Amanda Lau1, Alfred Tau Liang Gan1, Yar Li Tan1,4, Damon W K Wong2,5, Rachel Shujuan Chong1,3, Marcus Ang1,3, Rahat Husain1,4, Leopold Schmetterer6,3,7,8,9,10.
Abstract
BACKGROUND/AIMS: To assess the agreement in measuring retinal nerve fibre layer (RNFL) thickness between spectral-domain (SD; Cirrus HD, Carl Zeiss Meditec, USA) optical coherence tomography (OCT) and swept-source (SS; Plex Elite 9000, Carl Zeiss Meditec) OCT using an OCT angiography (OCTA) scanning protocol.Entities:
Keywords: imaging; retina
Mesh:
Year: 2019 PMID: 31585965 PMCID: PMC7361033 DOI: 10.1136/bjophthalmol-2019-314706
Source DB: PubMed Journal: Br J Ophthalmol ISSN: 0007-1161 Impact factor: 4.638
Figure 1Example of circular circumpapillary RNFL extraction from a 6×6 mm volumetric scan obtained from SS-OCT using an angiography protocol. (A) The boundary of the optic disc was labelled manually (white line), and the centre of the optic disc (white dot) was defined by fitting the optic disc boundary with an ellipse. Three concentric scans with 3.44 mm, 3.46 mm and 3.48 mm diameters from the optic disc centre were extracted and averaged into (C). The non-averaged single circular scan from 3.46 mm is also displayed (B). Zoomed in features showing lower noise and better RNFL boundary are shown in figure parts E and G), compared with the non-averaged images (D and F). The global average, as well as quadrant average, was calculated. OCT, optical coherence tomography; RNFL, retinal nerve fibre layer; SS, swept-source.
Characteristics of participants
| Total | Glaucoma | OHT | Normal | |
| No. of participants | 57* | 9 | 5 | 45 |
| No. of eyes | 94 | 12 | 8 | 74 |
| Age, years (SD) | 62.98 (7.38) | 61.78 (9.31) | 56.25 (4.26) | 63.44 (6.94) |
| Gender, M/F | 34/23* | 8/1 | 4/1 | 23/22 |
| Visual field, dB | −2.38 (3.07) | −5.75 (3.22) | −1.28 (3.24) | −1.91 (3.18) |
*Two patients have one eye with glaucoma and the other eye with OHT.
Comparison of RNFL thickness measured by two OCT systems, stratified by glaucoma status
| Region | SS-OCT | SD-OCT | Mean difference (95% CI)* | ICC (95% CI)† | |||
| Mean (SD) | COV | Mean (SD) | COV | ||||
| Total | Inferior | 114.1 (24.3) | 4.7 | 113.0 (22.4) | 5.05 | 1.1 (−0.7 to 3.0) | 0.92 (0.88 to 0.95) |
| Superior | 106.9 (22.3) | 4.79 | 107.0 (20.3) | 5.26 | −0.1 (−1.9 to 1.6) | 0.91 (0.87 to 0.94) | |
| Nasal | 69.7 (11.5) | 6.04 | 66.3 (9.3) | 7.12 |
| 0.73 (0.64 to 0.82) | |
| Temporal | 75.6 (13.7) | 5.52 | 67.9 (12.3) | 5.51 |
| 0.79 (0.71 to 0.85) | |
| Global | 91.6 (13.5) | 6.78 | 88.6 (12.5) | 7.08 |
| 0.91 (0.86to 0.94) | |
| Glaucoma | Inferior | 77.8 (20.9) | 3.72 | 77.8 (21.0) | 3.7 | −0.0 (−4.6 to 4.6) | 0.95 (0.89 to 0.98) |
| Superior | 81.3 (23.8) | 3.42 | 80.6 (21.2) | 3.8 | 0.7 (−3.5 to 4.9) | 0.93 (0.78to 0.98) | |
| Nasal | 65.0 (13.9) | 4.68 | 63.5 (11.6) | 5.47 | 1.5 (−4.7 to 7.7) | 0.76 (0.41 to 0.92) | |
| Temporal | 63.7 (15.2) | 4.18 | 56.5 (13.7) | 4.13 |
| 0.81 (0.65 to 0.92) | |
| Global | 71.9 (14.5) | 4.97 | 69.6 (13.6) | 5.13 |
| 0.95 (0.89 to 0.98) | |
| OHT | Inferior | 122.8 (22.1) | 5.56 | 119.1 (18.2) | 6.54 | 3.7 (−4.6 to 11.9) | 0.88 (0.13 to 0.92) |
| Superior | 109.0 (13.2) | 8.28 | 108.3 (16.4) | 6.6 | 0.7 (−6.8 to 8.2) | 0.88 (0.64 to 0.96) | |
| Nasal | 70.4 (9.6) | 7.32 | 65.8 (9.9) | 6.67 | 4.6 (−7.1 to 16.3) | 0.31 (0.00 to 0.61) | |
| Temporal | 79.1 (22.3) | 3.55 | 73.2 (19.0) | 3.85 |
| 0.94 (0.82 to 0.98) | |
| Global | 95.3 (12.3) | 7.76 | 91.6 (10.0) | 9.12 | 3.7 (−1.0 to 8.4) | 0.82 (0.67 to 0.92) | |
| Normal | Inferior | 119.1 (19.7) | 6.03 | 118.1 (17.5) | 6.74 | 1.0 (−1.2 to 3.2) | 0.87 (0.81 to 0.92) |
| Superior | 110.8 (20.2) | 5.48 | 111.2 (17.3) | 6.41 | −0.4 (−2.5 to 1.8) | 0.88 (0.83 to 0.91) | |
| Nasal | 70.4 (11.3) | 6.23 | 66.9 (8.9) | 7.51 |
| 0.76 (0.68 to 0.84) | |
| Temporal | 77.2 (11.4) | 6.79 | 69.2 (10.2) | 6.77 |
| 0.71 (0.61 to 0.79) | |
| Global | 94.4 (10.7) | 8.86 | 91.3 (9.7) | 9.41 |
| 0.86 (0.79 to 0.90) | |
*The sandwich estimator that allows for clustering for patient was used to calculate SEs.
†Non-parametric cluster-resampled bootstrapping was used to derive the sampling distribution of ICC estimates.
COV, coefficient of variance; ICC, intraclass correlation coefficient; OCT, optical coherence tomography; SS, swept-source.
Figure 2Scatter plots (A–E) with linear regression analysis and Bland-Altman plots (F–J) of circumpapillary RNFL thickness measurements by SD-OCT and SS-OCT using an angiography protocol. Black square: glaucoma participants. Red circle: normal participants. Blue triangle: OHT participants. OCT, optical coherence tomography; RNFL, retinal nerve fibre layer; SD, spectral-domain.