| Literature DB >> 31581906 |
Abstract
OBJECTIVES: Cyanoacrylate glue is injected for incompetent great saphenous vein (GSV) treatment 5 cm distal to the saphenofemoral junction (SFJ). Although a few reports have investigated the postprocedural remnant stump length, none have focused on the factors affecting glue extension length and the consequent remnant stump length.Entities:
Keywords: GSV; VenaSeal; cyanoacrylate closure; great saphenous vein; stump length; varicose veins
Mesh:
Substances:
Year: 2019 PMID: 31581906 PMCID: PMC6900567 DOI: 10.1177/1538574419879563
Source DB: PubMed Journal: Vasc Endovascular Surg ISSN: 1538-5744 Impact factor: 1.089
Patient Demographics.
| No. of patients | 50 |
|---|---|
| Female, no (%) | 40 (80%) |
| Age, mean (±SD) | 47.4 ± 13.8 |
| BMI, mean (±SD) | 22.7 ± 3.4 |
| CEAP no. | |
| C1 | 2 |
| C2 | 25 |
| C3 | 19 |
| C4 | 3 |
| C5 | 1 |
Abbreviations: BMI, body mass index; CEAP, clinical etiologic anatomic pathophysiologic classification; SD, standard deviation.
Figure 1.Association between the GSV diameter and glue propagation length. GSV indicates great saphenous vein.
Treatment Characteristics.
| No. of treated GSV | 79 |
|---|---|
| GSV diameter in supine position (cm) mean ± SD (range) | 0.61 ± 0.24 (0.30-1.63) |
| Glue extension length (cm) mean ± SD (range) | 1.13 ± 1.12 (0-4.31) |
| Remnant stump length (cm) mean ± SD (range) | 3.87 ± 1.12 (0.69-5.0) |
Abbreviations: GSV, great saphenous vein; SD, standard deviation.
Mean Glue Extension and Remnant Stump Length According to Vein Diameter.
| Variables | GSV Diameter <0.7 cm (N = 59) | GSV Diameter ≥0.7 cm (N = 20) |
|
|---|---|---|---|
| Mean glue extension length | 1.40 ± 1.17 | 0.34 ± 0.25 | <.001 |
| Mean remnant stump length | 3.60 ± 1.17 | 4.67 ± 0.25 | <.001 |
Abbreviation: GSV, great saphenous vein.