Jin Li1, Weiwei Hu2, Min Li1, Shaoli Deng1, Qing Huang1, Weiping Lu1.
Abstract
BACKGROUND: Metallo-β-lactamase-producing Pseudomonas aeruginosa (MBL-PA) are important causative agents of nosocomial infections and are associated with significant mortality rates, especially in intensive care units. The timely detection and typing of these strains is essential for surveillance, outbreak prevention and antibiotic therapy optimization. In this study, fifteen VIM-type and fifteen SPM-type MBL-PA strains were selected as strains to establish MALDI-TOF MS SuperSpectra.Entities:
Keywords: SPM; SuperSpectra; VIM; VITEK MS plus system; matrix-assisted laser desorption/ionization time-of-flight mass spectrometry
Year: 2019 PMID: 31564928 PMCID: PMC6735624 DOI: 10.2147/IDR.S211984
Source DB: PubMed Journal: Infect Drug Resist ISSN: 1178-6973 Impact factor: 4.003
Figure 1UPGMA clustering of mass spectra of the established strains (VIM- and SPM-type MBL-producing P. aeruginosa). The VIM- and SPM-type MBL-PA isolates collected from Daping Hospital were identified as two cluster groups by the VITEK MS plus System, which means that there were some different peaks for each subspecies.
Figure 2Peak profiles of the spectra from VIM- and SPM-type MBL-producing P. aeruginosa. The SuperSpectra of VIM- and SPM-type MBL-PA were generated using the RUO saramis knowledge base database. Among these peaks, 17 peaks overlap, which used as the mass exclusion list (3601 m/z, 4434 m/z, 4544 m/z, 4849 m/z, 4992 m/z, 5114 m/z, 5210 m/z, 5737 m/z, 6046 m/z, 6348 m/z, 6676 m/z, 7203 m/z, 7595 m/z, 8786 m/z, 9089 m/z, 9987 m/z and 10,229 m/z).
Characteristics of the fifty isolates used for validation of the SuperSpectra
| Isolate | Patient location | Isolation site | MBL sequencing identification | MALDI-TOF MS identification | Confidence values (%) |
|---|---|---|---|---|---|
| 1 | ICU | Sputum | VIM-type MBL-PA | VIM-type MBL-PA | 99.9 |
| 2 | ICU | Sputum | VIM-type MBL-PA | VIM-type MBL-PA | 99.9 |
| 3 | ICU | Secretion | VIM-type MBL-PA | VIM-type MBL-PA | 92.0 |
| 4 | ICU | Sputum | VIM-type MBL-PA | VIM-type MBL-PA | 92.0 |
| 5 | Surgical ward | Urine | VIM-type MBL-PA | VIM-type MBL-PA | 89.0 |
| 6 | Medical ward | Sputum | VIM-type MBL-PA | VIM-type MBL-PA | 85.0 |
| 7 | Pediatrics | Sputum | VIM-type MBL-PA | VIM-type MBL-PA | 85.8 |
| 8 | CCU | Sputum | VIM-type MBL-PA | VIM-type MBL-PA | 99.9 |
| 9 | ICU | Urine | VIM-type MBL-PA | VIM-type MBL-PA | 99.9 |
| 10 | ICU | Sputum | VIM-type MBL-PA | VIM-type MBL-PA | 81.2 |
| 11 | ICU | Urine | VIM-type MBL-PA | VIM-type MBL-PA | 99.0 |
| 12 | ICU | Sputum | VIM-type MBL-PA | VIM-type MBL-PA | 99.0 |
| 13 | Surgical ward | Urine | VIM-type MBL-PA | VIM-type MBL-PA | 91.2 |
| 14 | Surgical ward | Sputum | VIM-type MBL-PA | VIM-type MBL-PA | 99.9 |
| 15 | Pediatrics | Blood | VIM-type MBL-PA | VIM-type MBL-PA | 99.9 |
| 16 | CCU | Sputum | VIM-type MBL-PA | VIM-type MBL-PA | 82.5 |
| 17 | ICU | Sputum | VIM-type MBL-PA | VIM-type MBL-PA | 99.9 |
| 18 | ICU | Sputum | VIM-type MBL-PA | VIM-type MBL-PA | 99.9 |
| 19 | Medical ward | Urine | VIM-type MBL-PA | VIM-type MBL-PA | 96.0 |
| 20 | Medical ward | Sputum | VIM-type MBL-PA | VIM-type MBL-PA | 95.2 |
| 21 | Surgical ward | Drainage | VIM-type MBL-PA | VIM-type MBL-PA | 95.2 |
| 22 | ICU | Sputum | VIM-type MBL-PA | VIM-type MBL-PA | 90.0 |
| 23 | ICU | BALF | VIM-type MBL-PA | VIM-type MBL-PA | 90.0 |
| 24 | Surgical ward | Pus | VIM-type MBL-PA | VIM-type MBL-PA | 88.0 |
| 25 | CCU | Sputum | VIM-type MBL-PA | VIM-type MBL-PA | 89.0 |
| 26 | Surgical ward | Bile | SPM-type MBL-PA | SPM-type MBL-PA | 99.9 |
| 27 | Medical ward | BALF | SPM-type MBL-PA | SPM-type MBL-PA | 99.9 |
| 28 | Medical ward | Pharyngeal swab | SPM-type MBL-PA | SPM-type MBL-PA | 99.9 |
| 29 | Pediatrics | Sputum | SPM-type MBL-PA | SPM-type MBL-PA | 99.9 |
| 30 | CCU | Urine | SPM-type MBL-PA | SPM-type MBL-PA | 88.0 |
| 31 | ICU | Sputum | SPM-type MBL-PA | SPM-type MBL-PA | 91.2 |
| 32 | ICU | Sputum | SPM-type MBL-PA | VIM-type MBL-PA | 85.0 |
| 33 | Surgical ward | Wound | SPM-type MBL-PA | SPM-type MBL-PA | 89.0 |
| 34 | Surgical ward | Catheter | SPM-type MBL-PA | SPM-type MBL-PA | 99.9 |
| 35 | Surgical ward | Joint fluid | SPM-type MBL-PA | SPM-type MBL-PA | 99.9 |
| 36 | ICU | Sputum | SPM-type MBL-PA | VIM-type MBL-PA | 81.2 |
| 37 | ICU | Sputum | SPM-type MBL-PA | VIM-type MBL-PA | 92.0 |
| 38 | CCU | Sputum | SPM-type MBL-PA | SPM-type MBL-PA | 99.0 |
| 39 | Surgical ward | Secretion | SPM-type MBL-PA | SPM-type MBL-PA | 99.9 |
| 40 | Pediatrics | Sputum | SPM-type MBL-PA | SPM-type MBL-PA | 99.9 |
| 41 | ICU | Sputum | SPM-type MBL-PA | SPM-type MBL-PA | 99.9 |
| 42 | ICU | Urine | SPM-type MBL-PA | SPM-type MBL-PA | 99.0 |
| 43 | Surgical ward | Urine | SPM-type MBL-PA | SPM-type MBL-PA | 99.9 |
| 44 | ICU | BALF | SPM-type MBL-PA | SPM-type MBL-PA | 99.9 |
| 45 | ICU | Urine | SPM-type MBL-PA | SPM-type MBL-PA | 99.9 |
| 46 | ICU | Sputum | SPM-type MBL-PA | SPM-type MBL-PA | 88.0 |
| 47 | ICU | Sputum | SPM-type MBL-PA | SPM-type MBL-PA | 89.0 |
| 48 | CCU | BALF | SPM-type MBL-PA | SPM-type MBL-PA | 99.9 |
| 49 | Surgical ward | Secretion | SPM-type MBL-PA | VIM-type MBL-PA | 82.9 |
| 50 | Medical ward | Sputum | SPM-type MBL-PA | SPM-type MBL-PA | 92.0 |
Results of validation of VITEK MS plus system in identification of VIM- and SPM-type MBL-producing P. aeruginosa isolates using the two SuperSpectra
| Validation strains | No. of strains | Identification by VITEK MS | Confidence values (%) | Percentage |
|---|---|---|---|---|
| VIM-type MBL-PA | 4 | VIM-type MBL-PA | 99.9 | 16.0 |
| 12 | VIM-type MBL-PA | 90.0–99.9 | 48.0 | |
| 9 | VIM-type MBL-PA | 80.0–90.0 | 36.0 | |
| Subtotal | 25 | VIM-type MBL-PA accuracy rate | 100.0 | |
| SPM-type MBL-PA | 5 | SPM-type MBL-PA | 99.9 | 20.0 |
| 8 | SPM-type MBL-PA | 90.0–99.9 | 32.0 | |
| 8 | SPM-type MBL-PA | 80.0–90.0 | 32.0 | |
| 4 | VIM-type MBL-PA | 90.0–99.9 | 16.0 | |
| Subtotal | 25 | SPM-type MBL-PA accuracy rate | 84.0 | |
| Total | 50 | Total accuracy rate | 92.0 |
Note: The validation results showed that all the VIM-type MBL-PA isolates were correctly identified, just four SPM-type MBL-PA isolates were incorrectly identified as VIM-type MBL-PA isolates which got conflicting results in repeat testing by VITEK MS plus system.