Literature DB >> 31560132

Effect of Screw Channel Angulation on Reverse Torque Values of Dental Implant Abutment Screws.

Eric Hu1,2, Anton Petrich1,2, Glen Imamura1,2, Christopher Hamlin1,2.   

Abstract

PURPOSE: To compare the reverse torque values (RTVs) of abutment screws tightened from three different angles.
MATERIALS AND METHODS: Implant abutment screws (n = 48), abutments (3), and regular platform implant analogs (3) were divided into three angulation groups (n = 16/group). Custom guides of 0°, 10°, and 20° were fabricated to verify driver angulation. The implant components for each group were assembled and all screws torqued to 35 Ncm using a universal screwdriver in a manual torque wrench at the appropriate angle. Torque was reapplied 10 minutes after initial torque. A digital gauge was then used to measure reverse torque at a position parallel (0°) to the implant analog. RTVs were recorded and compared using one-way ANOVA and Tukey HSD post-hoc comparisons (α = 0.05).
RESULTS: All mean RTVs fell below the targeted torque value of 35Ncm, with some values in each angulation group 10% below of the target value. Mean RTVs in descending order from targeted torque value were: 10° group = 32.07 ± 0.97 Ncm, 0° group = 31.16 ± 1.12 Ncm, and 20° group = 30.08 ± 0.88 Ncm. One-way ANOVA revealed significant differences between angulation groups (F = 15.954, p < 0.001). Tukey HSD post hoc comparisons revealed that the mean RTVs of the three angulation groups were significantly different from each other (0° vs. 10°: p = 0.033; 0° vs. 20°: p = 0.011; 10° vs. 20°: p < 0.001).
CONCLUSIONS: All RTVs did not reach the targeted torque value of 35 Ncm. Mean screw RTVs were significantly influenced by screwdriver insertion angulation.
© 2019 by the American College of Prosthodontists.

Entities:  

Keywords:  Angled screw channel; dental implants; reverse torque; screw loosening

Year:  2019        PMID: 31560132     DOI: 10.1111/jopr.13111

Source DB:  PubMed          Journal:  J Prosthodont        ISSN: 1059-941X            Impact factor:   2.752


  4 in total

1.  The effect of the digital manufacturing technique of cantilevered implant-supported frameworks on abutment screw preload.

Authors:  Shahad Mohammmed Altuwaijri; Hanan Nejer Alotaibi; Talal Mughaileth Alnassar
Journal:  J Adv Prosthodont       Date:  2022-02-25       Impact factor: 1.904

2.  Titanium Base Abutments in Implant Prosthodontics: A Literature Review.

Authors:  Ahmad M Al-Thobity
Journal:  Eur J Dent       Date:  2021-11-18

3.  Clinical and Laboratory Outcomes of Angled Screw Channel Implant Prostheses: A Systematic Review.

Authors:  Vanya Rasaie; Jaafar Abduo; Mehran Falahchai
Journal:  Eur J Dent       Date:  2022-02-21

4.  Aesthetic Outcomes and Peri-Implant Health of Angled Screw Retained Implant Restorations Compared with Cement Retained Crowns: Medium Term Follow-Up.

Authors:  Livia Nastri; Ludovica Nucci; Vincenzo Grassia; Rino Miraldi
Journal:  J Funct Biomater       Date:  2021-05-16
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.