| Literature DB >> 31555032 |
Yahdiana Harahap1, Devina Devina1, Harmita Harmita1.
Abstract
OBJECTIVE: Ethinyl estradiol and levonorgestrel as a combination of oral contraceptive drugs have very low dosage levels; hence, a highly sensitive and selective method of using ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) is needed.Entities:
Keywords: Ethinyl estradiol; human plasma; levonorgestrel; prednisone; ultra-performance liquid chromatography–tandem mass spectrometry; validation
Year: 2019 PMID: 31555032 PMCID: PMC6662037 DOI: 10.4103/jpbs.JPBS_68_19
Source DB: PubMed Journal: J Pharm Bioallied Sci ISSN: 0975-7406
The first gradient elution profile
| Min to- | Mobile phase A (%) | Mobile phase B (%) |
|---|---|---|
| 35 | 65 | |
| 15 | 85 | |
| 10 | 90 | |
| 5 | 95 | |
| 5 | 95 | |
| 35 | 65 | |
| 35 | 65 |
Figure 1Fragmentation spectrum of (a) ethinyl estradiol as Internal Standard; (b) levonorgestrel as Internal Standard; (c) prednisone as Internal Standard
Figure 2The reaction of ethinyl estradiol derivatizated by dansil chloride
The second gradient elution profile
| Min to- | Mobile phase A (%) | Mobile phase B (%) |
|---|---|---|
| 75 | 25 | |
| 5 | 95 | |
| 5 | 95 | |
| 75 | 25 | |
| 75 | 25 |
Figure 3Representative UPLC-MS/MS chromatograms of ethinyl estradiol, levonorgestrel, and prednisone in (a) human blank plasma spiked with analyte at LLOQ; (b) human plasma spiked with analyte at LLOQ
Data of inter-day calibration curve of ethinyl estradiol
| Replica inter-day | R | Slope (b) | Intercept (a) |
|---|---|---|---|
| 0.9958 | 0.5066 | −0.0428 | |
| 0.9984 | 0.3883 | 0.0078 | |
| 0.9989 | 0.6911 | −0.0110 | |
| 0.9977 | 0.5287 | −0.0153 |
The accuracy and precision from LLOQ of ethinyl estradiol
| LLOQ conc. (pg/mL) | Measured conc. (pg/mL) | Accuracy (%diff) | Precision (%CV) |
|---|---|---|---|
| 4.51 | −9.75 | 2.17 | |
| 4.76 | −4.71 | ||
| 4.73 | −5.32 | ||
| 4.62 | −7.54 | ||
| 4.62 | −7.61 |
The intra- and inter-day accuracy and precision of ethinyl estradiol
| Nominal conc. (pg/mL) | Intra-day | Inter-day | ||
|---|---|---|---|---|
| Mean accuracy (%diff) | Precision (%CV) | Mean accuracy (%diff) | Precision (%CV) | |
| −3.81 to 9,01 | 5.73 | −15.05 to 9.01 | 6.27 | |
| −14.79 to 14.69 | 12.45 | −14.80 to 14.69 | 9.21 | |
| −12.59 to 5.69 | 7.64 | −14.72 to 11.63 | 7.85 | |
| −14.56 to 1.68 | 7.76 | −14.84 to 4.74 | 7.06 | |
The stability test results of ethinyl estradiol and levonorgestrel in human plasma
| Stability experiments | Stable to- | |
|---|---|---|
| Ethinyl estradiol | Levonorgestrel | |
| 24 h | 24 h | |
| 10 days | 15 days | |
| 24 h | 24 h | |
| 3 cycles | 3 cycles | |
Data of inter-day calibration curve of levonorgestrel
| Replica inter-day | R | Slope (b) | Intercept (a) |
|---|---|---|---|
| 0.9984 | 0.0023 | −0.000012 | |
| 0.9992 | 0.0015 | 0.000100 | |
| 0.9974 | 0.0021 | 0.000043 | |
| 0.9983 | 0.0020 | 0.000044 |
The accuracy and precision from LLOQ of levonorgestrel
| LLOQ conc. (pg/mL) | Measured conc. (pg/mL) | Accuracy (%diff) | Precision (%CV) |
|---|---|---|---|
| 92.98 | −7.02 | 5.49 | |
| 90.71 | −9.29 | ||
| 104.29 | 4.29 | ||
| 93.68 | −6.32 | ||
| 96.10 | −3.90 |
The intra- and inter-day accuracy and precision of levonorgestrel
| Nominal conc. (pg/mL) | Intra-day | Inter-day | ||
|---|---|---|---|---|
| Mean accuracy (%diff) | Precision (%CV) | Mean accuracy (%diff) | Precision (%CV) | |
| −13.16 to −0.02 | 5.33 | −13.16 to 13.22 | 8.80 | |
| 12.70 to 13.47 | 12.13 | −12.70 to 14.60 | 8.91 | |
| −7.00 to 14.05 | 8.33 | −13.60 to 14.05 | 8.31 | |
| 8.71 to 14.53 | 2.24 | −14.75 to 14.53 | 10.82 | |