| Literature DB >> 31548740 |
Balram Babu Rajanbabu1, Kalaimani Dhandapani2.
Abstract
Perfusion-related problems are still reported at incidence, varying from .72 to 2.86%, and add to the morbidity and mortality of cardiac surgery. Awareness of such problems and establishing protocols to avoid and manage them is a key factor in further improving patient outcomes. In this report, we describe a homemade setup which was used as a desperate salvage measure to rewarm a 1-year-old baby on cardiopulmonary bypass for intracardiac repair, who was inadvertently cooled to 24°C because of a malfunctioning heater/cooler machine.Entities:
Keywords: cardiopulmonary bypass; heater/cooler failure
Mesh:
Year: 2019 PMID: 31548740 PMCID: PMC6749159
Source DB: PubMed Journal: J Extra Corpor Technol ISSN: 0022-1058