| Literature DB >> 31520230 |
François Beloncle1, Lise Piquilloud2,3, Pierre-Yves Olivier2,4, Alice Vuillermoz2, Elise Yvin2, Alain Mercat2, Jean-Christophe Richard5,6.
Abstract
BACKGROUND: Occlusion pressure at 100 ms (P0.1), defined as the negative pressure measured 100 ms after the initiation of an inspiratory effort performed against a closed respiratory circuit, has been shown to be well correlated with central respiratory drive and respiratory effort. Automated P0.1 measurement is available on modern ventilators. However, the reliability of this measurement has never been studied. This bench study aimed at assessing the accuracy of P0.1 measurements automatically performed by different ICU ventilators.Entities:
Keywords: Inspiratory effort; Mechanical ventilation; Occlusion pressure; Respiratory drive; Respiratory failure
Year: 2019 PMID: 31520230 PMCID: PMC6744533 DOI: 10.1186/s13613-019-0576-x
Source DB: PubMed Journal: Ann Intensive Care ISSN: 2110-5820 Impact factor: 6.925
Fig. 1Experimental setup. A chamber of a two-chamber test lung was connected to a driving ventilator set in volume-controlled ventilation mode. The second chamber (2) was connected to the tested ventilator by a double limb circuit with an active humidifier. The two chambers of the test lung were linked by a rigid metal piece, so that the positive pressure insufflated by the driving ventilator into the first chamber (1) generates a negative pressure in the second chamber, recognized as an inspiratory effort by the tested ventilator. Data were acquired via an analog-to-digital converter and stored in a laptop computer for subsequent analysis
Fig. 2Accuracy of automatically measured P0.1 displayed on the ventilator screen (P0.1vent). a Correlation between P0.1vent and reference P0.1 (P0.1ref). The horizontal dotted lines represent P0.1 reference values of 2.5, 5 and 10 cm H2O, respectively. The diagonal dotted line represents the identity line (y = x). b Bland and Altman plots of differences between P0.1vent and P0.1ref versus their mean. The dotted lines represent the mean bias and the limits of agreement
Fig. 3Relationship between automatically measured P0.1 displayed by the ventilator (P0.1vent) and P0.1 measured on the airway pressure during automated P0.1 measurements (P0.1aw). a Correlation between P0.1vent and P0.1aw. The dotted line represents the identity line (y = x). b Bland and Altman plots of differences between P0.1vent and P0.1aw versus their mean. The dotted lines represent the mean bias and the limits of agreement
Fig. 4P0.1 measurements (P0.1circuit) according to the position of the airway occlusion. P0.1circuit measurements were performed with occlusions at different positions (i.e., at 0, 195, 360 and 690 cm from the airway opening). The second compartment of the test lung was occluding using a hermetic plug placed at the airway opening or on a circuit tube of 195, 360 or 690 cm connected to the airway opening. Dotted lines represent P0.1 values of 2.5, 5 and 10 cm H2O, respectively
Fig. 5Automatically measured P0.1 displayed by the ventilator (P0.1vent), P0.1 measured on the airway pressure during automated P0.1 measurement (P0.1aw) and P0.1 measurement corrected for the occlusion location (P0.1vent corrected). P0.1vent corrected = P0.1vent + ∆P0.1circuit with ∆P0.1circuit corresponding to the difference between P0.1circuit measured with an occlusion performed at 360 cm from the airway opening and P0.1ref (reference P0.1) measured with an occlusion performed at the airway opening