PET/CT-guided biopsy with respiratory motion correction.

Ruoqiao Zhang1,2, Dženan Zukić3, Darrin W Byrd1, Andinet Enquobahrie3, Adam M Alessio1, Kevin Cleary4, Filip Banovac5, Paul E Kinahan6.   

Abstract

PURPOSE: Given the ability of positron emission tomography (PET) imaging to localize malignancies in heterogeneous tumors and tumors that lack an X-ray computed tomography (CT) correlate, combined PET/CT-guided biopsy may improve the diagnostic yield of biopsies. However, PET and CT images are naturally susceptible to problems due to respiratory motion, leading to imprecise tumor localization and shape distortion. To facilitate PET/CT-guided needle biopsy, we developed and investigated the feasibility of a workflow that allows to bring PET image guidance into interventional CT suite while accounting for respiratory motion.
METHODS: The performance of PET/CT respiratory motion correction using registered and summed phases method was evaluated through computer simulations using the mathematical 4D extended cardiac-torso phantom, with motion simulated from real respiratory traces. The performance of PET/CT-guided biopsy procedure was evaluated through operation on a physical anthropomorphic phantom. Vials containing radiolabeled 18F-fluorodeoxyglucose were placed within the physical phantom thorax as biopsy targets. We measured the average distance between target center and the simulated biopsy location among multiple trials to evaluate the biopsy localization accuracy.
RESULTS: The computer simulation results showed that the RASP method generated PET images with a significantly reduced noise of 0.10 ± 0.01 standardized uptake value (SUV) as compared to an end-of-expiration image noise of 0.34 ± 0.04 SUV. The respiratory motion increased the apparent liver lesion size from 5.4 ± 1.1 to 35.3 ± 3.0 cc. The RASP algorithm reduced this to 15.7 ± 3.7 cc. The distances between the centroids for the static image lesion and two moving lesions in the liver and lung, when reconstructed with the RASP algorithm, were 0.83 ± 0.72 mm and 0.42 ± 0.72 mm. For the ungated imaging, these values increased to 3.48 ± 1.45 mm and 2.5 ± 0.12 mm, respectively. For the ungated imaging, this increased to 1.99 ± 1.72 mm. In addition, the lesion activity estimation (e.g., SUV) was accurate and constant for images reconstructed using the RASP algorithm, whereas large activity bias and variations (± 50%) were observed for lesions in the ungated images. The physical phantom studies demonstrated a biopsy needle localization error of 2.9 ± 0.9 mm from CT. Combined with the localization errors due to respiration for the PET images from simulations, the overall estimated lesion localization error would be 3.08 mm for PET-guided biopsies images using RASP and 3.64 mm when using ungated PET images. In other words, RASP reduced the localization error by approximately 0.6 mm. The combined error analysis showed that replacing the standard end-of-expiration images with the proposed RASP method in PET/CT-guided biopsy workflow yields comparable lesion localization accuracy and reduced image noise.
CONCLUSION: The RASP method can produce PET images with reduced noise, attenuation artifacts and respiratory motion, resulting in more accurate lesion localization. Testing the PET/CT-guided biopsy workflow using computer simulation and physical phantoms with respiratory motion, we demonstrated that guided biopsy procedure with the RASP method can benefit from improved PET image quality due to noise reduction, without compromising the accuracy of lesion localization.

Entities:  

Keywords:  Image registration; PET/CT-guided biopsy; Respiratory motion correction

Mesh:

Year:  2019        PMID: 31512193      PMCID: PMC6899076          DOI: 10.1007/s11548-019-02047-4

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  25 in total

1.  3D Slicer as an image computing platform for the Quantitative Imaging Network.

Authors:  Andriy Fedorov; Reinhard Beichel; Jayashree Kalpathy-Cramer; Julien Finet; Jean-Christophe Fillion-Robin; Sonia Pujol; Christian Bauer; Dominique Jennings; Fiona Fennessy; Milan Sonka; John Buatti; Stephen Aylward; James V Miller; Steve Pieper; Ron Kikinis
Journal:  Magn Reson Imaging       Date:  2012-07-06       Impact factor: 2.546

2.  Positron emission tomography-CT imaging in guiding musculoskeletal biopsy.

Authors:  Paul J O'Sullivan; Eric M Rohren; John E Madewell
Journal:  Radiol Clin North Am       Date:  2008-05       Impact factor: 2.303

3.  Abdominal masses sampled at PET/CT-guided percutaneous biopsy: initial experience with registration of prior PET/CT images.

Authors:  Servet Tatli; Victor H Gerbaudo; Marcelo Mamede; Kemal Tuncali; Paul B Shyn; Stuart G Silverman
Journal:  Radiology       Date:  2010-07       Impact factor: 11.105

4.  Standard PET/CT of the chest during shallow breathing is inadequate for comprehensive staging of lung cancer.

Authors:  Martin Allen-Auerbach; Kristen Yeom; John Park; Michael Phelps; Johannes Czernin
Journal:  J Nucl Med       Date:  2006-02       Impact factor: 10.057

5.  PET/CT of the abdomen: optimizing the patient breathing pattern.

Authors:  Gerhard W Goerres; Cyrill Burger; Michael R Schwitter; Thai-Nia H Heidelberg; Burkhardt Seifert; Gustav K von Schulthess
Journal:  Eur Radiol       Date:  2002-06-22       Impact factor: 5.315

Review 6.  18-fluorodeoxyglucose positron emission tomographic imaging in the detection and monitoring of infection and inflammation.

Authors:  Hongming Zhuang; Abass Alavi
Journal:  Semin Nucl Med       Date:  2002-01       Impact factor: 4.446

7.  PET-CT: accuracy of PET and CT spatial registration of lung lesions.

Authors:  Christian Cohade; Medhat Osman; Laura N T Marshall; Richard N T L Wahl
Journal:  Eur J Nucl Med Mol Imaging       Date:  2003-03-01       Impact factor: 9.236

8.  The impact of respiratory motion on tumor quantification and delineation in static PET/CT imaging.

Authors:  Chi Liu; Larry A Pierce; Adam M Alessio; Paul E Kinahan
Journal:  Phys Med Biol       Date:  2009-11-20       Impact factor: 3.609

9.  PET/CT-guided percutaneous biopsy of isolated intramuscular metastases from postcricoid cancer.

Authors:  Govindarajan J I Mallarajapatna; Kumar G Kallur; Nagaraj K Ramanna; Sridhar P Susheela; Prashanth G Ramachandra
Journal:  J Nucl Med Technol       Date:  2009-11-13

10.  Designing Tracking Software for Image-Guided Surgery Applications: IGSTK Experience.

Authors:  Andinet Enquobahrie; David Gobbi; Matt Turek; Patrick Cheng; Ziv Yaniv; Frank Lindseth; Kevin Cleary
Journal:  Int J Comput Assist Radiol Surg       Date:  2008-11-01       Impact factor: 2.924

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.