Thomas Senoner1, Fabian Barbieri1, Georg Semmler2,3, Agne Adukauskaite1, Andrea Rubatscher1, Wilfried Schgör1, Markus Stühlinger1, Axel Bauer1, Bernhard Erich Pfeifer4,5, Lukas Fiedler6, Franz Xaver Roithinger6, Florian Hintringer1, Alois Suessenbacher7, Christian Georg Wollmann2,3, Wolfgang Dichtl1.
Abstract
OBJECTIVES: To evaluate the long-term performance of the SonRtip atrial lead.Entities:
Mesh:
Year: 2019 PMID: 31498840 PMCID: PMC6733485 DOI: 10.1371/journal.pone.0222269
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1Kaplan Meier curve comparing the time of occurrence of malfunctions between SonR tip and non-SonR tip atrial leads.
Fig 2Kaplan Meier curve comparing the time of lead replacement between SonR tip and non-SonR tip atrial leads.
Fig 3Kaplan Meier surival curve for all-cause mortality between CRT patients with SonR tip and non SonR tip atrial leads.
Characteristics of the studied population.
| Baseline characteristics | P-value | ||
|---|---|---|---|
| 68.03 ± 9.4 | 70.07 ± 10.4 | 0.428 | |
| 106 (74.1) | 387 (73.6) | 0.894 | |
| 2.63 ± 1.9 | 2.87 ± 1.6 | 0.042 | |
| 429.6 ± 95.3 | 557.76 ± 118.1 | 0.003 | |
| 0.84 ± 0.5 | 0.87 ± 0.5 | 0.001 | |
| 0 (0) | 110 (100) | < 0.001 | |
| 143 (25.6) | 416 (74.4) | < 0.001 |
Values are mean ± SD or n (%)
CRT-D: defibrillator; CRT-P: pacemaker; mV: millivolt; Non-SonRtip: Non-SonRtip atrial leads; Ω: Ohm SonRtip: SonRtip atrial leads; V: Volt
Characteristics of patients with atrial lead malfunction.
| gender / | lead type | days until lead malfunction | chest x-ray | atrial sensing (mV) | lead impedance (Ω) | pacing threshold (V) | consequences |
|---|---|---|---|---|---|---|---|
| SonRtip | 830 | Unremarkable | 0.4 | > 3000 | 0.25 | Lead replacement | |
| SonRtip | 1439 | No chest x-ray obtained at time of malfunction | 0.4 | > 3000 | 0.5 | Lead replacement | |
| SonRtip | 1288 | Unremarkable | 0.6 | > 3000 | 4.5 | Lead replacement | |
| SonRtip | 718 | No chest x-ray obtained at time of malfunction | 0.7 | > 3000 | 0.75 | Lead replacement | |
| SonRtip | 840 | No chest x-ray obtained at time of malfunction | 2.2 | > 3000 | 0.75 | Lead replacement | |
| SonRtip | 904 | Unremarkable | Not measurable | > 3000 | Not measurable | Lead replacement | |
| Ingevity 7741 | 1166 | 1.4 | 523 | 5 | Lead replacement | ||
| Solia S53 | 398 | Unremarkable | 1.5 | > 3000 | Not measurable | Patient underwent heart transplant | |
| SonRtip | 876 | Unremarkable | 2 | 466 | 4 | Lead has not been replaced yet | |
| CapSure SP Novus 4592 | 1181 | Unremarkable | 1.6 | 448 | Not measurable | Lead replacement |
mV: millivolt; Ω: Ohm; V: Volt