| Literature DB >> 31485282 |
Marius Lebret1,2, Marie-Caroline Rotty3,4, Cyril Argento2, Jean-Louis Pepin1,5, Renaud Tamisier1,5, François Arbib6, Dany Jaffuel4,7,8, Nicolas Molinari3,9, Jean-Christian Borel1,2,5.
Abstract
Auto-CPAP may cause sleep fragmentation due to variations in pressure and unintentional leaks. The aim of this study was to compare air leak between fixed-CPAP and auto-CPAP after 4 months of CPAP treatment. This study is an ancillary analysis of a randomized, double-blind, parallel, controlled trial over 4 months, comparing fixed- and auto-CPAP in newly diagnosed patients with OSA. The following data were extracted from the CPAP devices: mean and 90th percentile pressure, residual apnea-hypopnea index, mean CPAP use, and amount of leak. Within each arm, patients were also randomly allocated to use of one of the three different brands of devices. Since the leak was reported differently for each device, median leak value was determined for each brand and leaks were classified as "above the median" or "below the median". Data from 269 patients were analyzed. The univariate analysis showed that tobacco consumption, CPAP level, and oronasal masks were associated with leaks above the median value but not the type of CPAP. The multivariate analysis showed that only CPAP level and oronasal masks were associated with leaks below the median. There were no differences in the types of mask used between fixed- and auto-CPAP. There was no impact of the type of CPAP on leaks or the type of interface used. We used a method based on the median leak value to standardize comparisons across devices which report leaks with different definitions.Entities:
Mesh:
Year: 2019 PMID: 31485282 PMCID: PMC6702837 DOI: 10.1155/2019/6310956
Source DB: PubMed Journal: Can Respir J ISSN: 1198-2241 Impact factor: 2.409
Leak data reported for each device brand and CPAP mode.
| Type of device | Definition of leak data | Fixed-CPAP group | Auto-CPAP group |
|---|---|---|---|
| WEINMANN | Time spent with large leakage expressed in % | 0.4 (0.0; 2.4) | 1.5 (0.1; 4.5) |
|
| |||
| ResMed devices | Median nonintentional leakage expressed in L/min | 1.2 (0.0; 2.4) | 1.2 (0.0; 3.6) |
| 95th percentile of nonintentional leakage expressed in L/min | 12.0 (6.0; 21.6) | 10.8 (6.0; 16.8) | |
|
| |||
| Philips devices | Mean total leakage expressed in L/min | 39.6 (35.2; 41.5) | 36.0 (33.5; 40.2) |
| 90th percentile of total leakage expressed in L/min | 47.9 (42.0; 51.1) | 45.1 (41.0; 49.5) | |
Data are expressed as medians (25th; 75th) and means (SD). Device names: SOMNOsmart 2 for WEINMANN, S8 Spirit 2 for ResMed, and Remstar Auto for Philips. This table shows how the 3 manufacturers report leakage for each mode of PAP used in the study: fixed or auto. Data reporting differ between each of the CPAP adherence tracking system devices.
Figure 1Classification of patients according to leakage reported by machines; median leak value was determined for each brand. (a) Distribution of leak magnitudes in the ResMed devices expressed as unintentional leakage (L/min) (x-axis). (b) Distribution of leak magnitudes in the Philips devices expressed as unintentional leakage (L/min) (x-axis). (c) Distribution of leak magnitudes in the WEINMANN devices expressed as unintentional leakage (L/min) (x-axis).
Figure 2Study flowchart: patients included in the analysis.
Patient characteristics.
| Anthropometric data and comorbidities | General population ( | Fixed-CPAP ( | Auto-CPAP ( |
|
|---|---|---|---|---|
| Age (years) | 58 (50; 64) | 58 (51; 65) | 57 (49; 64) | 0.51 |
| Sex, female, No. (%) | 76 (28.3) | 36 (27.7) | 40 (28.8) | 0.84 |
| BMI (kg/m2) | 30.1 (27.0; 34.8) | 31.2 (27.5; 35.1) | 29.6 (26.6; 34.5) | 0.11 |
| Current smoker, No. (%) | 51 (19.0) | 25 (19.2) | 26 (18.7) | 0.91 |
| Hypertension, No. (%) | 180 (66.9) | 95 (73.1) | 85 (61.2) | 0.04 |
| Diabetes, No. (%) | 49 (18.2) | 25 (19.2) | 24 (17.3) | 0.68 |
|
| ||||
|
| ||||
| Residual AHI (events/h) | 3.7 (2.2; 6.9) | 3.8 (2.2; 8.0) | 3.6 (2.1; 6.3) | 0.31 |
| CPAP use (hours) | 5.5 (3.9; 6.7) | 5.7 (3.9; 6.8) | 5.4 (3.8; 6.7) | 0.41 |
| CPAP level, (cmH2O) | 8.5 (7; 10) | 9.0 (8; 11) | 7.5 (6.5; 9.4) | <0.0001 |
| Oronasal maska, No. (%) | 99 (37.9) | 52 (41.9) | 47 (34.3) | 0.20 |
| Chin strap, No. (%)b | 5 (1.9) | 0 (0) | 5 (3.6) | 0.06 |
| Heated humidifier, No. (%)c | 130 (48.7) | 64 (50) | 66 (47.5) | 0.68 |
|
| ||||
|
| ||||
| AHI, (events/h) | 39.4 (29.6; 56.9) | 39.4 (29.0; 59.1) | 39.8 (30.0; 56.6) | 0.62 |
| Epworth sleepiness scale, 1–24 | 9 (6; 13) | 10 (6; 15) | 9 (6; 13) | 0.07 |
BMI: body mass index, AHI: apnea-hypopnea index, and CPAP: continuous positive airway pressure; data are expressed as % or median (IQR). aMask type from 8 patients was missing; bchin strap use from 6 patients was missing; cheated humidifier use from 5 patients was missing. Data are reported as medians and quartiles or numbers and percentages, as appropriate.
Logistic regression analysis with “leaks above the median value” as the dependent variable.
| Univariate analysis | Multivariate analysis | |||||
|---|---|---|---|---|---|---|
| OR | 95% (CI) |
| AOR | 95% (CI) |
| |
| Age (years) | 1.01 | (0.99; 1.03) | 0.37 | |||
| Sex, female | 0.78 | (0.46; 1.32) | 0.35 | |||
| BMI (kg/m2) | 1.02 | (0.98; 1.05) | 0.41 | |||
| Initial AHI (event/h) | 1.01 | (0.99; 1.02) | 0.36 | |||
| Current smoker | 2.10 | (1.12; 3.93) |
| |||
| Hypertension | 0.98 | (0.59; 1.63) | 0.93 | |||
| Diabetes | 1.42 | (0.76; 2.64) | 0.27 | |||
| Residual AHI (event/h) | 1.05 | (0.99; 1.11) | 0.10 | |||
| CPAP use (hours) | 0.95 | (0.84; 1.07) | 0.38 | |||
| Fixed-CPAP | 1.17 | (0.73; 1.89) | 0.52 | |||
| CPAP level (cmH2O) | 1.25 | (1.11; 1.41) |
| 1.19 | (1.05; 1.35) |
|
| Oronasal mask | 2.75 | (1.64; 4.61) |
| 2.23 | (1.30; 3.83) |
|
| Chin strap | 0.71 | (0.12; 4.35) | 0.71 | |||
| Heated humidifier | 1.18 | (0.73; 1.90) | 0.51 | |||
BMI: body mass index, AHI: apnea-hypopnea index, CPAP: continuous positive airway pressure, OR: odds ratio, and AOR: adjusted odds ratio. Data are reported as medians and quartiles or numbers and percentages, as appropriate.
Comparison between fixed- and auto-CPAP.
| Fixed-CPAP | Auto-CPAP |
| |
|---|---|---|---|
|
| |||
| Current AHI (event/h) | 3.7 (2.2; 8.0) | 3.6 (2.1; 6.5) | 0.38 |
| CPAP use (h/night) | 5.7 (3.9; 6.7) | 5.3 (3.8; 6.7) | 0.35 |
| CPAP level (cmH2O) | 9.0 (8.0; 11.0) | 7.5 (6.5; 9.4) |
|
|
| |||
|
| |||
| Oronasala | 52 (41.6) | 49 (35.3) | 0.29 |
| Chin strapb | 0 (0) | 6 (4.23) | NA |
| Heated humidifierc | 64 (49.6) | 67 (47.2) | 0.69 |
AHI: apnea-hypopnea index and CPAP: continuous positive airway pressure. Data are expressed as % or median (IQR); amask type from 8 patients was missing; bchin strap use from 6 patients was missing; cheated humidifier use from 5 patients was missing.