A 32-channel multi-coil setup optimized for human brain shimming at 9.4T.

Ali Aghaeifar1,2, Jiazheng Zhou1,2, Rahel Heule1, Behzad Tabibian3, Bernhard Schölkopf3, Feng Jia4, Maxim Zaitsev4, Klaus Scheffler1,5.   

Abstract

PURPOSE: A multi-coil shim setup is designed and optimized for human brain shimming. Here, the size and position of a set of square coils are optimized to improve the shim performance without increasing the number of local coils. Utilizing such a setup is especially beneficial at ultrahigh fields where B0 inhomogeneity in the human brain is more severe.
METHODS: The optimization started with a symmetric arrangement of 32 independent coils. Three parameters per coil were optimized in parallel, including angular and axial positions on a cylinder surface and size of the coil, which were constrained by cylinder size, construction consideration, and amplifiers specifications. B0 maps were acquired at 9.4T in 8 healthy volunteers for use as training data. The global and dynamic shimming performance of the optimized multi-coil were compared in simulations and measurements to a symmetric design and to the scanner's second-order shim setup, respectively.
RESULTS: The optimized multi-coil performs better by 14.7% based on standard deviation (SD) improvement with constrained global shimming in comparison to the symmetric positioning of the coils. Global shimming performance was comparable with a symmetric 65-channel multi-coil and full fifth-order spherical harmonic shim coils. On average, an SD of 48.4 and 31.9 Hz was achieved for in vivo measurements after global and dynamic slice-wise shimming, respectively.
CONCLUSIONS: An optimized multi-coil shim setup was designed and constructed for human whole-brain shimming. Similar performance of the multi-coils with many channels can be achieved with a fewer number of channels when the coils are optimally arranged around the target.
© 2019 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  B0 inhomogeneity; B0 shimming; echo planar imaging; multi-coil; optimization; ultrahigh field

Year:  2019        PMID: 31483527     DOI: 10.1002/mrm.27929

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  5 in total

1.  Analysis of coil element distribution and dimension for matrix gradient coils.

Authors:  Hongyan He; Shufeng Wei; Huixian Wang; Wenhui Yang
Journal:  MAGMA       Date:  2022-06-11       Impact factor: 2.310

2.  Physical limits to human brain B0 shimming with spherical harmonics, engineering implications thereof.

Authors:  Bruno Pinho Meneses; Alexis Amadon
Journal:  MAGMA       Date:  2022-07-13       Impact factor: 2.533

3.  Iterative static field map estimation for off-resonance correction in non-Cartesian susceptibility weighted imaging.

Authors:  Guillaume Daval-Frérot; Aurélien Massire; Boris Mailhe; Mariappan Nadar; Alexandre Vignaud; Philippe Ciuciu
Journal:  Magn Reson Med       Date:  2022-06-23       Impact factor: 3.737

4.  Assessing Age-Related Gray Matter Differences in Young Adults with Voxel-Based Morphometry: The Effect of Field Strengths.

Authors:  Feng-Yi Su; Jyun-Ru Chen; Chun-Ming Chen; Yen-Chih Huang; Shin-Lei Peng
Journal:  Brain Sci       Date:  2021-03-31

5.  An integrated RF-receive/B0-shim array coil boosts performance of whole-brain MR spectroscopic imaging at 7 T.

Authors:  Morteza Esmaeili; Jason Stockmann; Bernhard Strasser; Nicolas Arango; Bijaya Thapa; Zhe Wang; Andre van der Kouwe; Jorg Dietrich; Daniel P Cahill; Tracy T Batchelor; Jacob White; Elfar Adalsteinsson; Lawrence Wald; Ovidiu C Andronesi
Journal:  Sci Rep       Date:  2020-09-14       Impact factor: 4.996

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.