Alexander Breitenstein1, Jan Steffel1.
Abstract
Despite advances in heart failure treatment, this condition remains a relevant medical issue and is associated with a high morbidity and mortality. The cause of death in patients suffering from heart failure is not only a result of hemodynamic failure, but can also be due to ventricular arrhythmias. Implantable cardioverter defibrillators (ICDs) are these days the only tool to significantly reduce arrhythmic sudden death; but not all patients benefit to the same extend. In addition, cardiac resynchronization therapy (CRT) is another tool which is used in patients suffering from heart fialure. Even though both devices have been investigated in large randomized trials, both ICD and CRT remain underutilized in many countries. This brief review focuses on various aspects in this regard including a short overview on upcoming device novelties in the near future.Entities:
Keywords: cardiac resynchronization; heart failure; imaging; implantable cardioverter defibrillator; subcutaneous ICD (S-ICD)
Year: 2019 PMID: 31457018 PMCID: PMC6700378 DOI: 10.3389/fcvm.2019.00111
Source DB: PubMed Journal: Front Cardiovasc Med ISSN: 2297-055X
Current ICD and CRT indications according to the ESC guidelines.
| Secondary prevention |
In patients with documented VF or haemodynamically not tolerated VT Absence of reversible causes or within 48h after myocardial infarction Reasonable expectation of survival with a good functional status > 1 year | IA |
| Primary prevention | Severely impaired LV function (≤ 35 %)
With NYHA II-IVa | |
|
Despite optimal medical treatment for ≥ 3 months | ||
|
Ischemic cardiomyopathy Non-ischemic cardiomyopathy | IA | |
|
Severely impaired LV function (≤ 35 %) NYHA II-IVa Despite optimal medical treatment for ≥ 3 months | ||
|
With LBBB and QRS duration > 150 ms With LBBB and QRS duration 130-149 ms | IA | |
|
With non-LBBB and QRS duration > 150 ms With non-LBBB and QRS duration 130-149 ms | IIA |
Figure 1Electroanatomical voltage mapping of the left ventricle (posterior view). Red areas indicate scar tissue, whereas pink areas demonstrate healthy myocardium.
Figure 2Antero-posterior chest X-ray after subcutaneous ICD (S-ICD) implantation.
Figure 3Possible responses to cardiac resynchronization [Reproduced with permission from Circulation (36)].