Finite-element analysis and optimization of the mechanical properties of polyetheretherketone (PEEK) clasps for removable partial dentures.

Tzu-Yu Peng1, Yuki Ogawa2, Hiroyuki Akebono3, Shogo Iwaguro4, Atsushi Sugeta3, Saiji Shimoe5.   

Abstract

PURPOSE: Polyetheretherketone (PEEK), a high-strength, aesthetic, and non-allergic thermoplastic polymer, recently became a candidate for replacing metallic components in dental prosthesis. However, as PEEK is flexible, the need for retention presents a key challenge in terms of its clinical application. In this study, clasps prepared using PEEK were optimized and evaluated to provide the mechanical properties required by dentures.
METHODS: Seventy-two three-dimensional rod-shape models, based on four thickness/width ratios, three base widths, and six taper ratios were created. These models were analyzed using finite-element methods to determine which modified clasp arm shape provided the most appropriate mechanical properties. Three shape-optimized PEEK specimens and one standard-shape Co-Cr alloy specimen were then fabricated. Constant-displacement fatigue testing was performed to calculate load values and deformations after ten years of clinical use.
RESULTS: Shape optimization indicated a maximum stress concentration that was consistently located at the base of the specimen, a correlation between mean load values and thickness that was greater than that with the width, and a correlation between taper ratio and mean load values. Fatigue testing showed that although PEEK exhibited significantly lower average load values than the Co-Cr alloy, these were sufficient for clinical use. All specimens exhibited significant deformation during the first period of cycling; however, there was no significant difference in the deformation between the two materials after fatigue testing.
CONCLUSIONS: PEEK exerts fewer stresses on abutments compared to standard-alloy clasps, provides adequate retention, and satisfy aesthetic demands, indicating that PEEK presents a promising alternative to conventional metal clasps.
Copyright © 2019 Japan Prosthodontic Society. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  CAD/CAM; Dental clasp; Fatigue; Polyetheretherketone; Shape optimization

Year:  2019        PMID: 31455560     DOI: 10.1016/j.jpor.2019.07.012

Source DB:  PubMed          Journal:  J Prosthodont Res        ISSN: 1883-1958            Impact factor:   4.642


  7 in total

Review 1.  PEEK in Fixed Dental Prostheses: Application and Adhesion Improvement.

Authors:  Biyao Wang; Minghao Huang; Pengrui Dang; Jiahui Xie; Xinwen Zhang; Xu Yan
Journal:  Polymers (Basel)       Date:  2022-06-08       Impact factor: 4.967

Review 2.  Review on Development and Dental Applications of Polyetheretherketone-Based Biomaterials and Restorations.

Authors:  Ludan Qin; Shuo Yao; Jiaxin Zhao; Chuanjian Zhou; Thomas W Oates; Michael D Weir; Junling Wu; Hockin H K Xu
Journal:  Materials (Basel)       Date:  2021-01-15       Impact factor: 3.623

3.  Retention force of polyetheretherketone and cobalt-chrome-molybdenum removable dental prosthesis clasps after artificial aging.

Authors:  Felicitas Mayinger; Danka Micovic; Andreas Schleich; Malgorzata Roos; Marlis Eichberger; Bogna Stawarczyk
Journal:  Clin Oral Investig       Date:  2020-10-16       Impact factor: 3.573

4.  Performance of Graphene-Based and Polyether-Ether-Ketone Polymers as Removable Partial Denture Esthetic Clasp Materials after Cyclic Fatigue.

Authors:  Mostafa Omran Hussein
Journal:  Polymers (Basel)       Date:  2022-07-23       Impact factor: 4.967

5.  Laser-Milled Microslits Improve the Bonding Strength of Acrylic Resin to Zirconia Ceramics.

Authors:  Saiji Shimoe; Tzu-Yu Peng; Yuki Wakabayashi; Hiroto Takenaka; Shogo Iwaguro; Masato Kaku
Journal:  Polymers (Basel)       Date:  2020-04-04       Impact factor: 4.329

Review 6.  The use of PEEK in digital prosthodontics: A narrative review.

Authors:  Ioannis Papathanasiou; Phophi Kamposiora; George Papavasiliou; Marco Ferrari
Journal:  BMC Oral Health       Date:  2020-08-02       Impact factor: 2.757

7.  Is the high-performance thermoplastic polyetheretherketone indicated as a clasp material for removable dental prostheses?

Authors:  Danka Micovic; Felicitas Mayinger; Sebastian Bauer; Malgorzata Roos; Marlis Eichberger; Bogna Stawarczyk
Journal:  Clin Oral Investig       Date:  2020-10-07       Impact factor: 3.573

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.