How the rise time of carbon dioxide analysers influences the accuracy of carbon dioxide measurements.

J X Brunner1, D R Westenskow.   

Abstract

Carbon dioxide measurements are not accurate, especially in children, if the response time of the carbon dioxide analyser is too slow and its output fails to reach the actual carbon dioxide concentration at the end of the breath. The distortion of the carbon dioxide waveform is a function of the "rise time" of the analyser. We have simulated an expired carbon dioxide curve and calculated the rise time required to measure accurately end-tidal carbon dioxide and VCO2 in adults and children. A rise time of 80 ms (10-70%) is sufficient to measure end-tidal carbon dioxide concentration with 5% accuracy in patients with rates of ventilation less than 100 b.p.m. and I:E ratios less than 2:1. We have measured the rise time of 11 commercially available carbon dioxide analysers and found that only six of the 11 responded quickly enough to be accurate for rates up to 100 b.p.m. All 11 responded rapidly enough to measure end-tidal carbon dioxide concentration with 5% accuracy when ventilatory rates were less than 30 b.p.m. To measure VCO2 with 5% accuracy, an analyser should have a rise time of 20 ms. Only one analyser met this specification. An analyser's rise time can be estimated clinically to within 10 (SD 8) ms by a simple breath hold and forced exhalation, thus providing an estimate of the accuracy of carbon dioxide measurements in adults or children.

Entities:  

Mesh:

Substances:

Year:  1988        PMID: 3144999     DOI: 10.1093/bja/61.5.628

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  13 in total

Review 1.  Capnometry and anaesthesia.

Authors:  K Bhavani-Shankar; H Moseley; A Y Kumar; Y Delph
Journal:  Can J Anaesth       Date:  1992-07       Impact factor: 5.063

2.  Laboratory evaluation of the vital signs (ICOR) piezoelectric anesthetic agent analyzer.

Authors:  D R Westenskow; F H Silva
Journal:  J Clin Monit       Date:  1991-04

3.  Computerization of anesthesia information management.

Authors:  J H Eichhorm; D W Edsall
Journal:  J Clin Monit       Date:  1991-01

4.  Can the Raman scattering analyzer compete with mass spectrometers: an affirmative reply.

Authors:  D R Westenskow; D L Coleman
Journal:  J Clin Monit       Date:  1989-01

5.  The effect of airway narrowing and dead space on the shape of the capnogram.

Authors:  Hiroyuki Kinoshita; Kenichi Sakamoto; Masaharu Ito
Journal:  J Anesth       Date:  1995-06       Impact factor: 2.078

6.  Automated evaluation of functional residual capacity by oxygen washout.

Authors:  Wolfgang Eichler; Jan Schumacher; Angela Roth-Isigkeit; Jörg Braun; Hermann Kuppe; Karl-Friedrich Klotz
Journal:  J Clin Monit Comput       Date:  2002 Apr-May       Impact factor: 2.502

7.  A simple method to estimate functional residual capacity in mechanically ventilated patients.

Authors:  R Fretschner; H Deusch; A Weitnauer; J X Brunner
Journal:  Intensive Care Med       Date:  1993       Impact factor: 17.440

8.  Improved response time with a new miniaturised main-stream multigas monitor.

Authors:  Mattias Berggren; Nasser Hosseini; Krister Nilsson; Ola Stenqvist
Journal:  J Clin Monit Comput       Date:  2009-10-10       Impact factor: 2.502

9.  Use of end-tidal PCO2 and transcutaneous PCO2 as noninvasive measurement of arterial PCO2 in extubated patients recovering from general anesthesia.

Authors:  B P Kavanagh; A N Sandler; K E Turner; V Wick; S Lawson
Journal:  J Clin Monit       Date:  1992-07

Review 10.  Terminology and the current limitations of time capnography: a brief review.

Authors:  K Bhavani-Shankar; A Y Kumar; H S Moseley; R Ahyee-Hallsworth
Journal:  J Clin Monit       Date:  1995-05
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.