Immediate Impact of Prosthetic Graft Replacement of the Ascending Aorta on Circumferential Strain in the Descending Aorta.

Lisa Q Rong1, Maria C Palumbo2, Mohamed Rahouma3, Massimiliano Meineri4, Gabriel R Arguelles5, Jiwon Kim2, Christopher Lau3, Richard B Devereux2, Kane O Pryor5, Leonard N Girardi3, Alberto Redaelli6, Mario F L Gaudino3, Jonathan W Weinsaft2.   

Abstract

OBJECTIVES: Prosthetic replacement of the ascending aorta (AA) can potentially modify energy propagation to the distal aorta and contribute to adverse aortic remodelling. This preliminary study employed intra-operative transoesophageal echocardiography (TOE) to assess the immediate impact of prosthetic graft replacement of the AA on circumferential strain in the descending aorta.
METHODS: Intra-operative TOEs in patients undergoing AA graft replacement were analysed for circumferential strain, fractional area change (FAC), dimensions (end diastolic area [EDA], and end systolic area [ESA]) in the descending aorta immediately before and after graft replacement. Deformation was assessed via global peak circumferential aortic strain (CAS), together with pulse pressure corrected strain, time to peak strain (TTP), and aortic distensibility.
RESULTS: Forty-five patients undergoing AA replacement with prosthetic graft (91% elective) were studied. Following grafting, descending thoracic aortic circumferential strain increased (6.3 ± 2.8% vs. 8.9 ± 3.4%, p = .001) paralleling distensibility (5.7 [3.7-8.6] 10-3 mmHg vs. 8.5 [6.4-12.4] 10-3 mmHg, p < .001). Despite slight increments in post graft left ventricular ejection fraction (LVEF) (52.3 ± 10.8% vs. 55.0 ± 11.9, p < .001), stroke volume was similar (p = .41), and magnitude of increased strain did not correlate with change in stroke volume (r = -.03, p = .86), LVEF (r = .18, p = .28), or pulse pressure (r = .28, p = .06). Descending aortic size (EDA 4 [2.7-4.6] cm2vs. 3.7 [2.5-5] cm2, p = .89; ESA 4.3 [3.2-5.3] cm2vs. 4.5 [3.3-5.8] cm2, p = .14) was similar pre- and post graft. In subgroup analysis, patients with cystic medial necrosis had a significantly higher post procedure CAS than patients with atherosclerotic aneurysms (9.7 ± 3.5% vs. 7.0 ± 2.3%, p = .03).
CONCLUSIONS: Prosthetic graft replacement of the AA increases immediate aortic circumferential strain of the descending aorta, particularly in patients with cystic medial necrosis. Our findings suggest that grafts augment energy transfer to the distal aorta, a potential mechanism for progressive distal aortic dilation and/or dissection.
Copyright © 2019 European Society for Vascular Surgery. Published by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Aortic aneurysm; Echocardiography; Prosthetic grafts; Strain

Mesh:

Year:  2019        PMID: 31445862     DOI: 10.1016/j.ejvs.2019.05.003

Source DB:  PubMed          Journal:  Eur J Vasc Endovasc Surg        ISSN: 1078-5884            Impact factor:   7.069


  5 in total

1.  Impact of ascending aortic prosthetic grafts on early postoperative descending aortic biomechanics on cardiac magnetic resonance imaging.

Authors:  Maria C Palumbo; Alberto Redaelli; Matthew Wingo; Katherine A Tak; Jeremy R Leonard; Jiwon Kim; Lisa Q Rong; Christine Park; Hannah W Mitlak; Richard B Devereux; Mary J Roman; Arindam RoyChoudury; Christopher Lau; Mario F L Gaudino; Leonard N Girardi; Jonathan W Weinsaft
Journal:  Eur J Cardiothorac Surg       Date:  2022-03-24       Impact factor: 4.191

2.  Prosthetic aortic graft replacement of the ascending thoracic aorta alters biomechanics of the native descending aorta as assessed by transthoracic echocardiography.

Authors:  Maria C Palumbo; Lisa Q Rong; Jiwon Kim; Pedram Navid; Razia Sultana; Jonathan Butcher; Alberto Redaelli; Mary J Roman; Richard B Devereux; Leonard N Girardi; Mario F L Gaudino; Jonathan W Weinsaft
Journal:  PLoS One       Date:  2020-03-12       Impact factor: 3.240

3.  Combined thoracic endovascular aortic repair and endovascular aneurysm repair and the long-term consequences of altered cardiovascular haemodynamics on morbidity and mortality: case series and literature review.

Authors:  Sherif Sultan; Yogesh Acharya; Mohiey Hazima; Hiba Salahat; Juan Carlos Parodi; Niamh Hynes
Journal:  Eur Heart J Case Rep       Date:  2021-09-15

Review 4.  The contemporary design of endovascular aneurysm stent-graft materials: PTFE versus polyester.

Authors:  Niamh Hynes; Yogesh Acharya; Sherif Sultan
Journal:  Front Surg       Date:  2022-08-16

Review 5.  TEVAR and EVAR, the unknown knowns of the cardiovascular hemodynamics; and the immediate and long-term consequences of fabric material on major adverse clinical outcome.

Authors:  Sherif Sultan; Yogesh Acharya; Osama Soliman; Juan Carlos Parodi; Niamh Hynes
Journal:  Front Surg       Date:  2022-08-30
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.