| Literature DB >> 31440021 |
Alberto Mavilio1, Dario Sisto2, Paolo Ferreri2, Rosanna Dammacco2, Giovanni Alessio2.
Abstract
PURPOSE: To evaluate reliability of steady-state pattern electroretinogram (ssPERG) phase variability in re-test (procedure called RE-PERG) in the presence of myopia, which is known to affect ssPERG amplitude, in glaucomatous patients (GP), normal controls (NC), and myopic patients (MY).Entities:
Keywords: glaucoma; myopia; pattern electroretinogram; perg; re-perg; steady-state
Year: 2019 PMID: 31440021 PMCID: PMC6664252 DOI: 10.2147/OPTH.S211337
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Figure 1Five consecutive blocks of steady-state pattern electroretinogram (time domain – 512 samples).
Figure 2Five consecutive blocks of steady-state pattern electroretinogram (frequency domain – rps).
Demographic and relevant ocular characteristic of study participants
| Patients (n=112 cases) | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| GP (n=50) | NC (n=35) | My (n=19) | |||||||
| Mean | SD ± | Mean | SD ± | Mean | SD ± | GP vs NC | GP vs My | NC vs My | |
| Age | 67.4 | 11.1 | 65.2 | 9.9 | 64.0 | 7.7 | |||
| Male (%) | 50 | 51 | 68 | ||||||
| MD (db) | −2.5 | 1.91 | 0.35 | 1.00 | −1.27 | 1.26 | |||
| PSD (db) | 2.66 | 2.05 | 1.29 | 0.36 | 1.45 | 0.27 | |||
| RNFL (µm) | 75.84 | 13.34 | 92.03 | 8.93 | 85.5 | 10..5 | |||
| GCC (µm) | 67.26 | 11.70 | 80.14 | 4.94 | 79.3 | 9.4 | |||
| ssPERG Amp (µV) | 1.36 | 0.14 | 1.67 | 0.15 | 1.41 | 0.2 | |||
| ssPERG SD ph (SD) | 0.30 | 0.14 | 0.10 | 0.04 | 0.12 | 0.03 | |||
Notes: *One-way analysis of variance (Bonferroni corrected); **chi-square.
Abbreviations: MD, mean deviation; PSD, pattern standard deviation; RNFL, retinal nerve fiber layer thickness; GCC, ganglion cell complex; ssPERG SDph, steady-state intrinsic phase variability; ssPERG Amp, steady-state pattern electroretinogram mean amplitude value; GP, glaucomatous patients; NC, normal controls; My, myopic subjects.
Clinical characteristics of myopic subjects
| Normality range | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Gender | Age | ssPERG Amp | ssPERG SDph | MD | PSD | RNFL | GCC | ssPERG Amp | ssPERG SDph | |
| #1 | m | 44 | 1.42 | 0.09 | −0.4 | 1.8 | 88 | 72 | a | n |
| #2 | m | 74 | 1.48 | 0.13 | −0.5 | 1.3 | 74 | 64 | a | n |
| #3 | m | 78 | 1.36 | 0.1 | −2.01 | 1.2 | 88 | 70 | a | n |
| #4 | m | 68 | 1.15 | 0.12 | −3.5 | 1.5 | 104 | 91 | a | n |
| #5 | m | 60 | 1.18 | 0.11 | −3.21 | 1.55 | 101 | 91 | a | n |
| #6 | f | 74 | 1.19 | 0.12 | −2.4 | 1.88 | 87 | 71 | a | n |
| #7 | f | 56 | 1.09 | 0.1 | −2.13 | 1.25 | 75 | 69 | a | n |
| #8 | m | 55 | 1.13 | 0.11 | −1.5 | 1.24 | 79 | 77 | a | n |
| #9 | f | 57 | 1.54 | 0.11 | 0.35 | 1.88 | 76 | 77 | n | n |
| #10 | f | 57 | 1.56 | 0.14 | −2.89 | 1.24 | 98 | 92 | n | n |
| #11 | f | 68 | 1.15 | 0.16 | −1.4 | 1.4 | 102 | 84 | a | a |
| #12 | m | 60 | 1.51 | 0.11 | −2.1 | 1.53 | 99 | 90 | n | n |
| #13 | m | 68 | 1.53 | 0.16 | 0.93 | 1.46 | 84 | 80 | n | a |
| #14 | m | 65 | 1.44 | 0.07 | −3.22 | 1.7 | 78 | 79 | a | n |
| #15 | m | 61 | 1.55 | 0.07 | −1.52 | 1.21 | 70 | 68 | n | n |
| #16 | m | 69 | 1.53 | 0.12 | −0.06 | 1.34 | 72 | 55 | n | n |
| #17 | f | 62 | 1.5 | 0.15 | −0.45 | 1.34 | 70 | 87 | a | a |
| #18 | m | 57 | 1.75 | 0.2 | 1 | 1.4 | 100 | 90 | n | a |
| #19 | m | 55 | 1.76 | 0.17 | 0.4 | 1.5 | 99 | 90 | n | a |
Abbreviations: ssPERG Amp, steady-state pattern electroretinogram mean amplitude value; ssPERG SDph, standard deviation of steady-state PERG phase; MD, mean deviation; PSD, pattern standard deviation; RNFL, retinal nerve fiber layer thickness; GCC, ganglion cell complex in 19 Myopic subjects.
Pearson correlation coefficient (CC) and significance level P (SL-P) between mean deviation (MD) pattern standard deviation (PSD), retinal Nerve fiber layer thickness (RNFL)*, ganglion cell complex (GCC)*, steady-state pattern electroretinogram mean amplitude value (ssPERG Amp)*, standard deviation of steady-state PERG phase (ssPERG SDph) in all participants
| MD | PSD | GCC | RNFL | Amplitude | SD Phase | ||
|---|---|---|---|---|---|---|---|
| MD | CC | −0.59 | 0.55 | 0.52 | 0.72 | −0.63 | |
| SL-P | <0.0001 | <0.0001 | <0.0001 | <0.0001 | <0.0001 | ||
| PSD | CC | −0.59 | −0.34 | −0.33 | −0.34 | 0.41 | |
| SL-P | <0.0001 | 0.0005 | 0.0005 | 0.0004 | <0.0001 | ||
| GCC | CC | 0.55 | −0.34 | 0.81 | 0.49 | −0.52 | |
| SL-P | <0.0001 | 0.0005 | <0.0001 | <0.0001 | <0.0001 | ||
| RNFL | CC | 0.52 | −0.33 | 0.81 | 0.48 | −0.47 | |
| SL-P | <0.0001 | 0.0005 | <0.0001 | <0.0001 | <0.0001 | ||
| ssPERG Amp | CC | 0.72 | −0.34 | 0.49 | 0.48 | −0.27 | |
| SL-P | <0.0001 | 0.0004 | <0.0001 | <0.0001 | 0.006 | ||
| ssPERG SDph | CC | −0.63 | 0.41 | −0.52 | −0.5 | −0.55 | |
| SL-P | <0.0001 | <0.0001 | <0.0001 | <0.0001 | <0.0001 |
Notes: *RNFL, GCC, and ssPERG Amp were converted to a logarithmic scale (dB) in association with visual field indices.
Pearson correlation coefficient (CC) and significance level P (SL-P) between mean deviation (MD) pattern standard deviation (PSD), retinal nerve fiber layer thickness (RNFL)*, ganglion cell complex (GCC)*, steady-state pattern electroretinogram mean amplitude value (ssPERG Amp)*, standard deviation of steady-state PERG phase (ssPERG SDph) in myopic and glaucomatous patients
| MD | PSD | GCC | RNFL | Amplitude | SD Phase | ||
|---|---|---|---|---|---|---|---|
| MD | CC | −0.19 | −0.07 | −0.12 | 0.48 | −0.06 | |
| SL-P | 0.06 | 0.63 | 0.38 | 0.001 | 0.67 | ||
| PSD | CC | −0.19 | 0.01 | 0.14 | −0.15 | −0.09 | |
| SL-P | 0.06 | 0.93 | 0.31 | 0.28 | 0.54 | ||
| GCC | CC | −0.07 | 0.01 | 0.65 | −0.01 | 0.24 | |
| SL-P | 0.63 | 0.92 | <0.001 | 0.93 | 0.08 | ||
| RNFL | CC | −0.12 | 0.14 | 0.65 | 0.03 | 0.22 | |
| SL-P | 0.38 | 0.31 | <0.001 | 0.84 | 0.11 | ||
| ssPERG Amp | CC | 0.48 | −0.15 | −0.01 | 0.03 | −0.14 | |
| SL-P | 0.001 | 0.28 | 0.93 | 0.84 | 0.32 | ||
| ssPERG SDph | CC | −0.06 | −0.09 | 0.24 | 0.22 | −0.14 | |
| SL-P | 0.67 | 0.54 | 0.08 | 0.11 | 0.32 |
Notes: *RNFL, GCC, and ssPERG Amp were converted to a logarithmic scale (dB) in association with visual field indices.
(A) Receiver operating characteristics (ROCs) without myopic subjects. (B) ROCs myopic subjects included
| AUC | SEa | 95% CIb | Specificity | 95% CIb | |
|---|---|---|---|---|---|
| MD | 0.83 | 0.05 | 0.738–0.906 | 71.79 | 48.72–89.74 |
| PSD | 0.78 | 0.05 | 0.683–0.867 | 62.56 | 33.33–84.62 |
| GCC | 0.79 | 0.05 | 0.685–0.868 | 54.36 | 28.45–77.55 |
| RNFL | 0.77 | 0.05 | 0.664–0.853 | 64.10 | 47.20–78.80 |
| ssPERG Amp | 0.86 | 0.04 | 0.765–0.924 | 82.10 | 66.50–92.50 |
| ssPERG SDph | 0.93 | 0.04 | 0.857–0.976 | 84.60 | 69.50–94.10 |
| MD | 0.76 | 0.05 | 0.671–0.842 | 84.48 | 72.6–92.7 |
| PSD | 0.79 | 0.05 | 0.701–0.865 | 77.59 | 64.7–87.5 |
| GCC | 0.77 | 0.04 | 0.676–0.846 | 81.03 | 68.6–90.1 |
| RNFL | 0.73 | 0.04 | 0.640–0.817 | 58.62 | 44.9–71.4 |
| ssPERG Amp | 0.76 | 0.05 | 0.664–0.837 | 70.69 | 57.3–81.9 |
| ssPERG SDph | 0.86 | 0.04 | 0.778–0.920 | 93.10 | 83.3–98.1 |
Notes: Data from these studies.12,13,30 aDeLong et al, 1988. bBinomial exact.
Abbreviations: MD, mean deviation; PSD, pattern standard deviation; RNFL, retinal nerve fiber layer thickness; GCC, ganglion cell complex; ssPERG Amp, steady-state pattern electroretinogram mean amplitude value; ssPERG SDph, standard deviation of steady-state PERG phase.
Figure 3Box-plot of amplitude values of steady-state perg in glaucoma patients, myopic subjects and normal control.Abbreviation: ssPERG Amp, steady-state pattern electroretinogram mean amplitude value.
Figure 4Box-plot of intrinsic variability of the phase of steady-state perg in glaucoma patients, myopic subjects and normal control. Abbreviation: ssPERG SDph, standard deviation of steady-state pattern electroretinogram phase.