Literature DB >> 31438734

Intrasession Repeatability of Biometric Measurements Obtained with a Low-Coherence Interferometry System in Pseudophakic Eyes.

Hideki Fukumitsu1,2,3, Vicent J Camps1, David P Piñero1,3.   

Abstract

Purpose: To evaluate the intrasession repeatability of the biometric measurements obtained with a low-coherence reflectometry optical biometer in pseudophakic eyes implanted with two different types of intraocular lens (IOL).
Methods: Prospective, single-center, comparative study including 69 eyes of 69 patients with ages ranging from 51 to 92 years. Previous uncomplicated cataract surgery had been performed in all patients 1 to 2 months before measurements, with implantation of the Acrysof SN60WF IOL in 35 eyes (35 patients, group 1) and the IOL Akreos MI60 in 34 eyes (34 patients, group 2). A complete postoperative ophthalmological examination was performed including three consecutive measurements with the "Aladdin" system from (Topcon, Japan). Intrasession repeatability of axial length (AXL), anterior chamber depth (ACD) and IOL thickness (IOLT) were assessed with the within-subject standard deviation (Sw), intraobserver precision (1.96 × Sw), coefficient of variation (CV) and intraclass correlation coefficient (ICC).
Results: The Sw for AXL measurements was 0.03 and 0.05 mm in groups 1 and 2, respectively, with ICC of 1.000 and 0.999 (CV: 0.14% and 0.22%) (p ≤ 0.031). Concerning pseudophakic ACD, the Sw was 0.03 and 0.09 mm in groups 1 and 2, respectively, with ICC of 0.992 and 0.956 (CV: 0.55% and 1.75%) (p ≤ 0.021). The variability of IOLT measurements was high in both groups, with Sw of 0.12 and 0.29 mm for groups 1 and 2 (p = .008), respectively, and ICC of 0.065 and 0.770 (CV: 20.84% and 62.39%).Conclusions: The optical biometer "Aladdin" (Topcon, Japan) provides consistent measurements of AXL and ACD in pseudophakic eyes. However, there is a limitation in the consistency of IOLT measurements that should be investigated further.

Entities:  

Keywords:  Aladdin; anterior chamber depth; axial length; optical biometry; pseudophakic

Year:  2019        PMID: 31438734     DOI: 10.1080/02713683.2019.1659976

Source DB:  PubMed          Journal:  Curr Eye Res        ISSN: 0271-3683            Impact factor:   2.424