| Literature DB >> 31436186 |
Rajesh S Joshi1, Shrutika A Chavan1.
Abstract
Purpose: To study the effect of rotation of intraocular lens (IOL) on posterior capsular opacification (PCO) in eyes with phacoemulsification.Entities:
Keywords: Intraocular lens; phacoemulsification; posterior capsular opacification
Mesh:
Year: 2019 PMID: 31436186 PMCID: PMC6727723 DOI: 10.4103/ijo.IJO_1854_18
Source DB: PubMed Journal: Indian J Ophthalmol ISSN: 0301-4738 Impact factor: 1.848
Figure 1Retro illumination image showing PCO
Figure 2Image having PCO imported to EPCO - Software
Figure 3PCO - Score of the image
Showing axial length in two groups
| Axial length (mm) | Rotation group | Control group |
|---|---|---|
| 15-16.99 | 02 | 03 |
| 17-18.99 | 05 | 06 |
| 19-20.99 | 09 | 08 |
| 21-22.99 | 22 | 24 |
| 23-24.99 | 10 | 09 |
| 25-26.99 | 01 | 0 |
| More than 27 | 01 | 0 |
| Total | 50 | 50 |
Showing morphological types of cataract and Posterior capsular opacification 2 groups
| Type of cataract | Rotation group | Number of patients with PCO | Control group | Number of patients with PCO |
|---|---|---|---|---|
| Cortical | 10 | 01 | 09 | 01 |
| Nuclear | 23 | 01 | 21 | 02 |
| Posterior sub capsular | 06 | 00 | 05 | 03 |
| Posterior polar cataract | 09 | 01 | 10 | 03 |
| Mixed | 02 | 01 | 05 | 01 |
| Total | 50 | 04 | 50 | 10 |
Median PCO score in 2 groups
| Duration (months) | Median PCO score (Rotation group) | Median PCO score (Control group) | |
|---|---|---|---|
| 6 | 0.15 | 0.22 | 0.02 |
| 12 | 0.13 | 0.23 | 0.03 |
| 24 | 0.22 | 0.25 | 0.04 |
| 36 | 0.23 | 0.23 | 0.2 |
PCO=Posterior capsular opacification. The median PCO score at 6,12, and 24 months was significantly low in the rotation group (0.15, 0.13,0.22) compared to the control group (0.22, 0.23, 0.25)
Comparative data on prevention of Posterior capsular opacification by various methods (Cochrane data base Findl et al.)
| Strategies on prevention of PCO* | Conclusion from the study |
|---|---|
| Influence of IOL optic material | No significant difference in the prevention of PCO in different IOL material (PMMA, hydrogel, hydrophobic acrylic, silicon). Although hydrogel IOL’s have higher and silicon IOL’s have lower PCO rate |
| Influence of IOL optic design | Significantly lower PCO score (−8.65 (−10.72-−6.59), scale 0 to 100) and Nd:YAG rate (0.19 (0.11-0.35)) in sharp edged than in round edged IOLs, however not between 1-piece and 3- piece IOLs. |
| Surgical techniques and drugs | No significant difference between different types of intraoperative/postoperative anti-inflammatory treatment except for treatment with an immunotoxin (MDX-A) which led to a significantly lower PCO rate. |
| Present study (360 degree IOL rotation) | Significant reduction in PCO and Nd:YAG capsulotomy rate. |