Emily Garnett1, Jieli Li2, Deepthi Rajapakshe1, Estella Tam1, Qing H Meng2, Sridevi Devaraj1.
Abstract
OBJECTIVES: Assessment of Vitamin D status by measurement of 25-Hydroxyvitamin D (25-OH-D) is widely performed by immunoassay. Yet, the ability of these assays to detect Vitamin D2 (as 25-OH-D2) or Vitamin D3 (as 25-OH-D3) varies. It is important to recognize the ability of an assay to quantitate either form of 25-OH-D to evaluate Vitamin D status of supplemented patients. We evaluated detection of 25-OH-D2 and 25-OH-D3 by two assays in our medical center. DESIGN AND METHODS: The Abbott Architect i1000 SR 25-OH Vitamin D assay and Roche Cobas 8000 Vitamin D assay were compared for their recovery of 25-OH-D2 or D3 from spiked serum samples. Samples with known endogenous concentrations of 25-OH-D2 or D3 by LC-MS/MS were also measured to calculate bias between our assays and LC-MS/MS.Entities:
Keywords: 25-Hydroxyvitamin D; Assay comparison; Immunoassay; Recovery; Vitamin D2
Year: 2019 PMID: 31428669 PMCID: PMC6695248 DOI: 10.1016/j.plabm.2019.e00130
Source DB: PubMed Journal: Pract Lab Med ISSN: 2352-5517
Recovery studies of 25-OH-D with spiked samples by Architect or Cobas assays.
| Sample | 25-OH-D species | [25-OH-D] added, nmol | Architect | Cobas | ||||
|---|---|---|---|---|---|---|---|---|
| Measured [25-OH-D], nmol/L | Expected [25-OH-D], nmol/L | Recovery | Measured [25-OH-D], nmol/L | Expected [25-OH-D], nmol/L | Recovery | |||
| Low | none | 0 | 35.38 ± 2.1 | 40.86 ± 3.3 | ||||
| Low | D2 | 130.93 | 86.81 ± 7.0 | 166.30 ± 2.1 | 40 ± 5% | 130.47 ± 11.7 | 171.79 ± 3.8 | 69 ± 10% |
| Low | D3 | 134.35 | 151.75 ± 14.5 | 169.72 ± 2.1 | 87 ± 10% | 161.42 ± 14.7 | 175.21 ± 3.8 | 90 ± 12% |
| Moderate | none | 0 | 82.56 ± 7.3 | 97.72 ± 3.6 | ||||
| Moderate | D2 | 130.93 | 132.94 ± 11.6 | 213.49 ± 7.3 | 39 ± 5% | 189.56 ± 17.1 | 228.64 ± 3.1 | 71 ± 15% |
| Moderate | D3 | 134.35 | 194.63 ± 9.9 | 216.91 ± 7.3 | 84 ± 2% | 217.12 ± 19.8 | 232.06 ± 3.1 | 90 ± 14% |
| High | none | 0 | 151.68 ± 4.2 | 183.68 ± 15.1 | ||||
| High | D2 | 130.93 | 198.19 ± 2.3 | 282.62 ± 4.2 | 37 ± 4% | >250 | 314.60 ± 18.4 | |
| High | D3 | 134.35 | 265.50 ± 6.1 | 286.03 ± 4.2 | 86 ± 4% | >250 | 318.02 ± 18.4 | |
Unable to calculate recovery due to measured values above instrument reportable range.
Fig. 1Bias plots of total 25-OH-D assay results relative to 25-OH-D2 or D3 by LC-MS/MS. Samples were measured in singlicate. Percent bias was calculated for Architect and Cobas assays for measurements of samples with A) known 25-OH-D3 or B) known 25-OH-D2 concentrations.