Dosimetric comparison between three- and four-dimensional computerised tomography radiotherapy for breast cancer.

Yanli Yan1, Zhou Lu2, Zi Liu1, Wei Luo1, Shuai Shao1, Li Tan1, Xiaowei Ma1, Jiaxin Liu1, Emmanuel Kwateng Drokow1, Juan Ren1.   

Abstract

At present, methods of radiotherapy simulation for breast cancer based on four-dimensional computerised tomography (4D-CT) or three-dimensional CT (3D-CT) simulation remain controversial. In the present study, 7 patients with residual breast tissue received whole breast radiotherapy based on 3D-CT and 4D-CT simulation. For the 4D-CT plan, four types of CT images were produced, including images of the end of inspiration and the end of expiration, and images acquired by the maximal intensity projection (MIP) and average intensity projection (AIP). In the 3D-CT plan, the clinical target volume (CTV) and plan target volume (PTV) were marginally higher compared with the 4D-CT plan. In addition, the minimum point dose of the target volume (D min), the maximum point dose of the target volume (D max) and the mean point dose of the target volume (D mean) of the CTV and PTV in the MIP and AIP plans were marginally higher compared with the 3D-CT plan. For the contralateral breast (C-B), volumes of the 4D-CT plan were markedly lower compared with the 3D-CT plan. Furthermore, D min, D max and D mean of the 3D-CT plan were higher compared with the AIP and MIP plans. For the ipsilateral lungs (I-L), volumes of the 3D-CT and AIP plans were higher compared with the MIP plan. Furthermore, when breast lesions were on the left side, for the heart, the volume receiving no less than 40% of the prescription dose (V 40) and the volume receiving no less than 30% of the prescription dose (V 30) of the MIP and AIP plans were slightly lower compared with those of the 3D plan. In conclusion, 4D-CT radiotherapy based on the MIP and AIP plans provides a slightly smaller radiation area and slightly higher radiotherapy dosage of the CTV and PTV compared with 3D-CT radiotherapy for breast radiotherapy. Therefore, the MIP and AIP plans prevent C-B radiation exposure and improve sparing of the heart and I-L.

Entities:  

Keywords:  average intensity projection; breast cancer; dosimetry; four-dimensional computerised tomography; maximal intensity projection; radiotherapy; three-dimensional computerised tomography

Year:  2019        PMID: 31423248      PMCID: PMC6607180          DOI: 10.3892/ol.2019.10467

Source DB:  PubMed          Journal:  Oncol Lett        ISSN: 1792-1074            Impact factor:   2.967


Introduction

The extensive use of whole breast radiotherapy has rendered it a standard treatment for patients with early breast cancer undergoing breast-conserving surgery (1–4). Previous advancements of individualised treatment for breast cancer have facilitated the gradual application of whole breast radiotherapy to neoadjuvant radiotherapy (5,6) and palliative radiotherapy (7,8), which have enhanced opportunities for surgical intervention, and improved the survival time and quality of life of patients with breast cancer. At present, radiotherapy techniques for breast cancer are typically based on three-dimensional computerised tomography (3D-CT) simulation. However, the 3D-CT plan overlooks the target volume and radiotherapy dosage bias induced by breathing movements. Reportedly, 3D-CT images depend on different respiration phases when the CT scan starts (9). Previous studies have established that the accuracy and efficiency of radiotherapy in the chest and abdomen can easily be reduced by motions of tumours and organs at risk (OAR) (10,11). To overcome these problems, a novel technological innovation, four-dimensional (4D)-CT radiotherapy technique, has emerged. Compared with conventional scanning, more accurate images of tumours and normal organs can be acquired by the 4D-CT process (11). At present, 4D-CT radiotherapy has been extensively applied in the treatment of lung cancer, liver cancer, oesophageal cancer, gastric cancer and kidney cancer; however, it has rarely been applied for breast cancer (11–19). For 4D-CT simulation, a breathing cycle is evenly divided into ten respiration phases and ten sets of CT images are acquired, respectively (10,11). Typically, the outline of targets and OAR on all ten sets of images is delineated by radiation oncologists, which increases the workload compared with 3D-CT radiotherapy (17,20,21). A number of composite methods have been used to decrease and optimise the work, including maximal intensity projection (MIP), average intensity projection (AIP), minimum intensity projection and two-extreme-phases fusion (21,22). While MIP and MIP-CT images are acquired by finding the maximum and minimum CT value along the slices at the same pixel location, AIP-CT images are acquired by averaging all CT values along the slices at the same pixel location (22,23). The two-extreme-phases fusion method involves delineating target outlines on CT images of the two extreme respiration phases (T00 for the end of inhalation and T50 for the end of exhalation) and then fusing them (21). Notably, MIP-CT images are effective for assessing the motion of the organ but not for determining the tumour boundary near the diaphragm and chest wall, which can be readily demonstrated by AIP-CT and MIP-CT images (22). For partial breast irradiation, previous studies have demonstrated that 4D-CT radiotherapy improves the target definition and decreases the radiation dose of OAR (18,19). Furthermore, certain previous studies have compared different composite methods for lung cancer and liver cancer (21,23–28); however, to the best of our knowledge, at present, there is no study that has been published in English for breast cancer. Therefore, the present study aimed to compare the target volumes and the dosimetric difference between the 3D-CT and 4D-CT plans for whole breast radiotherapy to determine the more effective radiotherapy technique for breast cancer.

Patients and methods

Patients and design

In the present study, seven female patients with breast cancer with residual breast tissue received whole breast radiotherapy based on 3D-CT and 4D-CT between March 2016 and April 2017 at The First Affiliated Hospital of Xi'an Jiaotong University (Xi'an, China). Inclusion criteria: i) Female patients were digonsed with breast cancer by pathology and and clinical examination; ii) patients were willing to accept and could tolerate breast radiotherapy; and iii) patients had no radiotherapeutic contraindication. Exclusion criteria: i) patients with metastatic or recurrent breast cancer; and ii) patients who had previously received chest radiotherapy. Table I summarises the characteristics of all patients. The present study was approved by The Ethics Committee of First Affiliated Hospital of Xi'an Jiaotong University (approval no. 2015-101). Written informed consent was obtained from all participants.
Table I.

Characteristics of the patients (n=7).

CharacteristicMedian (range)n
Age, years50.57 (39–76)
Disease stage
  Ia2
  IIa3
  IIb1
  IV1
Radiotherapy location
  Right breast4
  Left breast3
Type of radiotherapy
  Postoperative4
  Neoadjuvant2
  Palliative1
Surgical treatment
  Yes5
  No2
Plan target volume, cm3757.702 (500.37–1,063.08)

CT simulation

All seven patients were fixed in position using thermoplastic sheets or breast brackets. Subsequently, 3D-CT or 4D-CT scans were received sequentially with free breathing in the supine position. For each patient, images were obtained using a Philips Big Bore CT-Simulator (Philips Medical Systems, Inc., Bothell, WA, USA) with 5-mm slice thickness and a scan range from the submentum to the subphrenic, including the heart, bilateral breasts and bilateral lungs. All CT images were then uploaded and reconstructed on the Monaco 5.11.01 radiation treatment planning system (TPS) (ELEKTA Co., Sweden). In addition, the 4D-CT scan was acquired using the Cine model and supplemented by the real-time position management (RPM) system (Philips Co., Holland) during breathing. Notably, the scan time was >1 respiratory cycle. For each patient, the respiratory cycle was evenly divided into ten respiration phases by the RPM system. T00 was defined as the end of inhalation and T50 was defined as the end of exhalation. Furthermore, ten sets of 4D-CT images of the ten respiration phases were acquired, and the MIP-CT and AIP-CT images of each patient were fused and reconstructed.

Targets and OAR delineation, and dose prescription

All acquired CT images were uploaded and rebuilt on the Monaco 5.11.01 TPS. Clinical doctors delineated the outlines of target areas and OAR, and medical physicists formulated radiotherapy plans. In addition, all delineations and the five types of plans (3D, T00, T50, MIP and AIP) were separately implemented for each patient by the same skilled doctor and medical physicist.

Target delineation

The clinical target volume (CTV) consisted of the whole residual breast tissue. The upper and lower boundaries of the CTV indicated the edges of breast tissue, the inner boundary indicated the sternal line, and the outer boundary indicated the anterior axillary line. The anterior boundary was 5 mm below the skin surface and the posterior boundary was the ectopectoralis fascia. In addition, the plan target volume (PTV) was attained by adding 5-mm isotropic expansion of the CTV and the anterior boundary was refined 3 mm below the skin surface simultaneously.

OAR delineation

In the present study, the delineation method of the contralateral breast (C-B) was similar to the aformentioned method of the CTV. The heart was delineated from the right atrium and the right ventricle to the cardiac apex, excluding the pulmonary trunk, ascending aorta and vena cava. In addition, the right and left lungs were delineated by the automatic function of the Monaco TPS and manual modification. Furthermore, the spinal cord and bilateral humeral heads were delineated on all layers of the CT scans.

Plan evaluation

The present study used a dose volume histogram (DVH) to evaluate the quality of the radiation plan. For the CTV and PTV, D represents the minimum dose delivered to x% of the target volume and V represents the volume receiving no less than x% of the prescription dose (29,30). In addition, Dmin, Dmax and Dmean of the CTV and PTV represent the minimum, maximum and mean point dose of the target volume, respectively. The conformity index (CI) and the homogeneity index (HI) of the PTV were automatically evaluated by the Monaco TPS to assess the PTV coverage rate. The CI indicates the ratio between the PTV and the irradiated volume at the prescription dose, and the HI implies the uniformity of the dose distribution in the target volume (31). The computational formulas of the CI and HI were as follows: CI=TV12/TV × VR1, where TV1 represents the volume of the target that received the prescription dose, TV represents the target volume and VR1 represents the total volume of the prescription isodose. Notably, values of CI closer to 1.0 represent a better dose conformity of the PTV. HI=Dmax/Dmin, where Dmax represents the maximum point dose and Dmin represents the minimum point dose of the target volume. Notably, values of HI closer to 1.0 indicate a plan with less heterogeneity. Definitions of Dmin, Dmax and Dmean for all OAR and V for the C-B are similar to definitions for the target volume. Other V represents the volume receiving no less than x Gray (Gy) (29,30).

Dose prescription

Intensity-modulated radiotherapy (IMRT) plans were performed with 6-mV x-ray, and 5–9 coplanar and isocenter radiation treatment fields for five groups of each patient. Subsequently, a dose of 50 Gy in 25 fractions of 2 Gy was prescribed to the PTV. Notably, 95% of the target volume should be included by 95% of the prescribed dose (4,750 centigray, cGy) and not >5% should be encompassed by 105% of the prescribed dose (5,250 cGy). In the present study, the dose limits of OAR were as follows: For the ipsilateral lungs (I-L), V20<25% and Dmean<15Gy; for the bilateral lungs, V20<20%; and for the heart, V30<10% and V40<5%.

Statistical analysis

The disease stage of patients was evaluated according to the 7th American Association of Cancer (AJCC) staging system (32). D, CI and HI were extracted using the Monaco system. All data were analysed using SPSS software (version 21.0; IBM Corp., Armonk, NY, USA) with a randomised block design. The Shapiro-Wilk test and Levene test were used to evaluate the normality and homogeneity of data. Data that are normaly distributed are expressed as mean ± standard deviation and presented as bar plots; all other data are expressed as median (interquartile range) and presented as box and whisker plots (Figs. 1–9). For each evaluation index, analysis of variance followed by a Least Significant Difference test were used when data satisfied normal distribution and homoscedasticity. Otherwise, a Friedman test and pairwise comparison were used. P<0.05 was considered to indicate a statistically significant difference.
Figure 1.

CI and HI values for the 3D computerised tomography and four-dimensional computerised tomography plans in 7 patients. (A) CI values. (B) HI values. The horizontal axis represents the five different plans and the vertical axis represents the median values of the CI and HI. CI, conformity index; HI, homogeneity index; AIP, average intensity projection; MIP, maximal intensity projection; T00, end of inhalation; T50, end of exhalation; 3D, three-dimensional; Gy, gray.

Figure 9.

Parameters of the heart (breast lesions in the right) between the 3D and four-dimensional computerised tomography plans in seven patients. (A) Heart (lesions in the right breast)-Volume. (B) Heart (lesions in the right breast)-Dmin. (C) Heart (lesions in the right breast)-Dmax. (D) Heart (lesions in the right breast)-Dmean. The horizontal axis represents the five different plans and the vertical axis represents the mean or median dose size. 3D, three-dimensional; AIP, average intensity projection; MIP, maximal intensity projection; T00, end of inhalation; T50, end of exhalation; D, the minimum dose delivered to x% of the target volume; V, represents the volume receiving no less than x Gy; Dmax, maximum point dose; Dmin, minimum point dose of the target volume; Dmean, mean point dose of the target volume; Gy, gray.

Results

Dosage comparison of the CTV and PTV between the 3D-CT plan and the 4D-CT plan

The present study compared the dosimetric characteristics between the 3D-CT radiotherapy plan and four different 4D-CT radiotherapy plans (AIP, MIP, T00 and T50) in the same order for the seven patients with residual breast tissue. For target dose parameters, no significant difference was observed in the CI and HI of the PTV between the 3D-CT and 4D-CT plans (Table II; Fig. 1). In addition, target volumes (including the total volume, V100, V95, V90 and V50) of the CTV and PTV of the 4D-CT plan were slightly lower compared with the 3D-CT plan (Table II; Figs. 2 and 3). Furthermore, Dmin, Dmax and Dmean of the CTV and PTV in the MIP and AIP plans were slightly higher compared with that of the 3D-CT plan (Table II; Figs. 2 and 3).
Table II.

Comparisons of the CTV and PTV between the 3D-CT plan and 4D-CT plans.

A, Comparisons of the CTV between the 3D-CT plan and 4D-CT plans

4D-CT

Dose parameters3D-CTAIPMIPT00T50
Volume, cm3524.034±128.999509.609±132.411504.793±121.959507.669±125.227492.642±131.491
V100, cm3513.763±123.248502.839±130.326508.235±142.437501.892±123.518487.116±129.799
V95, cm3523.997±128.984470.608±79.413504.319±121.564507.525±125.144500.237±120.540
V90, cm3524.033±128.998509.376±132.225504.566±121.769507.623±125.200500.334±120.610
V50, cm3524.034±128.999509.610±132.409504.793±121.959507.669±125.227500.357±120.627
D95, cGy5,053.800 (22.900)5,046.400±18.6345,050.343±28.3395,051.371±24.4205,063.600 (359.900)
D90, cGy5,061.614±19.7075,072.657±14.5355,075.471±22.4745,076.157±21.6945,075.214±12.667
D50, cGy5,157.857±13.9895,168.357±20.0565,171.300±15.5925,170.800±16.4695,170.100 (30.800)
D5, cGy5,283.629±16.1165,309.086±42.3715,301.171±29.1225,299.757±19.3665,294.3 (41.800)
Dmin, cGy4,580.300 (398.200)4,813.900 (295.200)4,881.100 (79.400)4,755.500 (211.500)4,848.700 (169.700)
Dmax, cGy5,529.571±35.7635,561.643±49.2715,566.429±60.8865,543.700±54.5255,554.171±58.020
Dmean, cGy5,157.986±14.6125,173.371±19.3735,170.900 (43.700)5,172.243±16.6065,171.700 (6.800)

B, Comparisons of the PTV between the 3D-CT plan and 4D-CT plans

4D-CT

Dose parameters3D-CTAIPMIPT00T50

Volume, cm3776.249±166.189763.390±171.117754.455±162.964758.134±173.683736.284±185.678
V100, cm3703.954 (703.954)710.172 (147.870)722.857±152.885732.503±161.832706.833±172.300
V95, cm3771.304±165.881758.704±170.350747.790±161.510754.069±173.358730.658±184.550
V90, cm3776.115±166.052762.838±170.593753.770±162.274757.892±173.439735.909±185.307
V50, cm3776.249±166.189763.390±171.117754.455±162.964758.134±173.683736.284±185.678
D95, cGy5,022.000 (28.600)5,048.700 (37.300)5,035.400 (46.100)5,058.100 (38.700)5,038.200 (43.2000)
D90, cGy5,050.300 (26.600)5,058.400 (32.800)5,061.200 (36.700)5,084.400 (36.900)5,065.000 (35.100)
D50, cGy5,151.614±19.8835,176.000 (41.500)5,165.229±21.1775,166.171±21.2925,169.214±18.955
D5, cGy5,286.186±17.4355,312.900±37.3725,305.486±28.6465,304.314±21.1775,302.400 (21.400)
Dmin, cGy3,902.786±553.1583,999.829±570.4324,139.871±612.8714,043.829±445.6343,898.843±583.582
Dmax, cGy5,580.943±32.9605,602.914±42.0955,596.357±41.6325,628.786±74.5225,608.300±56.299
Dmean, cGy5,149.371±23.2925,165.586±23.3525,162.686±26.8615,165.514±25.1065,167.086±21.546
CI0.840 (0.080)0.840 (0.050)0.840 (0.060)0.840 (0.050)0.830 (0.040)
HI1.050 (0)1.050 (0.020)1.050 (0.010)1.050 (0)1.050 (0.030)

Data that are normaly distributed are expressed as mean ± standard and all other data are expressed as median (interquartile range). 3D-CT, three-dimensional computed tomography; 4D-CT, four-dimensional computed tomography; AIP, average intensity projection; CI, conformity index; CTV, clinical target volume; Dmax, maximum point dose of the target volume; Dmean, mean point dose of the target volume; Dmin, minimum point dose of the target volume; D, minimum dose delivered to x% of the target volume; V, the volume receiving no less than x% of the prescription dose; HI, homogeneity index; PTV, plan target volume; cGy, centigray.

Figure 2.

Parameters of the CTV between the 3D and four-dimensional computerised tomography plans in seven patients. (A) CTV–Volume. (B) CTV–V100. (C) CTV–V95. (D) CTV–V90. (E) CTV–V50. (F) CTV-D95. (G) CTV-D90. (H) CTV-D50. (I) CTV-D5. (J) CTV-Dmin. (K) CTV-Dmax. (L) CTV-Dmean. The horizontal axis represents five different plans and the vertical axis represents the mean or median dose size. CTV, clinical target volume; 3D, three-dimensional; AIP, average intensity projection; MIP, maximal intensity projection; T00, end of inhalation; T50, end of exhalation; D, the minimum dose delivered to x% of the target volume; V, the volume receiving no less than x% of the prescription dose; Dmax, maximum point dose; Dmin, minimum point dose of the target volume; Dmean, mean point dose of the target volume; Gy, gray.

Figure 3.

Parameters of the PTV between the 3D and four-dimensional computerised tomography plans in seven patients. (A) PTV–Volume. (B) PTV–V100. (C) PTV–V95. (D) PTV–V90. (E) PTV–V50. (F) PTV-D95. (G) PTV-D90. (H) PTV-D50. (I) PTV-D5. (J) PTV-Dmin. (K) PTV-Dmax. (L) PTV-Dmean. The horizontal axis represents the five different plans and the vertical axis represents the mean or median dose size. PTV, plan target volume; 3D, three-dimensional; AIP, average intensity projection; MIP, maximal intensity projection; T00, end of inhalation; T50, end of exhalation; D, the minimum dose delivered to x% of the target volume; V, the volume receiving no less than x% of the prescription dose; Dmax, maximum point dose; Dmin, minimum point dose of the target volume; Dmean, mean point dose of the target volume; Gy, gray.

Dosage comparison of OAR between the 3D-CT and 4D-CT plans

In the present study, the following comparisons of dose parameters of OAR were made. For the C-B (Table III; Fig. 4), the total volume of the 4D-CT plan was markedly lower compared with the 3D-CT plan. In addition, Dmin, Dmax and Dmean of the AIP plan were lower than those of 3D-CT and MIP plans. No marked differences were observed in dose parameters between the MIP and AIP plans. For the I-L (Table III; Fig. 5), volumes (including the total volume, V30, V20, and V10) of the T00 plan were the highest, followed by the 3D-CT and AIP plans, and the volumes of the T50 and MIP plans were the lowest. For the contralateral lungs (C-L) (Table III; Fig. 6), the total volumes of the MIP plan were markedly lower compared with that of the 3D-CT, AIP and T00 plans; however, the volume, of the T00 plan were higher compared with the T50 plan. For the I-L and C-L (Table III; Figs. 5 and 6), no statistical differences were observed in the dosage among the five plans. In addition, for the contralateral and ipsilateral humeral head (Table III; Fig. 7), no significant differences were observed in dose parameters between the 3D-CT and 4D-CT plans. For the heart (Table III; Figs. 8 and 9), regardless of whether breast lesions were in the right or left side, the volume of the MIP and AIP plans were slightly higher compared with that of the 3D-CT plan, with no significant difference in dose among the 3D-CT, MIP and AIP plans. When breast lesions were on the left side, for the heart, V40 and V30 of the MIP and AIP plans were slightly lower compared with those of the 3D-CT plan (Table III; Fig. 8).
Table III.

Comparison of organs at risk between the 3D-CT plan and 4D-CT plans.

A, Comparison of contralateral breast between the 3D-CT plan and 4D-CT plans

4D-CT
Dose parameters3D-CTAIPMIPT00T50
Volume, cm3443.632±125.842[bcde]412.736±123.953[a]420.366±119.146[a]415.838±122.482[a]405.546±122.296[a]
V100, cm30 (0)0 (0)0 (0)0 (0)0 (0)
V95, cm30 (0)0 (0)0 (0)0 (0)0 (0)
V90, cm30 (0)0 (0)0 (0)0 (0)0 (0)
V50, cm30.051 (2.025)0 (1.492)0 (2.918)1.079±1.3550 (1.174)
D95, cGy235.100±85.804222.943±60.973218.500±78.801256 (127.100)249.943±109.144
D90, cGy268.471±88.399259.257±65.953252.157±86.770270.857±69.291292.329±118.998
D50, cGy468.743±110.044464.371±107.091453.971±143.022499.000±123.028514.757±193.549
Dmin, cGy130.000±60.833118.757±28.180125.257±50.198137.414±44.952145.243±72.691
Dmax, cGy3,095.100±1,134.2382,985.429±1,126.1333,097.929±1,097.6163,169.829±1,202.6512,606.800±1,223.300
Dmean, cGy552.814±160.896548.000±135.008552.214±153.072585.171±153.823576.814±224.615

B, Comparison of contralateral lungs between the 3D-CT plan and 4D-CT plans

4D-CT

Dose parameters3D-CTAIPMIPT00T50

Volume, cm31,277.545±299.910[c]1,282.555±316.948[c]1,160.681±303.191[abd]1,292.970±305.323[c]1,192.409±315.200
V30, cm30 (0.028)0 (0.114)0 (0)0 (0.029)0 (0.046)
V20, cm30.033 (8.205)0.430 (10.358)0 (3.852)0.767 (10.545)0.023 (8.181)
V10, cm3104.212±94.178100.521±74.60580.887±71.87899.002±79.668120.292±124.364
V5, cm3571.566±215.984547.182±223.574467.601±261.863591.245±291.803533.447±282.082
Dmin, cGy70.314±47.18358.886±29.40061.743±33.86154.614±32.90962.186±33.771
Dmax, cGy2,411.371±926.2602,475.143±861.4172,098.871±796.7632,340.914±815.5782,291.971±923.247
Dmean, cGy506.500 (48.600)506.786±69.115478.943±114.060498.743±89.702522.500±137.952

C, Comparison of ipsilateral lungs between the 3D-CT plan and 4D-CT plans

4D-CT

Dose parameters3D-CTAIPMIPT00T50

Volume, cm31,293.366±296.783[cde]1,283.719±304.658[cde]1,184.856±269.076[abd]1,373.630±315.333[abce]1,207.775±273.286[abd]
V30, cm3221.643±38.121[cde]228.896±46.964[ce]203.222±40.910[abd]238.280±43.928[ace]205.054±43.384[abd]
V20, cm3314.498±66.481[cde]316.125±62.331[ce]286.039±62.517[abd]334.568±65.860[ace]289.286±60.105[abd]
V10, cm3486.178±117.650[ce]473.562±111.914436.175±101.303[ad]506.874±103.594[ce]446.127±86.055[ad]
V5, cm3729.099±165.369[c]642.457 (174.459)[e]587.799 (125.114)[ad]673.485 (119.525)[ce]591.269 (101.470)[bd]
Dmin, cGy82.371±51.20479.386±47.89384.300±49.68268.314±50.05381.357±54.990
Dmax, cGy5,284.614±76.1285,283.543±62.5395,315.743±120.1795,305.214±79.7385,288.486±53.484
Dmean, cGy1,281.600 (211.200)1,267.300 (63.000)1,275.700 (74.700)1,259.200 (143.300)1,275.400 (150.300)

D, Comparison of contralateral humeral head between the 3D-CT plan and 4D-CT plans

4D-CT

Dose parameters3D-CTAIPMIPT00T50

Volume, cm328.462±10.35725.734±10.14126.783±11.03725.989±10.21526.011±9.347
V30, cm30 (0)0 (0)0 (0)0 (0)0 (0)
V20, cm30 (0)0 (0)0 (0)0 (0)0 (0)
V10, cm30 (0)0 (0)0 (0)0 (0)0 (0)
V5, cm30 (0)0 (0)0 (0)0 (0)0 (0)
Dmin, cGy22.800 (81.300)43.671±25.40635.971±20.88846.786±33.60033.071±23.574
Dmax, cGy84.3 (251.200)217.500±146.926194.157±170.316214.600±181.153248.014±171.875
Dmean, cGy44.300 (212.500)117.957±96.78887.600 (75.500)69.200 (171.700)139.329±112.214

E, Comparison of Ipsilateral humeral head between the 3D-CT plan and 4D-CT plans

4D-CT

Dose parameters3D-CTAIPMIPT00T50

Volume, cm326.099±12.28719.600 (21.870)26.470±13.64025.693±12.29825.506±12.668
V30, cm30 (0)0 (0)0 (0)0 (0)0 (0)
V20, cm30 (0)0 (0)0 (0)0 (0)0 (0)
V10, cm30 (0)0 (0)0 (0)0 (0)0 (0)
V5, cm30 (4.104)0 (10.989)0 (0.103)0 (1.498)0 (4.481)
Dmin, cGy66.3 (72.900)52.000 (145.500)57.900 (56.300)57.700 (77.200)68.000 (77.500)
Dmax, cGy190.500 (873.600)408.200 (1230.20)363.300 (385.700)217.800 (690.300)420.800 (752.200)
Dmean, cGy108.300 (253.000)113.100 (562.100)98.800 (96.400)95.400 (170.100)113.400 (285.000)

F, Comparison of Heart (lesions in the left breast) between the 3D-CT plan and 4D-CT plans

4D-CT

Dose parameters3D-CTAIPMIPT00T50

Volume, cm3536.802±57.042552.668±111.836597.627±144.324550.867±100.890574.107±117.065
V40, cm324.698±9.71416.812±3.54621.360±3.06619.028±3.06616.127±6.813
V30, cm353.045±21.85243.779±2.34845.758±5.37943.508±4.89642.364±9.319
Dmin, cGy167.767±36.235168.533±17.404154.933±49.980145.667±47.388171.067±31.825
Dmax, cGy5,171.133±97.1845,135.067±129.8535,262.267±80.1235,246.333±109.4705094.500±109.390
Dmean, cGy1,049.933±249.937929.733±62.684966.733±76.793953.200±70.634975.767±59.327

G, Comparison of Heart (lesions in the right breast) between the 3D-CT plan and 4D-CT plans

4D-CT

Dose parameters3D-CTAIPMIPT00T50

Volume, cm3579.013 (297.004)655.683±111.829606.943 (197.833)633.784±104.173576.575 (207.698)
V40, cm30 (0.040)0 (0)0 (0)0 (0)0.051(0.281)
V30, cm31.557±1.8011.382±2.0781.151±1.6232.452±3.1162.120±1.993
Dmin, cGy155.825±72.060163.850±44.906163.925±71.999153.350±39.108160.125±72.390
VDmax, cGy3,350.775±758.4033,313.925±541.5153,262.275±808.2233,658.650±588.1563,764.050±865.463
Dmean, cGy839.850±83.354809.775±92.947745.275±169.622809.375±44.920852.950±128.064

P<0.05 vs. 3D-CT plan

P<0.05 vs. AIP plan

P<0.05 vs. MIP plan

P<0.05 vs. T00 plan

P<0.05 vs. T50 plan. Data that are normaly distributed are expressed as mean ± standard and all other data are expressed as median (interquartile range). 3D-CT, three-dimensional computed tomography; 4D-CT, four-dimensional computed tomography; AIP, average intensity projection; CI, conformity index; CTV, clinical target volume; Dmax, maximum point dose of the target volume; Dmean, mean point dose of the target volume; Dmin, minimum point dose of the target volume; D, minimum dose delivered to x% of the target volume; V, the volume receiving no less than x Gy; cGy, centigray; Gy, gray.

Figure 4.

Parameters of the contralateral breast between the 3D and four-dimensional computerised tomography plans in seven patients. (A) Contralateral Breast-Volume. (B) Contralateral Breast-D95. (C) Contralateral Breast-D90. (D) Contralateral Breast-D50. (E) Contralateral Breast-Dmin. (F) Contralateral Breast-Dmax. (G) Contralateral Breast-Dmean. The horizontal axis represents the five different plans and the vertical axis represents the mean or median dose size. 3D, three-dimensional; AIP, average intensity projection; MIP, maximal intensity projection; T00, end of inhalation; T50, end of exhalation; D, the minimum dose delivered to x% of the target volume; V, the volume receiving no less than x% of the prescription dose; Dmax, maximum point dose; Dmin, minimum point dose of the target volume; Dmean, mean point dose of the target volume; Gy, gray.

Figure 5.

Parameters of the ipsilateral lungs between the 3D and four-dimensional computerised tomography plans in seven patients. (A) Ipsilateral Lungs-Volume. (B) Ipsilateral Lungs-V30. (C) Ipsilateral Lungs-V20. (D) Ipsilateral Lungs-V10. (E) Ipsilateral Lungs-V5. (F) Ipsilateral Lungs-Dmin. (G) Ipsilateral Lungs-Dmax. (H) Ipsilateral Lungs-Dmean. The horizontal axis represents the five different plans and the vertical axis represents the mean or median dose size. 3D, three-dimensional; AIP, average intensity projection; MIP, maximal intensity projection; T00, end of inhalation; T50, end of exhalation; D, the minimum dose delivered to x% of the target volume; V represents the volume receiving no less than x Gy; Dmax, maximum point dose; Dmin, minimum point dose of the target volume; Dmean, mean point dose of the target volume; Gy, gray.

Figure 6.

Parameters of the contralateral lungs between the 3D and four-dimensional computerised tomography plans in seven patients. (A) Contralateral Lungs-Volume. (B) Contralateral Lungs-V20. (C) Contralateral Lungs-V10. (D) Contralateral Lungs-V5. (E) Contralateral Lungs-Dmin. (F) Contralateral Lungs-Dmax. (G) Contralateral Lungs-Dmean. The horizontal axis represents the five different plans and the vertical axis represents the mean or median dose size. 3D, three-dimensional; AIP, average intensity projection; MIP, maximal intensity projection; T00, end of inhalation; T50, end of exhalation; D, the minimum dose delivered to x% of the target volume; V, represents the volume receiving no less than x Gy; Dmax, maximum point dose; Dmin, minimum point dose of the target volume; Dmean, mean point dose of the target volume; Gy, gray.

Figure 7.

Parameters of the contralateral and ipsilateral humeral head between the 3D and four-dimensional computerised tomography plans in seven patients. (A) Contralateral Humeral Head-Volume. (B) Contralateral Humeral Head-Dmin. (C) Contralateral Humeral Head-Dmax. (D) Contralateral Humeral Head-Dmean. (E) Ipsilateral Humeral Head-Volume. (F) Ipsilateral Humeral Head-Dmin. (G) Ipsilateral Humeral Head-Dmax. (H) Ipsilateral Humeral Head-Dmean. The horizontal axis represents the five different plans and the vertical axis represents the mean or median dose size. 3D, three-dimensional; AIP, average intensity projection; MIP, maximal intensity projection; T00, end of inhalation; T50, end of exhalation; D, the minimum dose delivered to x% of the target volume; V represents the volume receiving no less than x Gy; Dmax, maximum point dose; Dmin, minimum point dose of the target volume; Dmean, mean point dose of the target volume; Gy, gray.

Figure 8.

Parameters of the heart (breast lesions in the left) between the 3D and four-dimensional computerised tomography plans in seven patients. (A) Heart (lesions in the left breast)-Volume. (B) Heart (lesions in the left breast)-V40. (C) Heart (lesions in the left breast)-V30. (D) Heart (lesions in the left breast)-Dmin. (E) Heart (lesions in the left breast)-Dmax. (F) Heart (lesions in the left breast)-Dmean. The horizontal axis represents the five different plans and the vertical axis represents the mean or median dose size. 3D, three-dimensional; AIP, average intensity projection; MIP, maximal intensity projection; T00, end of inhalation; T50, end of exhalation; D, the minimum dose delivered to x% of the target volume; V, represents the volume receiving no less than x Gy; Dmax, maximum point dose; Dmin, minimum point dose of the target volume; Dmean, mean point dose of the target volume; Gy, gray.

Discussion

Previous studies have established that radiotherapy based on 4D-CT simulation enhances the accuracy of dosage and determines the locations of tumour(s) and OAR in the chest and abdomen (9–26,33). Certain previous studies have compared the composite methods of 4D-CT radiotherapy for lung cancer and liver cancer (9,17,20,23–26,33); however, at present, to the best of our knowldege, there is no study that has been published in English, which has investigated the same for breast cancer. To the best of the our knowledge, the present study is the first to compare the target volume and dosimetric differences between the 3D-CT radiotherapy plan and four 4D-CT radiotherapy plans (T00, T50, MIP and AIP plans) for whole breast radiotherapy. For dose parameters of targets, all target volumes (including the total volume, V100, V95, V90 and V50) of the CTV and PTV in the 3D-CT plan were slightly higher compared with four 4D-CT plans, with no difference between the T00 and T50 plans. For the C-B, the total volume in the 3D-CT plan was markedly higher compared with the 4D-CT plan. Therefore, it could be concluded that breast tissues exert little impact on the respiration movement, which suggests that the volume changes of the breasts should be ignored (34). In addition, differences in target volumes between the 3D-CT and 4D-CT plans were not due to respiration movement-caused displacement. In free breathing, partial target areas of the 3D-CT plan were scanned recurrently or missed due to uncertainty in the beginning time of the scan and the whole scan time was longer compared with the 4D-CT scan; therefore, 4D-CT images evaluated the location, size and shape of targets more precisely (35). However, Bedi et al (36) reported no significant differences in the left breast volume contoured on a 3D-CT scan (1,005±559 cm3), 4D-CT scan with full inspiration phase (1,019±563 cm3), 4D-CT scan with full expiration phase (1,023±573 cm3) and 4D-CT scan derived by AIP (1023±573 cm3). The difference between the two studies could be attributed to the following: i) Breast locations in western females are more easily affected by the respiration movement, as the breast size of western females is typically larger compared with Chinese females; and ii) all breast lesions in the previous study by Bedi et al (36) were in the left chest where the impact of the cardiac motion could not be ignored; however, in the present study, the number of lesions on the right side was greater than that on the left side (right-to-left ratio, 4:3) Respiration movements not only affect displacements of target volumes but also exert a considerable impact on the radiation dose distribution (37–39). In the clinical setting, during the process of treatment, changes in the dose distribution caused by target bias in the 3D-CT plan cannot satisfy the treatment requirements, which may increase the risk of local recurrence. In the present study, Dmin, Dmax and Dmean of the CTV and PTV in the MIP and AIP plans were slightly higher compared with that of the 3D-CT plan. However, Dmin, Dmax and Dmean of the C-B in the 3D-CT and T00 plans were the highest and those of the AIP and MIP plans were the lowest. In addition, no marked difference was observed in dose parameters between the MIP and AIP plans, indicating that the dose distribution of the MIP and AIP plans is better compared with that of the 3D-CT plan. In addition, this suggests that the MIP and AIP plans can prevent and decrease radiation exposure to normal breast. In the present study, no marked differences were observed in the CI and HI between the 3D-CT and four 4D-CT plans. As the target volumes in the five groups were similar, all plans were achieved by IMRT with the same pattern. Reportedly, the IMRT technique achieved improved CI and HI compared with 3D conformal radiotherapy, and increased OAR sparing and decreased the late effects, which enhanced the quality of life of the patients (31,40,41). Previous studies have compared different composite methods for lung cancer and liver cancer (21,23–28). Zhao et al (27) reported that lung volumes of the AIP plan were close to the 3D-CT volume in lung cancer and that of the MIP plan was smaller compared with the AIP plan by 11.4±2.3%. In addition, the DVH of the MIP plan revealed that the MIP plan was less sensitive to breathing movements (27). However, Simon et al (28) reported that the internal target volume of the MIP plan was closer to the actual volume compared with that of the AIP plan. Similarly, a previous study determined that the GTV of the MIP plan was markedly higher compared with that of the AIP plan for stereotactic body radiotherapy planning in lung cancer (20). The differences in the aforementioned studies may have resulted from different evaluated standard and research methods. MIP images are acquired by finding the maximum CT value of images from all respiratory phases. A previous study reported that MIP images include the motion and position extent of a lung tumour as the density of a tumour was higher compared with that of the surrounding normal lung tissue (42). However, Mohatt et al (25) reported that in clinical lung tumour cases with displacements ranging between 0.1 and 2.2 cm, the MIP plan typically underestimated target volumes and resulted in a PTV ratio of 0.95±0.15. When tumours were located close to the chest wall, MIP images were more easily affected by the surrounding structures with high or equal density, including the cartilage tissues and muscles, compared with AIP images (22,27). Park et al (43) reported that the MIP plan was markedly different from the ten-phase fusion plan when breathing was irregular or a tumour was close to similar-density tissues. However, in the present study, no marked dosimetric difference was observed between the MIP and AIP plans, which could be attributed to the fact that breast tissue movements are less sensitive to breathing; therefore, the impact of MIP images caused by surrounding structures was low. In addition to improving the sparing of the C-B, another objective of breast radiotherapy is to decrease the irradiation dose and volume of the heart and bilateral lungs. In the present study, for the I-L, marked differences were identified in volumes (including the total volume, V30, V20, V10 and V5) among the five plans (3D-CT, T00, T50, MIP and AIP). Additionally, volumes of the T00 plan were the highest, followed by the 3D-CT and AIP plans, and those of the T50 and MIP plans were the lowest. For the C-L, total volumes of the MIP plan were markedly lower compared with those of the 3D-CT, AIP and T00 plans, and total volumes of the T00 plan were higher compared with those of the T50 plan. Whether in the I-L or C-L, no statistical difference of dosage was observed among the five plans, indicating that in free breathing, although no apparent breast displacements occurred between the two extreme respiratory phases, the sufficiently apparent difference of bilateral lungs volumes resulting from thoracic movements were easily observed. Total volumes of the lungs in the MIP plan were easily affected by the surrounding structures with high or equal density (22,27), which induced smaller volumes in the MIP plan compared with others and then induced the smallest V30, V20, V10 and V5. Therefore, the present study could not completely establish that the MIP plan is superior to the others. For the heart, dosimetric parameters of plans are affected not only by respiratory movements but also by their own rhythm. In the present study, whether breast lesions were in the right or left side, heart volumes of the MIP and AIP plans were slightly higher compared with that of the 3D-CT plan, with no marked differences in dose among the 3D-CT, MIP and AIP plans. However, for hearts of patients with lesions in the left breast, V40 and V30 of the MIP and AIP plans were slightly lower compared with that of the 3D-CT plan. These results indicated that the MIP and AIP plans may improve sparing of the heart, particualy lesions in the left side, and the I-L. However, Bedi et al (36) reported that dosimetric results for the heart and the I-L exhibited no statistically significant differences between the 3D-CT and 4D-CT plans for patients with left-sided breast cancer, and that improved sparing of the heart and the lungs could only be attained by decreasing the posterior margins of the breast target volumes. For contralateral and ipsilateral humeral heads, no marked differences were observed in dose parameters between the 3D-CT and 4D-CT plans as humeral heads were far away from the targets and could not be affected by therapy plans. In conclusion, for whole breast radiotherapy of breast cancer with residual tissues (including postoperative radiotherapy, neoadjuvant radiotherapy and palliative radiotherapy), 4D-CT radiotherapy techniques based on the MIP and AIP plans provide a slightly smaller radiation area and slightly higher radiotherapy dosage of the CTV and PTV compared with 3D-CT radiotherapy. For the C-B, the dose distribution in the MIP and AIP plans is better compared with the 3D-CT plan; therefore, MIP and AIP plans prevent and reduce radiation exposure to normal breast. The MIP and AIP plans also improve sparing of the heart (particularly breast lesions in the left side) and the I-L. Furthermore, the dosimetric differences between the MIP and AIP plans are not significant. Therefore, these plans are worth considering for whole breast radiotherapy.
  39 in total

1.  Dose correlation for thoracic motion in radiation therapy of breast cancer.

Authors:  Meisong Ding; Jinsheng Li; Jun Deng; Eugene Fourkal; C M Ma
Journal:  Med Phys       Date:  2003-09       Impact factor: 4.071

2.  Dosimetry and radiobiologic model comparison of IMRT and 3D conformal radiotherapy in treatment of carcinoma of the prostate.

Authors:  Gary Luxton; Steven L Hancock; Arthur L Boyer
Journal:  Int J Radiat Oncol Biol Phys       Date:  2004-05-01       Impact factor: 7.038

3.  Impact of breathing motion on whole breast radiotherapy: a dosimetric analysis using active breathing control.

Authors:  Robert C Frazier; Frank A Vicini; Michael B Sharpe; Di Yan; Julie Fayad; Kathy L Baglan; Larry L Kestin; Vincent M Remouchamps; Alvaro A Martinez; John W Wong
Journal:  Int J Radiat Oncol Biol Phys       Date:  2004-03-15       Impact factor: 7.038

4.  Comparison of helical, maximum intensity projection (MIP), and averaged intensity (AI) 4D CT imaging for stereotactic body radiation therapy (SBRT) planning in lung cancer.

Authors:  Jeffrey D Bradley; Ahmed N Nofal; Issam M El Naqa; Wei Lu; Jubei Liu; James Hubenschmidt; Daniel A Low; Robert E Drzymala; Divya Khullar
Journal:  Radiother Oncol       Date:  2006-11-20       Impact factor: 6.280

5.  Dosimetric impact of intrafraction motion for compensator-based proton therapy of lung cancer.

Authors:  Li Zhao; George A Sandison; Jonathan B Farr; Wen Chien Hsi; X Allen Li
Journal:  Phys Med Biol       Date:  2008-06-03       Impact factor: 3.609

6.  Improvement of the cine-CT based 4D-CT imaging.

Authors:  Tinsu Pan; Xiaojun Sun; Dershan Luo
Journal:  Med Phys       Date:  2007-11       Impact factor: 4.071

7.  Randomized phase III study comparing Best Supportive Care to Biafine as a prophylactic agent for radiation-induced skin toxicity for women undergoing breast irradiation: Radiation Therapy Oncology Group (RTOG) 97-13.

Authors:  J Fisher; C Scott; R Stevens; B Marconi; L Champion; G M Freedman; F Asrari; M V Pilepich; J D Gagnon; G Wong
Journal:  Int J Radiat Oncol Biol Phys       Date:  2000-12-01       Impact factor: 7.038

8.  The effect of respiratory motion variability and tumor size on the accuracy of average intensity projection from four-dimensional computed tomography: an investigation based on dynamic MRI.

Authors:  Jing Cai; Paul W Read; Ke Sheng
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

9.  The intrafraction motion induced dosimetric impacts in breast 3D radiation treatment: a 4DCT based study.

Authors:  Ning J Yue; Xiang Li; Sushil Beriwal; Dwight E Heron; Marc R Sontag; M Saiful Huq
Journal:  Med Phys       Date:  2007-07       Impact factor: 4.071

10.  Initial evaluation of a four-dimensional computed tomography system using a programmable motor.

Authors:  Luc Simon; Philippe Giraud; Vincent Servois; Jean-Claude Rosenwald
Journal:  J Appl Clin Med Phys       Date:  2006-11-28       Impact factor: 2.102

View more
  1 in total

Review 1.  Myocardial Ischemia Related to Common Cancer Therapy-Prevention Insights.

Authors:  Minerva Codruta Badescu; Oana Viola Badulescu; Dragos Viorel Scripcariu; Lăcrămioara Ionela Butnariu; Iris Bararu-Bojan; Diana Popescu; Manuela Ciocoiu; Eusebiu Vlad Gorduza; Irina Iuliana Costache; Elena Rezus; Ciprian Rezus
Journal:  Life (Basel)       Date:  2022-07-12
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.