| Literature DB >> 31405991 |
Naoki Kato1, Ichiro Yuki1,2, Toshihiro Ishibashi1, Ayako Ikemura1, Issei Kan1, Kengo Nishimura1, Tomonobu Kodama1, Syougo Kaku1, Yukiko Abe3, Katharina Otani4, Yuichi Murayama1.
Abstract
PURPOSE: We used an imaging technique based on 3-dimensional (3D) C-arm CT to assess the apposition of three types of stents after coiling of intracranial aneurysms.Entities:
Keywords: CT; aneurysm; angiography; coil; stent
Mesh:
Year: 2019 PMID: 31405991 PMCID: PMC7029241 DOI: 10.1136/neurintsurg-2019-014966
Source DB: PubMed Journal: J Neurointerv Surg ISSN: 1759-8478 Impact factor: 5.836
Figure 1Representative images. Dual volume 3D fusion images show good stent apposition into the siphon of the left internal carotid artery (ICA) after stent-assisted coiling using Neuroform EZ for the IC-superior hypophysial artery aneurysm: (A) half lucent and (B) translucent imaging. ‘Crescent sign’ malapposition between the stent trunk of an Enterprise2 stent and the anterior wall of the ICA siphon is clearly depicted (arrow heads): (C) half lucent and (D) translucent imaging. ‘Edge malapposition’ floating distal edge of a Neuroform EZ stent is shown (arrow): (E) half lucent and (F) translucent imaging. Distance was measured using the electronic caliper of the workstation (G). The angle between the proximal and distal parent vessels was measured using the protractor function of SYNAPS work station (H).
Patient and aneurysm characteristics
| Total | Type of stent | P value | |||
| Enterprise2 | Neuroform EZ | Neuroform Atlas | |||
| Sample size | 75 | 22 | 26 | 27 | |
| Age (years) | 59 (10.8) | 58 (9.7) | 58 (10.2) | 61 (12) | 0.57 |
| Women | 57 (76) | 15 (68) | 22 (85) | 20 (74) | 0.40 |
| Location of aneurysm | 0.006 | ||||
| Internal carotid artery | 51 | 10 | 23 | 18 | |
| Other vessels | 24 | 12 | 3 | 9 | |
| Ruptured aneurysm | 2 | 1 | 1 | ||
| Maximum diameter of aneurysm (mm) | 6.8 (3.0) | 7.5 (4.2) | 6.8 (2.8) | 6.2 (1.7) | 0.13 |
| Diameter of parent vessel (mm) | 3.4 (0.6) | 3.2 (0.7) | 3.6 (0.6) | 3.4 (0.6) | 0.08 |
| Angle between parent vessels (degree) | 82.7 (42.0) | 90.4 (46.9) | 80.9 (38.3) | 78.3 (45.5) | 0.61 |
| Stent diameter (mm) | 4.1 (0.47) | 4.1 (0.2) | 4.3 (0.3) | 3.9 (0.7) | 0.03 |
| Stent length (mm) | 21.4 (3.5) | 23.6 (4.8) | 20.0 (3.5) | 21.0 (0.0) | 0.001 |
Values are n (%) or mean (SD).
Other vessels include anterior cerebral artery (n=5), middle cerebral artery (n=2), vertebral basilar artery (n=16), and posterior cerebral artery (n=1).
Stent malapposition by stent type and location
| Overall | Type of stent | P value | |||
| Enterprise2 | Neuroform EZ | Neuroform Atlas | |||
| Crescent sign | 8 (11) | 6 (27) | 2 (8) | 0.007 | |
| Internal carotid artery | 7 | 5 | 2 | ||
| Other vessels | 1 | 1 | |||
| Edge malapposition | 29 (39) | 6 (27) | 15 (58) | 8 (30) | 0.05 |
| Internal carotid artery | 25 | 3 | 14 | 8 | |
| Other vessels | 4 | 3 | 1 | ||
| Distance between stent and vessel wall, mm | 1.0 (0.5) | 1.2 (0.4) | 0.6 (0.2) | 0.14 | |
Values are n (%) or mean (SD).
Other vessels include anterior cerebral artery (n=5), middle cerebral artery (n=2), vertebral basilar artery (n=16), and posterior cerebral artery (n=1).
Figure 2Illustrative images of stent apposition. Case 1: Translucent view of the fusion image shows marked malapposition and the tortuous structure of the stent struts of the Enterprise2: (A) half lucent and (B) translucent imaging. Case 2: All stent struts and edge markers of a Neuroform Atlas stent are seen. Although the visibility of the stent is slightly blurred, good apposition of all stent struts was confirmed: (C) half lucent and (D) translucent imaging.