Jos IJspeert1, Hans C J W Kerstens2,3, Renske M J Janssen4, Alexander C H Geurts4, Nens van Alfen5, Jan T Groothuis4.
Abstract
BACKGROUND: Strength testing of the serratus anterior muscle with hand held dynamometry (HDD) in supine subjects has low reproducibility, and is influenced by compensatory activity of other muscles like the pectoralis major and upper trapezius. Previously, two manual maximum voluntary isometric contraction tests of the serratus anterior muscle were reported that recruited optimal surface electromyography (sEMG) activity in a sitting position. We adapted three manual muscle tests to make them suitable for HHD and investigated their validity and reliability.Entities:
Keywords: Muscle strength dynamometer; Reproducibility of results; Scapula; Serratus anterior; Strength testing
Mesh:
Year: 2019 PMID: 31391035 PMCID: PMC6686461 DOI: 10.1186/s12891-019-2741-7
Source DB: PubMed Journal: BMC Musculoskelet Disord ISSN: 1471-2474 Impact factor: 2.362
Fig. 1Test positions and surface electromyography placements for serratus anterior muscle testing, note; depicted angles differ from actual goniometry angles measured while testing. a 90°of shoulder flexion in the frontal plane. b 90° of shoulder flexion in scaption. c 125° of shoulder flexion in scaption
linear mixed models of electromyography activity difference estimations
| Model | Effect | Estimate | 95% CI | |
|---|---|---|---|---|
| LL | UL | |||
| Serratus anterior | Intercept (mV) | 239.54 | 173.78 | 330.13 |
| Pos A | 1.09 | .78 | 1.38 | |
| Pos B | 1.02 | .81 | 1.24 | |
| Pos C | 1 | – | – | |
| Pectoralis minor | Intercept (mV) | 33.04 | 23.23 | 45.04 |
| Pos A | 2.94 | 2.13 | 4.05 | |
| Pos B | 1.32 | .96 | 1.83 | |
| Pos C | 1 | – | – | |
| Trapezius descendens | Intercept (mV) | 89.88 | 28.34 | 53.82 |
| Pos A | .43 | .29 | .65 | |
| Pos B | .63 | .42 | .95 | |
| Pos C | 1 | – | – | |
Legend: 95% CI, 95% confidence interval; LL, Lower limit; UL, upper limit; mV, microvolts; Pos A, test position A; Pos B, test position B; Pos C, test position C (reference)
Fig. 2Linear mixed models analysis of EMG activity in micro Volts. EMG, electromyography; Pos A, test position A; Pos B, test position B; Pos C, test position C. Note: the scaling for serratus anterior sEMG activity is different
test-retest and interrater reproducibility of serratus anterior hand held dynamometer strenght testing
|
|
|
|
|
|
| ||
|---|---|---|---|---|---|---|---|
|
| |||||||
| Pos A | 383.88 (77.65) | 404.92 (84.96) | −21.05 (60.22) | − 139.08; 96.98 | 0.712 (0.420; 0.871 | 44,10 | 122,40 |
| Pos B | 314.43 (75.1) | 322.34 (82.34) | −7.90 (65.90) | − 137.06; 121.26 | 0.658 (0.324; 0.846) | 45,80 | 127,00 |
| Pos C | 351,54 (100.36) | 376.16 (93.55) | −24.62 (59.37) | − 140.99; 91.75 | 0.794 (0.490; 0.916) | 44,52 | 123,40 |
|
|
| ||||||
|
|
|
|
| ||||
| Pos A | 383.88 (77.65) | 340.44 (82.95) | 43.43* (92.74) | − 138.34; 225.20 | 0.794; (0.552; 0.912) | 55,80 | 196,80 |
|
| 314.43 (75.10) | 264.21 (61.76) | 50.21* (78.85) | − 104.34; 204.76 | 0.277 (− 0.089; 0.605) | 55,80 | 154,50 |
| Pos C | 351.54 (100.36) | 265.10 (40,0)) | 86.48* (86.21) | −82.49; 255.45 | 0.226 (−0.107; 559) | 85,31 | 236,47 |
N Newton, SD Standard deviation, Diff Difference, LoA limits of agreement, ICC Intraclass correlation coefficient, CI Confidence interval, SEM Standard error of measurement, SDC Smallest detectable change, % percentage, * p < 0.001
Legend: ICC 3.1 intraclass correlation coefficient model 3.1, 95% CI, 95% confidence interval, LL Lower limit, UL upper limit
Fig. 3Bland Altman plots for test-retest differences and their relation to the magnitude of strength measured with HHD in Newtons