Literature DB >> 31367564

Glomerular filtration rate measured by 99mTc-DTPA Gates method is not significantly affected by the premature or delayed initiation of image acquisition.

Guangyu Ma1, Mingzhe Shao1, Baixuan Xu1, Jiahe Tian1, Yingmao Chen1.   

Abstract

BACKGROUND: This study was performed to examine the effect of non-synchronization of the radiotracer injection and image acquisition on estimates of the glomerular filtration rate (GFR) by the Gates' method.
METHODS: A total of 218 volunteers were selected as the research subjects. Two-sample method (GFRdt) and 99mTc-DTPA dynamic renal imaging (GFRGates) were used for determination of the GFR. We took GFRdt as the reference method, and then took the peak time of blood perfusion phase as the new time origin to ensure that all patients were unified on the time-radioactivity count rate curve. We moved the radioactivity curve on 9 time points to simulate premature (+20/+15/+10/+5 seconds), synchronous (0 seconds), and delayed (-20/-15/-10/-5 seconds) image acquisition in relation to the completion of tracer injection; we then acquired 9 GFRGates. The correlation and consistency of GFRGates and GFRdt were analyzed. Variance analysis compared the differences between different GFRGates.
RESULTS: All 9 GFRGates had good correlation with GFRdt. GFRdt and GFRGates derived from -5, -10 and -15 s had the best correlation (r=0.827, P<0.01). The consistency between GFRGates derived from +20 s and GFRdt was the worst, and GFRGates derived from -15 s and GFRdt was the best. There were no significant differences between the 9 GFRGates.
CONCLUSIONS: Non-synchronization of the radiotracer injection and image acquisition has no significant effect on the estimates of the GFRGates if the premature or delayed time between image acquisition and tracer injection is not more than 20 seconds.

Entities:  

Keywords:  Glomerular filtration rate (GFR); SPECT; dual plasma sample method; radionuclide renography

Year:  2019        PMID: 31367564      PMCID: PMC6629559          DOI: 10.21037/qims.2019.06.14

Source DB:  PubMed          Journal:  Quant Imaging Med Surg        ISSN: 2223-4306


  12 in total

1.  Comparison of methods for determination of glomerular filtration rate: Tc-99m-DTPA renography, predicted creatinine clearance method and plasma sample method.

Authors:  Kazuo Itoh
Journal:  Ann Nucl Med       Date:  2003-10       Impact factor: 2.668

2.  A formula to estimate the approximate surface area if height and weight be known. 1916.

Authors:  D Du Bois; E F Du Bois
Journal:  Nutrition       Date:  1989 Sep-Oct       Impact factor: 4.008

3.  Factors affecting the Gates' measurement of glomerular filtration rate.

Authors:  M Awdeh; K Kouris; I M Hassan; H M Abdel-Dayem
Journal:  Am J Physiol Imaging       Date:  1990

4.  Attenuation correction in evaluating renal function in children and adults by a camera-based method.

Authors:  Y Inoue; K Yoshikawa; T Suzuki; N Katayama; I Yokoyama; T Kohsaka; Y Tsukune; K Ohtomo
Journal:  J Nucl Med       Date:  2000-05       Impact factor: 10.057

5.  Split renal function testing using Tc-99m DTPA. A rapid technique for determining differential glomerular filtration.

Authors:  G F Gates
Journal:  Clin Nucl Med       Date:  1983-09       Impact factor: 7.794

6.  Estimation of glomerular filtration rate from plasma clearance of 51-chromium edetic acid.

Authors:  C Chantler; T M Barratt
Journal:  Arch Dis Child       Date:  1972-08       Impact factor: 3.791

7.  Establish New Formulas for the Calculation of Renal Depth in Both Children and Adults.

Authors:  Guangyu Ma; Mingzhe Shao; Baixuan Xu; Jiahe Tian; Yingmao Chen
Journal:  Clin Nucl Med       Date:  2015-07       Impact factor: 7.794

Review 8.  The kinetic basis of glomerular filtration rate measurement and new concepts of indexation to body size.

Authors:  A M Peters
Journal:  Eur J Nucl Med Mol Imaging       Date:  2003-10-31       Impact factor: 9.236

9.  Comparison of methods for determination of glomerular filtration rate: low and high-dose Tc-99m-DTPA renography, predicted creatinine clearance method, and plasma sample method.

Authors:  Majid Assadi; Mohammad Eftekhari; Mehrdad Hozhabrosadati; Mohsen Saghari; Abdolali Ebrahimi; Iraj Nabipour; Mohammad Zaki Abbasi; Darab Moshtaghi; Moloud Abbaszadeh; Sakineh Assadi
Journal:  Int Urol Nephrol       Date:  2008-08-09       Impact factor: 2.370

10.  Improved formulas for the estimation of renal depth in adults.

Authors:  A Taylor; C Lewis; A Giacometti; E C Hall; K P Barefield
Journal:  J Nucl Med       Date:  1993-10       Impact factor: 10.057

View more
  5 in total

1.  The Application Value of the Renal Region of Interest Corrected by Computed Tomography in Single-Kidney Glomerular Filtration Rate for the Evaluation of Patients With Moderate or Severe Hydronephrosis.

Authors:  Haixia Yin; Wenli Liang; Deshan Zhao
Journal:  Front Physiol       Date:  2022-05-09       Impact factor: 4.755

2.  Establishment of normal reference ranges for total and separate glomerular filtration rates in living kidney donors by Gates' method after renal depth measured using single-photon emission computed tomography/computed tomography.

Authors:  Yushan Wei; Yan Liu; Qiang Li; Yuemin Zhang; Qi Wang; Yiyuan Yang; Yiqian Liang; Aomei Zhao; Aimin Yang; Jianjun Xue
Journal:  Quant Imaging Med Surg       Date:  2020-12

3.  Influence of Early Bladder Imaging in Experimental Rabbits on the Quantitative Determination of Glomerular Filtration Rate by the Gates Method.

Authors:  Changyin Wang; Shun Li; Chun Gao; Wasili Maimaiti; Qisheng Yang; Linglong Jiang
Journal:  Biomed Res Int       Date:  2020-11-26       Impact factor: 3.411

4.  The Application of 99mTc-DTPA Renal Dynamic Imaging to Measuring Renal Function of Children with Acute Lymphoblastic Leukemia after Induction Therapy.

Authors:  Lidan Wang; Kailan Chen; Qiong Xu
Journal:  Biomed Res Int       Date:  2020-11-17       Impact factor: 3.411

5.  The roles of protein load test in quantitatively evaluating renal function after severe unilateral ureteral obstruction in adult rabbits.

Authors:  Changyin Wang; Chun Gao; Shun Li; Maimaiti Wasili; Qisheng Yang; Linglong Jiang
Journal:  Quant Imaging Med Surg       Date:  2021-08
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.