Exploring a new method for quantitative sodium MRI in the human upper leg with a surface coil and symmetrically arranged reference phantoms.

Bastien Milani1,2,3, Jean Delacoste2,3, Michel Burnier1, Menno Pruijm1.   

Abstract

BACKGROUND: The aim of this study is to validate and evaluate the reproducibility of a new setup for the quantification of the tissue sodium concentration (TSC) in the human upper leg muscles with sodium MRI at 3 Tesla. This setup is making use of an emit and receive single loop surface coil together with a set of square, symmetrically arranged reference phantoms. As a second aim, the performances of two MRI protocols for the TSC quantification in the upper leg muscles are compared: one using an ultra-short echo time (UTE) 3-dimensional radial sequence (UTE-protocol), and the other one using standard gradient echo sequence (GRE-protocol).
METHODS: A validation test of the quantification of sodium concentration is performed in phantoms. The bias of the method is estimated and compared between both protocols. The reproducibility of TSC quantification is assessed in phantoms by the coefficient of variation (CV) and compared between both protocols. The reproducibility is also assessed in 11 health volunteers. Signal to noise ratio (SNR) maps are acquired in phantoms with both protocols in order to compare the resulting SNR.
RESULTS: The apparatus and post processing were successfully implemented. The bias of the method was smaller than 10% in phantoms (excepted for Na concentration of 10 mmol/L when using the GRE protocol). The reproducibility of the method using symmetrically arranged phantoms was high in phantoms and humans (CV <5%). The GRE-protocol leads to a better SNR than the UTE-protocol in 2D images.
CONCLUSIONS: The use of symmetrically arranged reference phantoms lead to reproducible results in phantoms and humans. Sodium imaging in the human upper leg with a single loop surface coil should be performed with a standard 2-dimensional GRE protocol if an optimal SNR is needed. However, the quantification of the fast and slow decay time constants of the sodium signal, which plays a role in the TSC quantification, still has to be done with a UTE sequence. Moreover, the quantification of sodium concentration is more accurate with the UTE protocol for small sodium concentrations (<20 mmol).

Entities:  

Keywords:  Gradient echo sequence (GRE); quantitative; sodium; surface coil; symmetrically; ultra-short echo time (UTE)

Year:  2019        PMID: 31367553      PMCID: PMC6629565          DOI: 10.21037/qims.2019.06.08

Source DB:  PubMed          Journal:  Quant Imaging Med Surg        ISSN: 2223-4306


  33 in total

1.  Measuring human cardiac tissue sodium concentrations using surface coils, adiabatic excitation, and twisted projection imaging with minimal T2 losses.

Authors:  Ronald Ouwerkerk; Robert G Weiss; Paul A Bottomley
Journal:  J Magn Reson Imaging       Date:  2005-05       Impact factor: 4.813

2.  Design and analysis of a practical 3D cones trajectory.

Authors:  Paul T Gurney; Brian A Hargreaves; Dwight G Nishimura
Journal:  Magn Reson Med       Date:  2006-03       Impact factor: 4.668

3.  Sodium MRI using a density-adapted 3D radial acquisition technique.

Authors:  Armin M Nagel; Frederik B Laun; Marc-André Weber; Christian Matthies; Wolfhard Semmler; Lothar R Schad
Journal:  Magn Reson Med       Date:  2009-12       Impact factor: 4.668

4.  Two-dimensional radial acquisition technique with density adaption in sodium MRI.

Authors:  Simon Konstandin; Armin M Nagel; Patrick M Heiler; Lothar R Schad
Journal:  Magn Reson Med       Date:  2010-11-04       Impact factor: 4.668

5.  Validation of sodium magnetic resonance imaging of intervertebral disc.

Authors:  Chenyang Wang; Erin McArdle; Matthew Fenty; Walter Witschey; Mark Elliott; Matthew Sochor; Ravinder Reddy; Arijitt Borthakur
Journal:  Spine (Phila Pa 1976)       Date:  2010-03-01       Impact factor: 3.468

6.  Phase-sensitive sodium B1 mapping.

Authors:  Steven P Allen; Glen R Morrell; Brock Peterson; Danny Park; Garry E Gold; Joshua D Kaggie; Neal K Bangerter
Journal:  Magn Reson Med       Date:  2010-11-30       Impact factor: 4.668

7.  Elevated tissue sodium concentration in malignant breast lesions detected with non-invasive 23Na MRI.

Authors:  Ronald Ouwerkerk; Michael A Jacobs; Katarzyna J Macura; Antonio C Wolff; Vered Stearns; Sarah D Mezban; Nagi F Khouri; David A Bluemke; Paul A Bottomley
Journal:  Breast Cancer Res Treat       Date:  2007-01-27       Impact factor: 4.872

8.  Quantitative sodium imaging with a flexible twisted projection pulse sequence.

Authors:  Aiming Lu; Ian C Atkinson; Theodore C Claiborne; Frederick C Damen; Keith R Thulborn
Journal:  Magn Reson Med       Date:  2010-06       Impact factor: 4.668

9.  Sodium magnetic resonance imaging of chemotherapeutic response in a rat glioma.

Authors:  Victor D Schepkin; Brian D Ross; Thomas L Chenevert; Alnawaz Rehemtulla; Surabhi Sharma; Mahesh Kumar; Jadranka Stojanovska
Journal:  Magn Reson Med       Date:  2005-01       Impact factor: 4.668

10.  Tissue sodium concentration in myocardial infarction in humans: a quantitative 23Na MR imaging study.

Authors:  Ronald Ouwerkerk; Paul A Bottomley; Meiyappan Solaiyappan; Amy E Spooner; Gordon F Tomaselli; Katherine C Wu; Robert G Weiss
Journal:  Radiology       Date:  2008-07       Impact factor: 11.105

View more
  1 in total

1.  Repeatability assessment of sodium (23Na) MRI at 7.0 T in healthy human calf muscle and preliminary results on tissue sodium concentrations in subjects with Addison's disease.

Authors:  Olgica Zaric; Hannes Beiglböck; Veronika Janacova; Pavol Szomolanyi; Peter Wolf; Michael Krebs; Siegfried Trattnig; Martin Krššák; Vladimir Juras
Journal:  BMC Musculoskelet Disord       Date:  2022-10-20       Impact factor: 2.562

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.