| Literature DB >> 31360010 |
Mohandoss Murugesan1, Chandran K Nair2, Sangeetha K Nayanar3, Kalyana Chakravarthy Pentapati4.
Abstract
BACKGROUND: Flow cytometric enumeration of CD34+ hematopoietic stem cells (HSC) is the reference point for undertaking apheresis and evaluation of adequacy for peripheral blood stem cell (PBSC) engraftment. AIMS: To determine whether single platform correlates with dual platform methods in CD34+ enumeration using ISHAGE protocol.Entities:
Keywords: CD34+ cells enumeration; International Society of Hematotherapy and Graft Engineering; flow cytometry; hematopoietic stem cells
Year: 2019 PMID: 31360010 PMCID: PMC6580824 DOI: 10.4103/ajts.AJTS_83_18
Source DB: PubMed Journal: Asian J Transfus Sci ISSN: 0973-6247
Enumeration of CD34 cells in peripheral blood and peripheral blood stem cell product using single- and dual-platform assay from a tertiary cancer center in South India
| Sample | Mean±SD | |||
|---|---|---|---|---|
| Single platform | Dual platform | |||
| Peripheral blood | 41 | 59.8±39.8 | 55.9±38.0 | 0.112 |
| PBSC product | 111 | 1159.4±922.3 | 991.0±951.6 | <0.001 |
PBSC=Peripheral blood stem cell, SD=Standard deviation
Correlation between single- and dual-platform assay of CD34+counts in peripheral blood sample and peripheral blood stem cell product sample from a tertiary cancer center in South India
| Sample | |||
|---|---|---|---|
| Peripheral blood | 41 | 0.92 | <0.001 |
| PBSC product | 111 | 0.85 | <0.001 |
PBSC=Peripheral blood stem cell
Figure 1Correlation of single- and dual-platform assay for CD34 enumeration in peripheral blood sample from a tertiary cancer center in South India