Bashir Zikria1, Nima Hafezi-Nejad1, Ian Patten1, Alex Johnson1, Arya Haj-Mirzaian1, John H Wilckens1, James R Ficke1, Shadpour Demehri1.
Abstract
BACKGROUND: Autologous chondrocyte implantation (ACI), a promising modality for repairing full-thickness cartilage defects, requires 2 consecutive arthroscopic procedures for chondrocyte harvesting and implantation. In the present study, we assessed the feasibility and efficacy of image-guided chondrocyte harvesting as an alternative to arthroscopic biopsy.Entities:
Year: 2019 PMID: 31334460 PMCID: PMC6613850 DOI: 10.2106/JBJS.OA.18.00039
Source DB: PubMed Journal: JB JS Open Access ISSN: 2472-7245
Fig. 1Fig. 1-A Preprocedural axial CT scan used for the determination of the target site (red arrow and dashed line) for the performance of image-guided chondrocyte retrieval. Fig. 1-B Axial CTA scan made following the intra-articular administration of Omnipaque 350, used to measure the size of the cartilage defect. Fig. 1-C Axial CTA scan made after CTA-guided chondrocyte harvesting; the distance between the predefined target on the preprocedural CT scan and the final insertion site of the needle (red arrow) was calculated as the distance between the 2 red dashed lines (marked by the yellow bidirectional arrow).
Fig. 2Figs. 2-A, 2-B, and 2-C Serial axial CTA slices demonstrating the harvesting needle pathway used to obtain cartilage tissue. Fig. 2-D Cartilage tissue and accompanying subchondral bone obtained with use of CTA-guided needle harvesting.
Absolute Agreement, Consistency, and Accuracy of CTA Measurements in Estimating Full-Thickness Femoral Cartilage Defects*
| Defect Height | Defect Width | |||||
| Defect Site | Absolute Agreement | Consistency | Accuracy | Absolute Agreement | Consistency | Accuracy |
| Trochlea | 0.996 | 0.997 | 0 (2.0) | 0.952 | 0.946 | 0 (3.0) |
| Medial condyle | 0.970 | 0.987 | 2.0 (2.0) | 0.848 | 0.828 | 0 (3.0) |
| Lateral condyle | 0.938 | 0.939 | 0.5 (3.5) | 0.404 | 0.429 | 1.0 (3.8) |
| All | 0.981 | 0.983 | 0.5 (2.0) | 0.936 | 0.933 | 0 (3.0) |
For interclass correlation coefficients (ICCs), a higher value (closer to 1) was indicative of higher absolute agreement and consistency. For accuracy, which was measured as the Euclidean difference (in millimeters) between the CTA estimates and the direct measurements of the full-thickness lesions, lower values were indicative of higher accuracy.
The values are given as the median, with the interquartile range in parentheses.
P < 0.001.
P < 0.01.
P < 0.05.
Performance Measures for 2 Operators When Obtaining Chondrocytes from Medial and Lateral Trochlear Ridges*
| Operator 1 | Operator 2 | |
| Time to obtain tissue | ||
| Medial ridge | 106 ± 36 | 111 ± 35 |
| Lateral ridge | 74 ± 34 | 72 ± 30 |
| P value | 0.056 | 0.014 |
| Accuracy | ||
| Medial ridge | 3.23 ± 1.72 | 2.81 ± 1.36 |
| Lateral ridge | 1.32 ± 1.01 | 1.17 ± 0.57 |
| P value | 0.007 | 0.040 |
| Needle readjustment attempts | ||
| Medial ridge | 1.10 ± 0.99 | 1.00 (0.81) |
| Lateral ridge | 1.00 (1.00) | 1.00 (1.25) |
| P value | 0.315 | 0.739 |
The values are given as the mean and the standard deviation (for normally distributed values) or as the median with the width of the interquartile range in parentheses (for non-normally distributed values). The values for the lateral and medial trochlear ridges were compared with use of the independent 2-samples t test or the non-parametric Mann-Whitney U test. Accuracy was measured as the Euclidean distance from the target (in millimeters); lower values were indicative of higher accuracy.
Significant (p < 0.05).
Comparison of Conventional Arthroscopic Harvesting and Needle Harvesting of Chondrocytes in Clinical Study*
| Conventional Arthroscopic Harvesting | Needle Harvesting | P Value | |
| Total weight prior to bone/synovium removal | 215 (497.5) | 101 ± 40.7 | 0.013 |
| Net weight after bone/synovium removal | 170 (272.5) | 82 ± 54.5 | 0.013 |
| Processed weight | 176 ± 100 | 82 ± 52.5 | 0.011 |
| No. of cells/mg | 2,749 ± 1,090 | 1,365 (2,022) | 0.093 |
| No. of viable cells | 1.35 × 107 ± 9.73 × 106 | 5.33 × 106 ± 1.48 × 106 | 0.030 |
| Growth index | 0.494 ± 0.089 | 0.488 ± 0.138 | 0.897 |
The values are given as the mean and the standard deviation (for normally distributed values) or as the median with the width of the interquartile range in parentheses (for non-normally distributed values). The values were compared with use of the paired t test or the paired Wilcoxon signed-rank test.
Needle-harvested chondrocytes were obtained from only 1 core sample from each subject.
Significant (p < 0.05).