Chang Heon Choi1,2,3,4, Jin Ho Kim1,2,3, Jung-In Kim1,2,3, Jong Min Park1,2,3,5.
Abstract
PURPOSE: This study compares the plan quality of magnetic-resonance image (MRI)-based intensity modulated radiation therapy (IMRT) using a linac (MR-linac-IMRT), MRI-based IMRT using tri-Co-60 sources (MR-Co-60-IMRT), and volumetric modulated arc therapy (VMAT) for spine stereotactic ablative radiotherapy (SABR).Entities:
Year: 2019 PMID: 31329641 PMCID: PMC6645671 DOI: 10.1371/journal.pone.0220039
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Average treatment plan parameters.
| Parameters | VMAT | MR-linac IMRT | MR-Co-60 IMRT | |
|---|---|---|---|---|
| Beam-off time (min) | 0.02 ± 0.00 | 5.61 ± 0.50 | - | |
| Beam-on time (min) | 4.0 ± 1.1 | 27.6 ± 5.1 | 49.7 ± 11.1 | |
| Monitor unit (MU) | 6037 ± 799 | 16430 ± 3117 | - | |
| Number of beam segments | 354 | 60 ± 1 | 60 ± 5 | 0.264 |
VMAT: volumetric modulated arc therapy, MR-linac: linear accelerator with magnetic resonance imaging system, IMRT: intensity modulated radiation therapy, MR-Co-60: radiation therapy system using tri-Co-60 sources with magnetic resonance imaging system, MLC: multileaf collimator, MU: monitor unit.
Average dose-volumetric parameters of the clinical target volume (CTV) of VMAT, MR-linac IMRT, and MR-Co-60 IMRT plans.
| Dose-volumetric parameter | VMAT | MR-linac IMRT | MR Co-60 IMRT | |
|---|---|---|---|---|
| CTV volume (cm3) | 57.0 ± 48.8 | |||
| D98% (Gy) | 15.29 ± 2.30 | 15.27 ± 1.67 | 10.89 ± 2.16 | |
| D90% (Gy) | 17.45 ± 1.15 | 17.74 ± 0.65 | 12.09 ± 3.06 | |
| D80% (Gy) | 17.67 ± 0.85 | 18.16 ± 0.09 | 12.80 ± 3.66 | |
| D5% (Gy) | 18.33 ± 0.84 | 19.55 ± 0.04 | 15.06 ± 5.17 | |
| D2% (Gy) | 18.48 ± 0.88 | 19.75 ± 0.31 | 15.27 ± 5.28 | |
| Minimum dose (Gy) | 10.54 ± 0.88 | 9.95 ± 0.34 | 9.04 ± 1.46 | |
| Maximum dose (Gy) | 19.24 ± 0.89 | 20.81 ± 0.37 | 16.11 ± 5.62 | |
| Mean dose (Gy) | 17.79 ± 3.18 | 18.53 ± 2.72 | 13.63 ± 4.21 | |
| Conformity index (CI) | 0.853 ± 0.209 | 1.027 ± 0.087 | 1.451 ± 0.551 | |
| Homogeneity index (HI) | 0.089 ± 0.046 | 0.140 ± 0.047 | 0.282 ± 0.035 | |
VMAT: volumetric modulated arc therapy, MR-linac: linear accelerator with magnetic resonance imaging system, IMRT: intensity modulated radiation therapy, MR-Co-60: radiation therapy system using tri-Co-60 sources with magnetic resonance imaging system, Dn%: dose received by n% volume of the CTV.
Fig 1Dose distributions of the volumetric modulated arc therapy (VMAT) plans, intensity modulated radiation therapy (IMRT) plans with the MRIdian Linac system (MR-linac), and IMRT plans with the ViewRay system (MR-Co-60) of two representative cases for spine stereotactic ablative radiotherapy (SABR).
One (patient No. 2) is a representative case with a prescription dose of 18 Gy for VMAT (a), MR-linac IMRT (b), and MR-Co-60 IMRT plans (c). The other (patient No. 17) is a representative case with prescription doses of 16 Gy, 18 Gy and 9.5 Gy for VMAT (d), MR-linac IMRT (e), and MR-Co-60 IMRT plans (f), respectively.
Fig 2Dose volume histograms (DVHs) of the volumetric modulated arc therapy (VMAT) plans, intensity modulated radiation therapy (IMRT) plans with the MRIdian Linac system (MR-linac), and IMRT plans with the ViewRay system (MR-Co-60) of two representative cases for spine stereotactic ablative radiotherapy (SABR).
One (patient No. 2) is a representative case with a prescription dose of 18 Gy for VMAT, MR-linac IMRT, and MR-Co-60 IMRT plans (a). The other (patient No. 17) is a representative case with prescription doses of 16 Gy, 18 Gy and 9.5 Gy for VMAT, MR-linac IMRT, and MR-Co-60 IMRT plans, respectively (b). The DVHs of VMAT, MR-linac IMRT, and MR-Co-60 IMRT plans are plotted as solid, dashed and dotted lines, respectively (black color: whole body, red color: clinical target volume, blue color: spinal cord, and green color: spinal cord planning organ-at-risk volume).
Average dose-volumetric parameters of the spinal cord, spinal cord PRV, and whole body of VMAT, MR-linac IMRT, and MR-Co-60 IMRT plans.
| Dose-volumetric parameter | VMAT | MR-linac IMRT | MR Co-60 IMRT | |
|---|---|---|---|---|
| Spinal cord | ||||
| Volume (cm3) | 3.8 ± 1.6 | 1.1 ± 1.0 | ||
| D10% (Gy) | 8.57 ± 1.69 | 7.99 ± 2.45 | 9.80 ± 1.56 | 0.020 |
| V10Gy (cm3) | 0.2 ± 0.1 | 0.1 ± 0.2 | 0.1± 0.2 | 0.279 |
| V14Gy (cm3) | 0.0 ± 0.0 | 0.0 ± 0.0 | 0.0 ± 0.0 | 0.145 |
| Maximum dose (Gy) | 11.39 ± 2.15 | 9.38 ± 3.13 | 10.40 ± 1.59 | |
| Spinal cord PRV | ||||
| Volume (cm3) | 8.0 ± 2.7 | 3.2 ± 1.9 | ||
| D10% (Gy) | 10.99 ± 2.69 | 9.81 ± 1.69 | 10.07 ± 1.45 | |
| V10Gy (cm3) | 1.4 ± 1.2 | 0.2 ± 0.5 | 0.5 ± 0.6 | |
| V14Gy (cm3) | 0.2 ± 0.2 | 0.0 ± 0.1 | 0.0 ± 0.0 | |
| Maximum dose (Gy) | 15.02 ± 2.36 | 12.24 ± 4.37 | 10.98 ± 1.58 | |
| Whole body | ||||
| V50% (cm3) | 240.1 ± 172.2 | 325.5 ± 256.2 | 501.2 ± 485.7 | 0.049 |
| V100% (cm3) | 56.4 ± 45.2 | 60.5 ± 54.9 | 76.8 ±73.4 | |
| Gradient index (GI) | 4.49 ± 0.64 | 5.74 ± 1.01 | 6.85 ±1.56 | |
PRV: planning organ-at-risk volume, VMAT: volumetric modulated arc therapy, MR-linac: linear accelerator with magnetic resonance imaging system, IMRT: intensity modulated radiation therapy, MR-Co-60: radiation therapy system using tri-Co-60 sources with magnetic resonance imaging system, Dn%: dose received by n% volume of a structure, VnGy (Vn%): volume receiving more than n Gy (or n% of the prescription dose) of a structure.