An automated approach to fully self-gated free-running cardiac and respiratory motion-resolved 5D whole-heart MRI.

Lorenzo Di Sopra1, Davide Piccini1,2, Simone Coppo1, Matthias Stuber1,3, Jérôme Yerly1,3.   

Abstract

PURPOSE: To develop a previously reported, electrocardiogram (ECG)-gated, motion-resolved 5D compressed sensing whole-heart sparse MRI methodology into an automated, optimized, and fully self-gated free-running framework in which external gating or triggering devices are no longer needed.
METHODS: Cardiac and respiratory self-gating signals were extracted from raw image data acquired in 12 healthy adult volunteers with a non-ECG-triggered 3D radial golden-angle 1.5 T balanced SSFP sequence. To extract cardiac self-gating signals, central k-space coefficient signal analysis (k0 modulation), as well as independent and principal component analyses were performed on selected k-space profiles. The procedure yielding triggers with the smallest deviation from those of the reference ECG was selected for the automated protocol. Thus, optimized cardiac and respiratory self-gating signals were used for binning in a compressed sensing reconstruction pipeline. Coronary vessel length and sharpness of the resultant 5D images were compared with image reconstructions obtained with ECG-gating.
RESULTS: Principal component analysis-derived cardiac self-gating triggers yielded a smaller deviation ( 17.4 ± 6.1 ms ) from the reference ECG counterparts than k0 modulation ( 26 ± 7.5 ms ) or independent component analysis ( 19.8 ± 5.2 ms ). Cardiac and respiratory motion-resolved 5D images were successfully reconstructed with the automated and fully self-gated approach. No significant difference was found for coronary vessel length and sharpness between images reconstructed with the fully self-gated and the ECG-gated approach (all P ≥ . 06 ).
CONCLUSION: Motion-resolved 5D compressed sensing whole-heart sparse MRI has successfully been developed into an automated, optimized, and fully self-gated free-running framework in which external gating, triggering devices, or navigators are no longer mandatory. The resultant coronary MRA image quality was equivalent to that obtained with conventional ECG-gating.
© 2019 International Society for Magnetic Resonance in Medicine.

Keywords:  3D radial trajectory; 5D whole-heart MRI; compressed sensing; coronary MRA; free-running; self-gating

Mesh:

Substances:

Year:  2019        PMID: 31321816     DOI: 10.1002/mrm.27898

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  14 in total

1.  Combined T2 -preparation and multidimensional outer volume suppression for coronary artery imaging with 3D cones trajectories.

Authors:  David Y Zeng; Corey A Baron; Mario O Malavé; Adam B Kerr; Phillip C Yang; Bob S Hu; Dwight G Nishimura
Journal:  Magn Reson Med       Date:  2019-11-05       Impact factor: 4.668

2.  Natively fat-suppressed 5D whole-heart MRI with a radial free-running fast-interrupted steady-state (FISS) sequence at 1.5T and 3T.

Authors:  Jessica A M Bastiaansen; Davide Piccini; Lorenzo Di Sopra; Christopher W Roy; John Heerfordt; Robert R Edelman; Ioannis Koktzoglou; Jérôme Yerly; Matthias Stuber
Journal:  Magn Reson Med       Date:  2019-08-27       Impact factor: 4.668

3.  Free-running cardiac and respiratory motion-resolved 5D whole-heart coronary cardiovascular magnetic resonance angiography in pediatric cardiac patients using ferumoxytol.

Authors:  Christopher W Roy; Lorenzo Di Sopra; Kevin K Whitehead; Davide Piccini; Jérôme Yerly; John Heerfordt; Reena M Ghosh; Mark A Fogel; Matthias Stuber
Journal:  J Cardiovasc Magn Reson       Date:  2022-06-27       Impact factor: 6.903

4.  Self-gated 3D stack-of-spirals UTE pulmonary imaging at 0.55T.

Authors:  Ahsan Javed; Rajiv Ramasawmy; Kendall O'Brien; Christine Mancini; Pan Su; Waqas Majeed; Thomas Benkert; Himanshu Bhat; Anthony F Suffredini; Ashkan Malayeri; Adrienne E Campbell-Washburn
Journal:  Magn Reson Med       Date:  2021-11-16       Impact factor: 3.737

5.  Volumetric coronary endothelial function assessment: a feasibility study exploiting stack-of-stars 3D cine MRI and image-based respiratory self-gating.

Authors:  Gabriele Bonanno; Robert G Weiss; Davide Piccini; Jérôme Yerly; Sahar Soleimani; Li Pan; Xiaoming Bi; Allison G Hays; Matthias Stuber; Michael Schär
Journal:  NMR Biomed       Date:  2021-07-21       Impact factor: 4.044

6.  5D Flow MRI: A Fully Self-gated, Free-running Framework for Cardiac and Respiratory Motion-resolved 3D Hemodynamics.

Authors:  Liliana E Ma; Jérôme Yerly; Davide Piccini; Lorenzo Di Sopra; Christopher W Roy; James C Carr; Cynthia K Rigsby; Daniel Kim; Matthias Stuber; Michael Markl
Journal:  Radiol Cardiothorac Imaging       Date:  2020-11-12

7.  Using 5D flow MRI to decode the effects of rhythm on left atrial 3D flow dynamics in patients with atrial fibrillation.

Authors:  Liliana Ma; Jérôme Yerly; Lorenzo Di Sopra; Davide Piccini; Jeesoo Lee; Amanda DiCarlo; Rod Passman; Philip Greenland; Daniel Kim; Matthias Stuber; Michael Markl
Journal:  Magn Reson Med       Date:  2021-01-05       Impact factor: 3.737

8.  Fully self-gated whole-heart 4D flow imaging from a 5-minute scan.

Authors:  Aaron Pruitt; Adam Rich; Yingmin Liu; Ning Jin; Lee Potter; Matthew Tong; Saurabh Rajpal; Orlando Simonetti; Rizwan Ahmad
Journal:  Magn Reson Med       Date:  2020-09-30       Impact factor: 4.668

Review 9.  From Compressed-Sensing to Artificial Intelligence-Based Cardiac MRI Reconstruction.

Authors:  Aurélien Bustin; Niccolo Fuin; René M Botnar; Claudia Prieto
Journal:  Front Cardiovasc Med       Date:  2020-02-25

10.  Variable anisotropic FOV for 3D radial imaging with spiral phyllotaxis (VASP).

Authors:  Guruprasad Krishnamoorthy; Jouke Smink; Joao Tourais; Marcel Breeuwer; Marc Kouwenhoven
Journal:  Magn Reson Med       Date:  2020-08-27       Impact factor: 4.668

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.